VanishPoint® 24331Recalls & reported problems
Retractable Technologies, Inc. · 21G x 3/4" w/12" tubing VanishPoint® Blood Collection Set, Without Luer Adapter · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613703243300), received between May 22, 2019 and May 23, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Backflow” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Backflow1
- Difficult or Delayed Activation1
- Loose or Intermittent Connection1
- Therapeutic or Diagnostic Output Failure1
- Unintended Ejection1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2019Therapeutic or Diagnostic Output Failure
I WAS STICKING A PATIENT WITH A VANISHPOINT 21 G. I CLICKED THE BUTTON AND BLOOD SHOT OUT THE SIDE ALL OVER MY HAND THAT WAS GLOVED. I HAD TO CHANGE GLOVES BEFORE DOING ANYTHING ELSE WITH PATIENT. MANUFACTURER RESPONSE FOR TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION, VANISHPOINT (PER SITE REPORTER). MFR SENT CLINICAL REP ALONG WITH TERRITORY MANAGER, WILL BE SENDING TRAINERS NEXT WEEK.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8635470 ↗MalfunctionMay 22, 2019Unintended Ejection
AFTER INITIATION OF RETRACTION OF SHARP INTO SHARP CHAMBER, BLOOD BACK FLOWED AND LEAKED OUT OF THE CHAMBER. WE HAVE HAD 4 OTHER REPORTS OF THIS ALONG WITH OTHER ISSUES WITH THIS PARTICULAR DEVICE. OTHER ISSUES IS THAT WHEN THE BUTTON IS DEPRESSED TO RETRACT THE SHARP, IT HAS LAUNCHED THE SHARP OUT INTO THE AIR.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8631071 ↗MalfunctionMay 22, 2019Backflow; Loose or Intermittent Connection; Difficult or Delayed Activation
THE VANISHPOINT BLOOD COLLECTION SET, 21 GA X 3/4" X 12" REF# 24331 THE HOUSING CAN COME APART EASILY IF DROPPED OR WITH THE SIMPLE USE OF A FINGERNAIL TO SEPARATE. IT IS NOT SEALED. ADDITIONALLY, THE BUTTON THAT MUST BE DEPRESSED IN ORDER TO RETRACT THE SHARP IS EXTREMELY DIFFICULT TO DEPRESS AND REQUIRES MANEUVERING IN ORDER TO ACHIEVE. THE CLINICIANS HAVE REPORTED THAT THE ITEM HAS ALSO HAD BACKFLOW OF BLOOD INTO THE AREA WHERE THERE IS POTENTIAL TO SEPARATE ALONG WITH THE SHARP BECOMING A PROJECTILE UPON DEPRESSING THE RETRACTION LEVER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8631067 ↗
Other Tubes, Vials, Systems, Serum Separators, Blood Collection devices
- BD Vacutainer® SST™ 367986Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls2,711 reports
- BD Vacutainer® Buffered Sodium Citrate (9NC) Blood Collection Tubes 363083Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall1,744 reports
- BD Vacutainer® SST™ II Advance 367955Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1,478 reports
- BD Vacutainer® Eclipse™ Blood Collection Needle 368607Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,220 reports
- BD Vacutainer® SST™ 367983Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls1,075 reports
- BD Vacutainer® K2 EDTA (K2E) 5.4mg Blood Collection Tubes 367856Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,010 reports
- BD Vacutainer® PST™ Gel and Lithium HeparinN 56 Units 367960Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall747 reports
- BD Vacutainer® SST™ II Advance 367957Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection767 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.