Skip to content

Bio-Medicus® 96370-023Recalls & reported problems

Medtronic, Inc. · CANNULA 96370-023 BIO-MEDICUS 23FR EA · Class 2 · Implant

All models of Bio-Medicus (59)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994504777), received between Jul 21, 2025 and Jul 21, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K924642); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1
  • Misassembled During Installation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2025Mechanical Problem; Misassembled During Installation; Mechanical Problem; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS ONE-PIECE FEMORAL VENOUS CANNULA, IT WAS REPORTED THAT THE ECMO (EXT RACORPOREAL MEMBRANE OXYGENATION) RETRIEVAL TEAM PICKED UP THE PATIENT FROM A REGIONAL SERVICE. THE RETRIEVAL TEAM CANNULATED THE PATIENT TO COMMENCE ECMO. HOWEVER, THEY DISCOVERED THAT THEY COULD NOT CONNECT THE CANNULA TO THE ECMO CIRCUIT DUE TO A HALF-INCH CONNECTOR, RATHER THAN THE STANDARD 3/8 CONNECTOR. THE PATIENT HAD TO BE RE-CANNULATED WITH A DIFFERENT CANNULA THAT HAD A 3/8 CONNECTOR. THE PATIENT WAS SUCCESSFULLY ATTACHED TO THE ECMO CIRCUIT. THE PATIENT WAS A CRITICALLY UNWELL PATIENT. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22553503 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K924642. One submission often covers many models, so check the product and lot details in each record.

  • Z-1928-2008 · May 9, 2008 · Terminated
    Medtronic Bio-Medicus Femoral Cannula/Introducer, for Cardiopulmonary Bypass. Catalog Numbers: CB96345-023, CB96345-023, and CB96345-029.

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page