AnchorFast SlimFit 9787Recalls & reported problems
Hollister Incorporated · Oral Endotracheal Tube Fastener · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00610075095834), received between Apr 17, 2025 and Apr 28, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 23, 2024Class IIOpen, Classified
Hollister received reports of decreased skin barrier wear time which could can lead to tube migration
Z-1179-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Loss of or Failure to Bond4
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Tissue Breakdown1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2025Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED THAT THE UPPER LIP FOAM SPACER FROM AN ANCHORFAST SLIMFIT ORAL ENDOTRACHEAL TUBE FASTENER SEPARATED FROM THE PLASTIC ANCHORFAST SLIMFIT TRACK ON A PATIENT WHO HAD THE DEVICE IN PLACE FOR 7 HOURS. IT WAS REPORTED THAT THIS SEPARATION WAS NOT WITNESSED, AND THE PIECE WAS FOUND HANGING ON THE OUTSIDE OF THE PATIENT'S MOUTH. IT WAS FURTHER REPORTED THAT THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21917495 ↗MalfunctionApr 17, 2025Loss of or Failure to Bond; Loss of or Failure to Bond
THIS IS THE THIRD OF THREE REPORTS FROM THE SAME ACCOUNT WITH THE SAME SKU AND LOT NUMBER. IT WAS REPORTED THAT THE UPPER LIP FOAM SPACER FROM AN ANCHORFAST SLIMFIT ORAL ENDOTRACHEAL TUBE FASTENER SEPARATED FROM THE PLASTIC ANCHORFAST SLIMFIT TRACK WHILE IN USE ON A PATIENT. IT WAS REPORTED THAT THE DETAILS REGARDING WHERE THE SEPARATED UPPER LIP FOAM SPACER LANDED WAS NOT AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21861341 ↗MalfunctionApr 17, 2025Loss of or Failure to Bond; Loss of or Failure to Bond
THIS IS THE SECOND OF THREE REPORTS FROM THE SAME ACCOUNT WITH THE SAME SKU AND LOT NUMBER. IT WAS REPORTED THAT THE UPPER LIP FOAM SPACER FROM AN ANCHORFAST SLIMFIT ORAL ENDOTRACHEAL TUBE FASTENER SEPARATED FROM THE PLASTIC ANCHORFAST SLIMFIT TRACK WHILE IN USE ON A PATIENT. IT WAS REPORTED THAT THE DETAILS REGARDING WHERE THE SEPARATED UPPER LIP FOAM SPACER LANDED WAS NOT AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21860423 ↗MalfunctionApr 17, 2025Loss of or Failure to Bond; Loss of or Failure to Bond
THIS IS THE FIRST OF THREE REPORTS FROM THE SAME ACCOUNT WITH THE SAME SKU AND LOT NUMBER. IT WAS REPORTED THAT THE UPPER LIP FOAM SPACER FROM AN ANCHORFAST SLIMFIT ORAL ENDOTRACHEAL TUBE FASTENER SEPARATED FROM THE PLASTIC ANCHORFAST SLIMFIT TRACK WHILE IN USE ON A PATIENT. IT WAS FURTHER REPORTED THAT THERE WAS AN UPPER LIP PRESSURE INJURY. IT WAS REPORTED THAT NEITHER THE DETAILS REGARDING THE UPPER LIP PRESSURE INJURY NOR INFORMATION ABOUT WHERE THE SEPARATED UPPER LIP FOAM SPACER LANDED WAS AVAILABLE.
Patient problems: Tissue Breakdown
MAUDE report 21860156 ↗
Other DEVICE, FIXATION, TRACHEAL TUBE devices
- AnchorFast 9799Hollister Incorporated · DEVICE, FIXATION, TRACHEAL TUBE1 recall1 death49 reports
- AnchorFast Guard 9800Hollister Incorporated · DEVICE, FIXATION, TRACHEAL TUBE1 recall4 reports
- AnchorFast Guard Select 9700Hollister Incorporated · DEVICE, FIXATION, TRACHEAL TUBE1 recall1 report
- Medline DYNJMBRC2Medline Industries, Inc. · DEVICE, FIXATION, TRACHEAL TUBE1 recall1 report
- Medline DYNJMBRC4BMedline Industries, Inc. · DEVICE, FIXATION, TRACHEAL TUBE1 recall0 reports
- AnchorFast 9799Hollister Incorporated · DEVICE, FIXATION, TRACHEAL TUBE21 reports
- HUDSON RCI IPN914324Teleflex Incorporated · DEVICE, FIXATION, TRACHEAL TUBE6 reports
- Dale Endotracheal Tube Holder 270Dale Medical Products, Inc. · Device, Fixation, Tracheal Tube6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page