Skip to content

VERTEX® Reconstruction System 7752526Recalls & reported problems

Medtronic Sofamor Danek, Inc. · CROSSLINK 7752526 MEDIUM MAS · Class 2 · Implant

All models of VERTEX Reconstruction System (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994245472), received between Sep 7, 2017 and Oct 18, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2024Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. PRE-OPERATIVE DIAGNOSIS FOR THIS PROCEDURE WAS CERVICAL MYELOPATHIE. IT WAS REPORTED THAT THE DEFECTIVE CROSSLINKS HAS BEEN CONNECTED, THE SETSCREW WAS BROKEN. LEVELS IMPLANTED WERE C3-T1. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT A NEW CROSSLINK AND SETSCREW WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20479046 ↗
  • MalfunctionSep 7, 2017Break; Break

    PRE-OP DIAGNOSIS: CANAL STENOSIS LEVELS IMPLANTED: C3-T1 IT WAS REPORTED THAT INTRA -OP, THE HEAD TO HEAD CROSSLINK BROKE DURING FINAL TIGHTENING. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO FRAGMENTS REMAINED INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS THE RESULT OF THE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6847553 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →