CD HORIZON® Spinal System 7570900Recalls & reported problems
Medtronic Sofamor Danek, Inc. · INSERTER 7570900 REDUX ROD · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00613994302748), received between Oct 12, 2017 and Apr 9, 2026 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Material Twisted/Bent1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 9, 2026Mechanical Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING VOYAGER CASE MIS LUMBAR ARTHRODESIS PROCEDURE. IT WAS REPORTED THAT THE INSERTER WAS NOT ABLE TO GRASP THE ROD AS EXPECTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24834876 ↗InjuryOct 11, 2024Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A ROD INSERTER USED ON PATIENT HAVING CASE OF THE L4-L5 MINIMALLY INVASIVE TRANSFORAMINAL LUMBAR INTERBODY FUSION (MIS TLIF) WITH SEXTANT II THERAPY FOR L4-L5 CENTRAL STENOSIS. IT WAS REPORTED THAT IN POST OPERATIVE X-RAY SURGEON FOUND A FOREIGN BODY ON THE RIGHT SIDE NEAR THE ZIG INSERTION SITE. IT WAS UNDERSTOOD THAT IT WAS PART OF THE ROD GRIPPER OF THE ZIG. THIS WAS NOT SEEN IN PRE-OP IMAGES AS THE BROKEN PIECE WAS FAR FROM THE IMPLANTED AREA. PATIENT HAD THE REVISION SURGERY NEXT DAY FOR REMOVAL OF THE BROKEN FLANGE OF THE ZIG. THERE WERE NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20426209 ↗MalfunctionOct 28, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT LUMBAR ARTHRODESIS OF 5 LEVELS. INTRA-OP, WHEN INSERTING THE FORCEPS, TIP OF THE ROD INSERTER BROKE. THE PRODUCT CAME IN CONTACT WITH THE PATIENT; BUT NO FRAGMENT OF THE BROKEN INSERTER REMAINED INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9246175 ↗MalfunctionSep 10, 2019Break; Material Twisted/Bent; Break; Material Twisted/Bent
SURGEON BENT THE ITEM WHEN REMOVING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8980122 ↗InjuryOct 12, 2017Break; Break
PRE-OP DIAGNOSIS: NERVE COMPRESSION, LEG PAIN PROCEDURE: ROD INSERTED DURING SPONDYLODESIS. THE INSERTER WAS 'BEENACHTIG' THE SKIN AND FASCIA. LEVELS IMPLANTED: L3-L4 IT WAS REPORTED THAT INTRA-OP, THE TIP OF THE ROD INSERTER WAS REPORTED BROKEN. THE TWO PINS OFF THE INNER SLEEVE FROM THE INSERTER WERE FIXED INTO THE TULIP OF THE SCREW. WHEN THE DOCTOR TRIED TO LOOSE THE INSERTER FROM THE ROD THE TIP BROKE OFF. ONE OF THE TWO PINS GOT OUT OF THE PATIENT, THE OTHER ONE WAS STUCK BETWEEN THE ROD AND THE SCREW. THE DOCTOR ACCEPTED THIS INSTEAD OF OPENING THE SKIN AND FASCIA. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. THERE WERE FRAGMENTS OF THE INSTRUMENT REMAINING IN THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6943511 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page