Skip to content

Midas Rex® AVA10Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT AVA10 10CM VARIABLE ANGLED · Class 2

All models of Midas Rex (154)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00613994424181), received between Aug 13, 2020 and Jul 2, 2026 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (4 reports).

9 other recalls cover devices cleared under the same FDA submission (K020069); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device4
  • No Apparent Adverse Event3
  • Detachment of Device or Device Component2
  • Difficult to Remove1
  • Electrical /Electronic Property Problem1
  • Fluid/Blood Leak1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Chemical Exposure1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 2, 2026Detachment of Device or Device Component

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE WAS NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25686726 ↗
  • MalfunctionJul 1, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THERE WERE NO ALLEGED ISSUE WITH THE DEVICE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25674875 ↗
  • MalfunctionJun 30, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE DEVICE HAS GENERAL DISSATISFACTION. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25662767 ↗
  • MalfunctionJun 5, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT WAS UNABLE TO LOCK TOOL OR BLADE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25405189 ↗
  • MalfunctionMar 18, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24625808 ↗
  • MalfunctionSep 11, 2024Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20191448 ↗
  • MalfunctionAug 28, 2024Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN. ON FOLLOW-UP IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20085589 ↗
  • MalfunctionAug 25, 2024Electrical /Electronic Property Problem; Detachment of Device or Device Component; Electrical /Electronic Property Problem; Detachment of Device or Device Component

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF SEIZED/NOT RUNNING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. REPAIR ESCALATED TO A PRODUCT EVENT ON EVALUATION DUE TO THE BEARING DETACHMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20059232 ↗
  • MalfunctionAug 20, 2024Overheating of Device; Difficult to Remove; Overheating of Device; Difficult to Remove

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF REMOVAL DIFFICULTY AND OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20008873 ↗
  • MalfunctionJun 17, 2021Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF HEATING UP WHEN IN USE. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN AND DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12016204 ↗
  • InjuryAug 13, 2020Fluid/Blood Leak; Fluid/Blood Leak

    ON FOLLOW-UP, IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE AND THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY. IT WAS ALSO REPORTED THAT THERE WAS NO FURTHER INFORMATION REGARDING THIS EVENT.

    Patient problems: Chemical Exposure

    MAUDE report 10402631 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K020069. One submission often covers many models, so check the product and lot details in each record.

  • Z-1602-2016 · Apr 8, 2016 · Terminated
    Medtronic Midas Rex Dissecting Tool Legend 10 cm 4mm BA DIAM. Intended for use in neurosurgery.
  • Z-0321-2016 · Oct 21, 2015 · Terminated
    Tool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
  • Z-0546-2010 · Nov 30, 2009 · Terminated
    Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotom
  • Z-2156-2008 · May 30, 2008 · Terminated
    Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General S
  • Z-1574-05 · Sep 1, 2005 · Terminated
    Medtronic Midas Rex Classic, GS and Legend dissecting tools manufactured by Medtronic Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, tX 76137 USA, Rx only, S
  • Z-1067-05 · Jun 29, 2005 · Terminated
    Medtronic MIDAS REX Legend High Speed Pneumatic System, Gold Touch Motor. Manufactured by Medtronic, Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, TX 76137
  • Z-0278-05 · Nov 11, 2004 · Terminated
    Medtronic Midas Rex Legend High Speed Pneumatic System
  • Z-0213-05 · Nov 3, 2004 · Terminated
    Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
  • Z-0419-03 · Nov 11, 2002 · Terminated
    Legend Lubricant System (Legend Gold, Gold Touch and Platinum motors and Legend Lubricant/Diffuser cartridge)

Other MOTOR, DRILL, PNEUMATIC devices

All MOTOR, DRILL, PNEUMATIC devices →

Sources and updates