NA HY11W65R1Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK HY11W65R1 SPECTRUM CPB · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000704841), received between Oct 17, 2024 and Jan 20, 2026 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

2 FDA recall records name this device's identifier, 2 still open.

25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 12, 2022Open, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1027-2023 · Root cause: Process change control · Recall details · FDA record ↗

  2. Dec 12, 2022Open, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1028-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Delivered as Unsterile Product3
  • Component Missing1
  • Crack1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Hemorrhage/Blood Loss/Bleeding3

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026Delivered as Unsterile Product

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM PACK, IT WAS REPORTED THAT THE OUTER PACKAGING, INNER PACKAGING AND STERILE BARRIER WERE DAMAGED. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24116889 ↗
  • MalfunctionDec 12, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM PACK, IT WAS REPORTED THAT THE SASH WRAP ON TABLE LINES APPEARED TO HAVE BEEN TORN BY THE FUSION OXYGENATOR DURING SHIPPING. THE CUSTOMER STATED THAT THERE WAS NO DAMAGE TO THE PACKAGING AND NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER WERE DAMAGED. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TOP LID (STERILE TOP) WAS INTACT WHEN THIS WAS FIRST NOTED. THE DEVICE WAS STILL WITHIN THE STERILE SEAL HOWEVER IT MUST HAVE BEEN MOVING AROUND DURING SHIPPING BECAUSE THE STERILE SASH WRAP FOR THE AV LOOP WAS RIPPED OPEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23789737 ↗
  • MalfunctionDec 12, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT THERE WAS A TEAR IN THE ARTERIAL VENTRICULAR LOOP BLUE STERILE WRAP. THERE WAS NO DAMAGE TO THE PACKAGING. NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER WERE NOT DAMAGED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23789723 ↗
  • MalfunctionApr 23, 2025Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THERE WAS A MINIMAL BLOOD LOSS AS A RESULT OF A BLOOD LEAK, AND THERE WAS NO UNPLANNED TRANSFUSION AS A RESULT OF THIS LEAK. THE LEAK OCCURRED AT LINE 4, VRV VALVE. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21889513 ↗
  • InjuryMar 26, 2025Crack; Fluid/Blood Leak; Material Puncture/Hole; Crack; Fluid/Blood Leak; Material Puncture/Hole

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM PACK, IT WAS REPORTED THAT THE ONE-WAY VALVE WAS COMPLETELY BROKEN IN HALF IN THE PACK SPEC LINE 15 . THE ONE-WAY VALVE IN THE AORTIC VENT BLEW, CAUSING A SIGNIFICANT MESS. THE FLOW THROUGH THE LINE WAS APPROXIMATELY 450 ML/MIN, AND PRESSURE WAS NOT TRANSDUCED THROUGH IT. THE DEVICE WAS REPLACED TO COMPLETE THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21695865 ↗
  • MalfunctionJan 27, 2025Fluid/Blood Leak; Component Missing; Fluid/Blood Leak; Component Missing

    IT WAS REPORTED THAT PRIOR TO USE OF A CUSTOM TUBING PACK A LEAK OCCURRED AT THE QUICK CONNECT ON THE AV LOOP. THE QUICK CONNECTOR THAT NORMALLY HAS A SPRING IN IT DID NOT HAVE A SPRING THIS TIME SO IT WAS LEAKING AROUND THE SIDES. AFTER PRIMING THE CIRCUIT AND HANDING LINES OFF TO THE SURGEON IT WAS OBSERVED THAT THE SPRING IN THE QUICK CONNECT FOR THE AV LOOP WAS NOT THERE AND CRYSTALLOID WAS LEAKING AROUND IT. THERE WAS NO PATIENT BLOOD LOSS AS A RESULT OF THIS LEAK. AN UNPLANNED BLOOD TRANSFUSION WAS NOT REQUIRED AS A RESULT OF THE LEAK. THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS. THE DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21238949 ↗
  • MalfunctionNov 11, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE AV LOOP WAS LEAKING. THE CUSTOMER STATED THEY HAD A PACK WITH A LEAKING PORT INSIDE THE STERILE WRAPPER. THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20651749 ↗
  • MalfunctionOct 17, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT IT HAD A LEAKY ONE WAY VALVE. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20468711 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.

  • Z-1872-2021 · Apr 29, 2021 · Terminated
    Medtronic HY8M63R4, Custom Pack HY8M63R4 Adult ECC, Custom Perfusion System, Sterile, Rx.
  • Z-1871-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10Z48R2, Ctr., Custom Pack HY10Z48R2 Adult ECC 1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1888-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10R55R5, Custom Pack CB10R55R5 3/8X3/8 Loop, Custom Perfusion System, Sterile, Rx.
  • Z-1876-2021 · Apr 29, 2021 · Terminated
    Medtronic CB7J97R8, Custom Pack CB7J97R8 U14 E Sup, Custom Perfusion System, Sterile, Rx.
  • Z-1889-2021 · Apr 29, 2021 · Terminated
    Medtronic CB7J98R7, Custom Pack CB7J98R7, Custom Pack CB7J98R7 3/8 Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1874-2021 · Apr 29, 2021 · Terminated
    Medtronic BB11J14R2, Custom Pack BB11J14R2 Bridge, Custom Perfusion System, Sterile, Rx.
  • Z-1890-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10L28R3, Custom Pack HY10L28R3 1/4 Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1884-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10W65R1, Custom Pack CB10W65R1 CH Access1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1887-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10R79R6, Custom Pack CB10R79R6 1/4X1/4 Loop, Custom Perfusion System, Sterile, Rx.
  • Z-1880-2021 · Apr 29, 2021 · Terminated
    Medtronic HY8R62R13,Custom Pack HY8R62R13 E Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1878-2021 · Apr 29, 2021 · Terminated
    Medtronic CB1H18R20, Custom Pack CB1H18R20 Accesory, Custom Perfusion System, Sterile, Rx.
  • Z-1894-2021 · Apr 29, 2021 · Terminated
    Medtronic BB9G78R8, Custom Pack BB9G78R8 ECC Supp, Custom Perfusion System, Sterile, Rx.
  • Z-1895-2021 · Apr 29, 2021 · Terminated
    Medtronic 1A46R11, Custom Pack 1A46R11 Neonatal, Custom Perfusion System, Sterile, Rx.
  • Z-1886-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10D90R2, Custom Pack HY10D90R2 3/8 ACC, Custom Perfusion System, Sterile, Rx.
  • Z-1892-2021 · Apr 29, 2021 · Terminated
    Medtronic BB8M63R5, Custom Pack BB8M63R5 Adult ECC, Custom Perfusion system, Sterile, Rx.

…and 10 more.

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices

All Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices →