Skip to content

Hollister 9782Recalls & reported problems

Hollister Incorporated · Vertical Tube Attachment Device - Accomodates 5Fr to 40Fr Tubes · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00610075097821), received between Apr 9, 2024 and Apr 9, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 11, 2021Class IITerminated

    The tube holder can separate from the barrier that secures the device to the patients' body. The tube holder has the potential failure to separate from the barrier. This separation might contribute to the potential risk of the tube/drain/catheter migration or loss.

    Z-1311-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT THE CLAMP PORTION SECURING A 16 FRENCH TUBE SEPARATED FROM THE BARRIER PORTION OF A HOLLISTER VERTICAL DRAIN TUBE ATTACHMENT DEVICE. IT WAS REPORTED THAT THE SEPARATION OCCURRED SOMETIME BETWEEN HOME HEALTH CARE NURSE VISITS WHICH OCCURRED EVERY 3 DAYS. IT WAS FURTHER REPORTED THAT THERE WAS NO INJURY OR HARM TO THE END USER ASSOCIATED WITH THIS SEPARATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19069779 ↗

FDA clearance and classification

Cleared through 510(k) K932855 on Oct 13, 1993 (decision: Substantially Equivalent). Submitted by Hollister, Inc. as “Hollister Vertical Drain/Tube Attachment Device”. The FDA review took 125 days. FDA record ↗

The FDA regulates product code KNT (tubes, gastrointestinal (and accessories)) under 21 CFR 876.5980 as a Class II device. 21 CFR 876.5980 ↗

Common questions about the Hollister 9782

Has the Hollister 9782 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-1311-2021 (Class II), began on Feb 11, 2021: The tube holder can separate from the barrier that secures the device to the patients' body. The tube holder has the potential failure to separate from the barrier. This separation might contribute to the potential… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Hollister 9782?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 9, 2024. That report lists device problem “Material Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Hollister 9782?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Hollister 9782 FDA approved?

It was cleared through 510(k) K932855 on Oct 13, 1993. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Hollister 9782, and is it still sold?

The FDA's device database lists Hollister Incorporated as the company responsible for it (brand Hollister, model 9782). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices

All TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page