Bio-Console® E560BCS1Recalls & reported problems

Medtronic, Inc. · CONSOLE E560BCS1 EVAL CONTROLLER · Class 2

All models of Bio-Console (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613994450456), received between May 19, 2023 and Aug 28, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Power Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Power Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2025Power Problem; Power Problem

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A BIO-CONSOLE CONTROLLER INSTRUMENT, IT WAS REPORTED THAT THERE WAS NO AC POWER. THE INSTRUMENT WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22913832 ↗
  • MalfunctionJun 21, 2023Power Problem; Power Problem

    DURING PREVENTIVE MAINTENANCE OF THIS BIO-CONSOLE INSTRUMENT THE SERVICE TECHNICIAN OBSERVED THAT THE FRONT PANEL CHARGE INDICATOR LIGHT WAS FLICKERING AT A HIGH RATE. THIS WAS DETECTED DURING SERVICE SO THERE WAS NO ADVERSE PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17171894 ↗
  • MalfunctionMay 19, 2023Power Problem; Power Problem

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THIS BIO-CONSOLE INSTRUMENT THERE WAS AN AC POWER ISSUE. USE OF INSTRUMENT WAS CONTINUED AND THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION: THE HAND CRANK WAS NOT USED TO MAINTAIN FLOW AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16960322 ↗

Other Control, pump speed, cardiopulmonary bypass devices