CD HORIZON® Spinal System 5442138Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CROSSLINK 5442138 4.75X10 MLTSPN 38-42MM · Class 2 · Implant
All models of CD HORIZON Spinal System (341)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00613994588463), received between Dec 8, 2019 and Aug 28, 2021 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K091974); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break4
- Device Fell1
- Device Slipped1
- Fracture1
- Material Deformation1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2021Fracture; Fracture
INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA FIELD REPRESENTATIVE REGARDING PATIENT WITH PRE-OP DIAGNOSIS-AIS NEUROFIBROMA INVOLVED IN THE NAVIGATE LOWER SCOLIOSIS SURGERY. IT WAS REPORTED THAT INTRA-OPERATIVELY, CAP WAS FRACTURED. PRODUCT BROKE AND NO FRAGMENTS LEFT IN THE PATIENT'S BODY. PRODUCT WAS UTILIZED CORRECTLY ACCORDING TO THE DIRECTIONS GIVEN IN THE IFU/LABELING. THERE WAS NO DELAY IN OVERALL PROCEDURE TIME. THE LEVEL OF TREATMENT OR IMPLANTATION WAS VERY HIGH, THE SURGEON OFTEN USED THE CROSSLINK DURING THE OPERATION, AND THE USE OF THE CROSSLINK WAS VERY FAST. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. PATIENT IS ALIVE AND NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12381162 ↗MalfunctionMar 15, 2021Break; Device Slipped; Break; Device Slipped
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA FIELD REPRESENTATIVE REGARDING A PATIENT WHO UNDERGONE SPINAL THERAPY WITH PRE-OP DIAGNOSIS OF LUMBAR 1 CONE FRACTURE, SEVERE OSTEOPOROSIS, ANKYLOSING SPONDYLITIS. IT WAS REPORTED THAT DURING THORACOLUMBAR POSTERIOR DEVELOPMENTAL SURGERY WHEN THE PRODUCT WAS IMPLANTED, THE SCREW CAP BROKE JUST WITH LIGHT FORCE AND THE CROSSLINK SCREW CAP SLIPPED. FRAGMENT OF THE DEVICE REMAINING IN THE PATIENT WAS UNKNOWN. PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT WAS UNKNOWN. IT WAS UNKNOWN WHETHER ANY TREATMENT OR ADDITIONAL SURGERY PERFORMED AS A RESULT OF THIS EVENT OR NOT. PATIENT WAS ALIVE AND NO INJURY. PRODUCT UTILIZED CORRECTLY ACCORDING TO THE DIRECTIONS GIVEN IN THE IFU/LABELING. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11482187 ↗MalfunctionAug 26, 2020Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH AN INDICATION OF LUMBAR DISC HERNIATION IN NEED OF LUMBAR DISCECTOMY USED IN SPINAL THERAPY. IT WAS REPORTED THAT WHEN THE CROSS LINK WAS IMPLANTED, THE TAPERED PROTRUSION OF THE CROSS LINK BROKE AND FELL OFF, AND THE FIXATION OF THE CROSS LINK FAILED. THERE WAS A DELAY OF LESS THAN 60 MINS IN OVERALL PROCEDURE TIME REPORTED. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FRAGMENTS IN THE PATIENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10452818 ↗MalfunctionAug 26, 2020Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH AN INDICATION OF LUMBAR DISC HERNIATION IN NEED OF LUMBAR DISCECTOMY USED IN SPINAL THERAPY. IT WAS REPORTED THAT WHEN THE CROSS LINK WAS IMPLANTED, THE TAPERED PROTRUSION OF THE CROSS LINK BROKE AND FELL OFF, AND THE FIXATION OF THE CROSS LINK FAILED. THERE WAS A DELAY OF LESS THAN 60 MINS IN OVERALL PROCEDURE TIME REPORTED. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FRAGMENTS IN THE PATIENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10452807 ↗MalfunctionAug 24, 2020Break; Device Fell; Break; Device Fell
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA FIELD REPRESENTATIVE REGARDING PATIENT WITH SPINAL FRACTURE INVOLVED IN SPINAL REDUCTION AND INTERNAL FIXATION PROCEDURE. IT WAS REPORTED THAT DURING PROCEDURE, THE FIXATION CROSSLINK BROKE AND FELL OFF, THAT MADE UNABLE TO FIX. THERE IS NO FRAGMENT LEFT IN THE PATIENT. THERE IS NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED AS THE RESULT OF EVENT. THERE IS NO ADDITIONAL SURGERY PERFORMED AS A RESULT OF THIS EVENT. PATIENT IS ALIVE AND NO INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10441251 ↗MalfunctionDec 8, 2019Material Deformation; Material Deformation
IT WAS REPORTED THAT THE PATIENT WAS UNDERGOING A REVISION SURGERY AT T9-T10. INTRA-OP, WHEN PLACING THE CROSSLINK, SET SCREW OF THE CROSSLINK STRIPPED; SO, THE SURGEON REPLACED THE ENTIRE CROSSLINK WITH A NEW ONE. NO PATIENT COMPLICATIONS WERE REPORTED DUE TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9435077 ↗
FDA clearance and classification
Cleared through Special 510(k) K091974 on Sep 2, 2009 (decision: Substantially Equivalent). Submitted by Medtronic Sofamor Danek USA, Inc. as “Modification To Cd Horizon Spinal System”. The FDA review took 63 days. FDA record ↗
The FDA regulates product code KWP (appliance, fixation, spinal interlaminal) under 21 CFR 888.3050 as a Class II device. 21 CFR 888.3050 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K091974. One submission often covers many models, so check the product and lot details in each record.
- Z-2876-2011 · Jun 28, 2011 · TerminatedCD HORIZON (R) SOLERA (TM) FAS, REF 54410006035, pedicle screw, Rx only, Medtronic Sofamor Danek USA, 1800 Pyramid Place, Memphis, TN 38132 Intended to help provide immobilizatio
Common questions about the CD HORIZON® Spinal System 5442138
Has the CD HORIZON® Spinal System 5442138 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00613994588463), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the CD HORIZON® Spinal System 5442138?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Dec 8, 2019 and Aug 28, 2021. In those reports, the most common specific device problems are “Break” (4), “Device Fell” (1) and “Device Slipped” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CD HORIZON® Spinal System 5442138?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CD HORIZON® Spinal System 5442138 FDA approved?
It was cleared through Special 510(k) K091974 on Sep 2, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the CD HORIZON® Spinal System 5442138 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the CD HORIZON® Spinal System 5442138, and is it still sold?
The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand CD HORIZON® Spinal System, model 5442138). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL433 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page