CP5 Drive UnitRecalls & reported problems

LivaNova Deutschland GmbH · Class 2

Quick summary

This LivaNova CP5 Drive Unit is the pump controller used to run cardiopulmonary bypass and control pump speed during surgery. Most of the 32 reports describe the pump stopping or failing to deliver steady flow, sometimes with error messages or alarms, and staff switching to another pump. There are2 reports of patient injury and no deaths. One recall, started in October 2017 and closed, concerned the automatic line closure after air bubble detection taking longer than intended.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 04033817900948), received between Dec 2, 2016 and Apr 17, 2026 — classified as 2 injuries, 30 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (18 reports).

1 FDA recall record name this device's identifier.

3 other recalls cover devices cleared under the same FDA submission (K112225); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 30, 2017Terminated

    Possibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take longer than is intended by design in the CP5 system.

    Z-0401-2018 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped18
  • Pumping Problem6
  • Infusion or Flow Problem3
  • No Flow2
  • Appropriate Device Problem Term/Code Not Available1
  • Decreased Pump Speed1
  • Device Displays Incorrect Message1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • No Consequences Or Impact To Patient4
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2026Pumping Stopped

    LIVANOVA RECEIVED A REPORT ABOUT A CP5 SYSTEM DISPLAYING THE ERROR MESSAGE "MOTOR CONTROL FAILURE". THE EVENT OCCURRED DURING PRIMING. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24911192 ↗
  • MalfunctionMar 11, 2026Pumping Problem

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT THE CENTRIFUGAL PUMP CP5, USED AS ARTERIAL PUMP, PROVIDED UNSTABLE FLOW DURING PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24573777 ↗
  • MalfunctionJun 12, 2025Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, JUST BEFORE THE PROCEDURE STARTED, THE CP5 PUMP WAS STOPPED. THIS WAS USED AS A SUCTION. THE PERFUSIONIST WAS THEN ABLE TO QUICKLY SWITCH TO ANOTHER PUMP THAT WAS FREE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22195034 ↗
  • MalfunctionMay 26, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT OF AN ISSUE WITH CP5 DRIVER UNIT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22084826 ↗
  • MalfunctionNov 8, 2024Pumping Problem; Pumping Problem

    LIVANOVA RECEIVED A REPORT THAT A YELLOW ICON SHOWED UP WHEN CUSTOMER TRIED TO FORWARD THE FLOW, THE CP5 HAD DROPPED THE FLOW TO 0 AND THERE WAS NO FORWARD FLOW. PERFUSIONIST CANNOT RECALL THE INFORMATION ASSOCIATED TO THE DISPLAYED ALARM (YELLOW TRIANGLE ALARM ON CP5 PANEL). THROUGH FOLLOW UP, THE PERFUSIONIST INFORMED LIVANOVA THE PUMP WOULD NOT HAVE FORWARD FLOW FOR ABOUT 5-7 MINUTES. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20637895 ↗
  • MalfunctionOct 31, 2024Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20578109 ↗
  • MalfunctionOct 30, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CP5 GAVE AN ERROR MESSAGE AND PUMP DID NOT RUN. THE ISSUE OCCURRED AT SETUP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20565932 ↗
  • MalfunctionOct 10, 2024Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20415955 ↗
  • MalfunctionAug 5, 2024Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19906206 ↗
  • MalfunctionJun 14, 2024Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19542361 ↗
  • MalfunctionJun 14, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT DURING PRIMING A CENTRIFUGAL PUMP 5 (CP5) TURNED OFF AND THEN BACK ON BY ITSELF, AT LEAST TWO TIMES, AND AN ERROR MESSAGE ON S5 CONSOLE CONTROL PANEL APPEARED INDICATING A TEMPORARY INTERNAL FAILURE OF THE CENTRIFUGAL PUMP. THE CUSTOMER SUBSEQUENTLY TURNED THE ENTIRE S5 CONSOLE OFF AND ON AGAIN WITHOUT CREATING ANY OTHER PROBLEMS DURING THE SURGERY. IT WAS CLAIMED BY THE CUSTOMER THAT ALSO PRESSURE MODULE WAS DEFECTIVE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19541733 ↗
  • MalfunctionMay 22, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CP5 DRIVE UNIT STOPPED RUNNING; THERE WAS NO FLOW AND RPM¿S WERE SHOWING ON THE CP5 DISPLAY. THEN IT STARTED RUNNING AGAIN BY ITSELF. THE ISSUE OCCURRED PRIOR TO GOING ON BYPASS. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19372014 ↗
  • MalfunctionMar 29, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CENTRIFUGAL PUMP 5 (CP5) STOPPED INVOLUNTARY DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19002290 ↗
  • MalfunctionMar 21, 2024Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18950066 ↗
  • MalfunctionJan 3, 2024Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP GAVE MOTOR CONTROL FAILURE ERROR MESSAGE.THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18434375 ↗
  • InjuryNov 8, 2023Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT THAT, DURING A CARDIOPULMONARY BYPASS, THE PUMP STOPPED WORKING. THERE WAS NO CIRCULATION UNTIL A NEW BYPASS MACHINE WAS HOOKED UP AND STARTED. ACCORIND TO INFORMATION THIS TOOK APPROXIMATELY 8-10 MINUTES AND REQUIRED MEDICATIONS, BLOOD PRODUCTS AND CARDIAC MASSAGE. STAFF FROM PERFUSION IDENTIFIED THE TUBING AS A POSSIBLE CAUSE OF MALFUNCTION. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18092429 ↗
  • InjuryNov 8, 2023Pumping Stopped; Pumping Problem; Pumping Stopped; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18092413 ↗
  • MalfunctionOct 12, 2023Infusion or Flow Problem; Infusion or Flow Problem

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT AN UNKNOWN EVENT, OCCURRED AT 10:57, TRIGGERED A NEGATIVE PRESSURE ALARM ON A CENTRIFUGAL PUMP 5 (CP5); THE MAXIMUM AMOUNT OF FLOW ACHIEVED WAS 1.74 LPM WITH MAXED OUT RPMS OF 3500. THE ISSUE OCCURRED DURING PROCEDURE, DURING CLAMPING. AFTER TWO (2) MINUTES OF LOW FLOW, AND CUSTOMER SWITCHED TO HAND CRANKING TO ALLOW MORE TIME TO TROUBLESHOOT; NO ALERTS OR ALARMS WAS DISPLAYED. THEN, THE INVOLVED DRIVE UNIT WAS REPLACED WITH A ONE BELONGING TO A NEW PUMP BY THE CUSTOMER. FULL FLOW WAS RE-ESTABLISHED. THE TOTAL TIME OF LOW FLOW/HAND CRANKING WAS APPROXIMATELY SEVEN (7) MINUTES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17921247 ↗
  • MalfunctionSep 21, 2023No Flow; No Flow

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17787008 ↗
  • MalfunctionJun 29, 2023Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17224712 ↗
  • MalfunctionJun 16, 2023Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17141511 ↗
  • MalfunctionJun 5, 2023Infusion or Flow Problem; Infusion or Flow Problem

    SEE INTIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17058512 ↗
  • MalfunctionApr 13, 2023Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16731695 ↗
  • MalfunctionOct 31, 2022No Flow; No Flow

    LIVANOVA RECEIVED REPORT THAT ABOUT AFTER 20 MINUTES FROM PROCEDURE STARTED, THERE WAS NO FLOW DESPITE RPM OF THE CENTIFUGAL PUMP 5 (CP5) BEING 3800. FLOW WAS DOWN AND THERE WAS NEGATIVE LINE PRESSURE. THE ORIGINAL THOUGHT WAS A PRESSURE EXCURSION WHICH COULD BE EXCLUDED BY USER BECAUSE AS THEY HAND CRANKED EVERYTHING FLOWED. THE VACUUM WAS REMOVED AND THE CANNULA WAS VERIFIED NOT TO BE OCCLUDED. THE USER TURNED THE CP5 OFF AND BACK ON WHILE CONTINUING TO HAND CRANK AND THE PROBLEM PERSISTED. THE USER TURNED THE DEVICE OFF AND BACK ON AGAIN AND THIS TIME EVERYTHING WENT BACK TO NORMAL AND HAND CRANK WAS STOPPED. THE PROCEDURE COULD BE COMPLETED WITH NO FURTHER ISSUES. REPORTEDLY, THERE WERE NO ALARMS OR ERROR MESSAGES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15697993 ↗
  • MalfunctionAug 9, 2022Decreased Pump Speed; Pumping Stopped; Decreased Pump Speed; Pumping Stopped

    LIVANOVA RECEIVED REPORT THAT WHEN TURNED ON AND RPMS INCREASED TO 2000 THE PUMP HEAD OF CENTRIFUGAL PUMP 5 (CP5) DID NOT TURN. THE RPM NUMBER STAYED YELLOW, THEN WENT TO ZERO WITH THE DIAL STILL AT 2000. AT THIS TIME, THE PUMP HEAD STARTED TO TURN VERY SLOWLY. THE RPMS WERE MANUALLY SET DOWN TO ZERO AND BACK UP VERY SLOWLY TO 2000. THE PUMP HEAD STILL ONLY TURNED SLOWLY. THE RPM NUMBER REMAINED YELLOW. I WATCHED FOR A FEW MINUTES AND WITHOUT ME TOUCHING ANYTHING THE PUMP HEAD SPEED INCREASED TO MATCH THE RPM NUMBER AND THEN RAN APPROPRIATELY UNTIL I TURNED IT OFF. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15189829 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112225. One submission often covers many models, so check the product and lot details in each record.

  • Z-0787-2022 · Dec 17, 2021 · Terminated
    Sorin Centrifugal Pump System, CP5, Drive Unit (Catalogue Item #60-01-04, UDI: 04033817900948) used in an HLM S5/C5 System configuration. A cardiopulmonary bypass speed control dev
  • Z-0857-2014 · Dec 4, 2013 · Terminated
    CP5 centrifugal pump system Product Usage: The Stockert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
  • Z-1329-2012 · Feb 23, 2012 · Terminated
    Pump control panel. The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.

Other Control, pump speed, cardiopulmonary bypass devices