Skip to content

AnchorFast Guard 9800 (UDI 00610075186518)Recalls & reported problems

Hollister Incorporated · Oral Endotracheal Tube Fastener · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00610075186518), received between Jun 13, 2023 and Mar 31, 2026 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 23, 2024Open, Classified

    Hollister received reports of decreased skin barrier wear time which could can lead to tube migration

    Z-1181-2025 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Loss of or Failure to Bond1

Patient problems

  • Pressure Sore/Ulcer2
  • Insufficient Information1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A SKIN INJURY WAS OBSERVED BENEATH THE ADHESIVE BARRIER CHEEK PAD DURING REMOVAL OF THE HOLLISTER ANCHORFAST GUARD DEVICE. IT WAS REPORTED THAT THE DEVICE HAD BEEN IN PLACE FOR 7 DAYS AND DURING A PERIOD OF THAT TIME, THE PATIENT HAD BEEN PLACED IN THE LATERAL POSITION. IT WAS REPORTED THAT A MEDICATED OINTMENT AND A SKIN BARRIER WERE USED AS TREATMENT FOR THE INJURY. HOLLISTER HAS SINCE PROVIDED EDUCATION TO THE ACCOUNT.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 24749193 ↗
  • InjuryFeb 17, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AN INTUBATED PATIENT DEVELOPED A STAGE IV UPPER LIP PRESSURE INJURY WITH FULL-THICKNESS SKIN AND TISSUE LOSS AND EXPOSED TEETH WHILE USING THE HOLLISTER ANCHORFAST GUARD ENDOTRACHEAL TUBE HOLDER. IT WAS FURTHER REPORTED THAT THE WOUND EVOLVED QUICKLY AND THE PATIENT REQUIRED DEBRIDEMENT OF THE WOUND AND SURGICAL REPAIR OF HIS LIP BY MAXILLOFACIAL ORAL SURGERY.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 24367433 ↗
  • InjuryFeb 17, 2026Loss of or Failure to Bond

    IT WAS REPORTED TO HOLLISTER THROUGH AN FDA MEDWATCH THE FOLLOWING: DURING HEAD TURN OF A PRONE PATIENT, THE ANCHORFAST CLIP BROKE CAUSING ENDOTRACHEAL TUBE DISLODGEMENT.

    Patient problems: Insufficient Information

    MAUDE report 24367086 ↗
  • MalfunctionJun 13, 2023Adverse Event Without Identified Device or Use Problem

    PATIENT DEVELOPED A NON-STAGEABLE INJURY TO THE MUCOSAL LIP WHERE THE ANCHORFAST WAS LOCATED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17117083 ↗

Other DEVICE, FIXATION, TRACHEAL TUBE devices

All DEVICE, FIXATION, TRACHEAL TUBE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.