ROHO LOW PROFILE Single Compartment Cushion 1R109LPCRecalls & reported problems
Roho, Inc. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613732005535), received between Nov 26, 2025 and Nov 26, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Gas/Air Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak1
- Material Integrity Problem1
- Naturally Worn1
Patient problems
- Pressure Sore/Ulcer1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 26, 2025Gas/Air Leak; Material Integrity Problem; Naturally Worn; Gas/Air Leak; Material Integrity Problem; Naturally Worn
USER CALLED TO REPORT HE HAD JUST RETURNED FROM VACATION AND HIS CUSHION DEVELOPED A LEAK WHILE ON VACATION. HE HAS PATCHED THE CUSHION SEVERAL TIMES BUT WAS UNABLE TO GET THE PATCH TO STOP THE LEAK. HE STATED THAT HE HAS DEVELOPED A PRESSURE SORE ON HIS BUTTOCKS DUE TO THE DEFLATED CUSHION. THE CUSHION WAS RETURNED TO ROHO FOR EVALUATION ON 11/6/2025. EVALUATION OF THE CUSHION SHOWED ONE OF THE PATCHES WAS LEAKING AIR. THE CELL WITH THE LEAKING PATCH WAS DISSECTED AND IT WAS DETERMINED THERE WAS A TEAR UNDER THE PATCH WHICH WAS LIKELY TOO LARGE FOR THE PATCH TO STOP THE LEAK. THERE WAS ANOTHER PATCH ON THE SAME CELL WHICH ALSO HAD A TEAR UNDERNEATH IT BUT THIS PATCH WAS PROPERLY SEALED. THE INSIDE OF THE LEAKING CELL SHOWED SIGNS OF FATIGUE INCLUDING A CREASE. A REPLACEMENT CUSHION WAS PROVIDED TO THE USER.
Patient problems: Pressure Sore/Ulcer
MAUDE report 23657766 ↗
Other Cushion, Flotation devices
- MEDLINE MSCBUB1818Medline Industries, Inc. · CUSHION, FLOTATION10 reports
- MEDLINE MSCBUB1818VLMedline Industries, Inc. · CUSHION, FLOTATION4 reports
- ROHO QUADTRO SELECT HIGH PROFILE Cushion QS1011CRoho, Inc. · Cushion, Flotation3 reports
- Medline MSCBUB2822VLMedline Industries, Inc. · CUSHION, FLOTATION3 reports
- ROHO QUADTRO SELECT HIGH PROFILE Cushion QS1010CRoho, Inc. · Cushion, Flotation3 reports
- ROHO QUADTRO SELECT HIGH PROFILE Cushion QS910CRoho, Inc. · Cushion, Flotation2 reports
- ROHO QUADTRO SELECT HIGH PROFILE Cushion QS89CRoho, Inc. · Cushion, Flotation2 reports
- MEDLINE MSCBUB2822Medline Industries, Inc. · CUSHION, FLOTATION2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page