Vertical Drain Tube Attachment Device (VTAD)… recallZ-1311-2021 · Hollister Incorporated · initiated Feb 11, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Hollister Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The tube holder can separate from the barrier that secures the device to the patients' body. The tube holder has the potential failure to separate from the barrier. This separation might contribute to the potential risk of the tube/drain/catheter migration or loss.
FDA root cause category: Under Investigation by firm
What the company is doing
Hollister has instructed the affected consignees to quarantine the affected inventory and discard the product per their local procedures for product destruction.
Recall details
- Recall number
- Z-1311-2021
- Status
- Terminated
- Initiated by the company
- Feb 11, 2021
- Recalling firm
- Hollister Incorporated
- Quantity
- 1,031 boxes (5,155 eaches)
- Product code
- KNT
- FDA submissions
- K932855
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of IL, CO, NJ, CA, AZ, TX, KS, NY, IN, SC, RI, ID, WV, OH, MT, MO, SD, MI, MA, PA, VA, FL, NE, OR, KY, GA, WA, OK, TN, NC, MD, MN, MS, LA , NM, ND and the country of Canada.
Product description
Vertical Drain Tube Attachment Device (VTAD) (Gastrointestinal tube and accessories) - Product Usage: designed to stabilize and secure a variety of catheters, surgical drains, and tubes from sizes 5-40 French.
Code information (models, lots, serial numbers)
Model: 9782. UDI: Primary Unit of Use DI Number: 00610075097821. Lot Numbers: 0K02, 1A09, and lAlO
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Hollister 9782Hollister Incorporated · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page