CD HORIZON® Spinal System 54840007545Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 54840007545 MAS 7.5X45 CC · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613994589293), received between Jan 23, 2017 and Apr 3, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K091974); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
Patient problems
- Pain2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 3, 2023Break; Break
INFORMATION WAS RECEIVED FROM A CONSUMER HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OPERATIVE DIAGNOSIS FOR BROKEN RIGHT SIDE S1 PEDICLE SCREW. PROCEDURE OR TECHNIQUE USED WAS O-ARM NAVIGATED TLIF PROCEDURE. LEVELS IMPLANTED S1. PATIENT HAD A MEDICAL HISTORY OF REGULAR ALCOHOL AND DRUG USE. IT WAS REPORTED THAT THE PATIENT FELL AND HEARD A POP AND FELT PAIN. WENT BACK TO SEE SURGEON WHO WITH IMAGING SAW THE BROKEN SCREW ON THE RIGHT SIDE. REOPERATION WAS PERFORMED TO REPLACE A BROKEN SCREW. AFTER FALL PATIENT SAW THE SURGEON WHO TOOK FILMS AND SAW THE SCREW IN S1 ON THE RIGHT SIDE WAS BROKEN. WHEN SURGEON OPENED THE PATIENT AND REMOVED THE SET SCREWS AND RODS, BOTH S1 SCREWS WERE BROKEN. SURGEON USED 8.0 TREPHINE AND 8.0 CONICAL EXTRACTION BOLT TO REMOVE BOTH BROKEN PIECED THAT REMAINED IN THE S1 BODY AFTER THE BROKEN TOP PORTION WAS REMOVED. PATIENT HAD PAIN AFTERWARDS.
Patient problems: Pain
MAUDE report 16661532 ↗InjuryApr 3, 2023Break; Break
INFORMATION WAS RECEIVED FROM A CONSUMER HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OPERATIVE DIAGNOSIS FOR BROKEN RIGHT SIDE S1 PEDICLE SCREW. PROCEDURE OR TECHNIQUE USED WAS O-ARM NAVIGATED TLIF PROCEDURE. LEVELS IMPLANTED S1. PATIENT HAD A MEDICAL HISTORY OF REGULAR ALCOHOL AND DRUG USE. IT WAS REPORTED THAT THE PATIENT FELL AND HEARD A POP AND FELT PAIN. WENT BACK TO SEE SURGEON WHO WITH IMAGING SAW THE BROKEN SCREW ON THE RIGHT SIDE. REOPERATION WAS PERFORMED TO REPLACE A BROKEN SCREW. AFTER FALL PATIENT SAW THE SURGEON WHO TOOK FILMS AND SAW THE SCREW IN S1 ON THE RIGHT SIDE WAS BROKEN. WHEN SURGEON OPENED THE PATIENT AND REMOVED THE SET SCREWS AND RODS, BOTH S1 SCREWS WERE BROKEN. SURGEON USED 8.0 TREPHINE AND 8.0 CONICAL EXTRACTION BOLT TO REMOVE BOTH BROKEN PIECED THAT REMAINED IN THE S1 BODY AFTER THE BROKEN TOP PORTION WAS REMOVED. PATIENT HAD PAIN AFTERWARDS.
Patient problems: Pain
MAUDE report 16661515 ↗MalfunctionJan 23, 2017Break; Break
PRE-OPERATIVE DIAGNOSIS: ADJACENT LEVEL HERNIATED DISC LEVELS IMPLANTED: L5-S1 IT WAS REPORTED THAT DURING SURGERY, WHEN SURGEON WENT TO REMOVE SET SCREWS AND RODS, ONE OF THE SCREW HEAD BROKE OFF. ADJACENT LEVEL SURGERY WAS PERFORMED WITH ADDITIONAL HARDWARE. PART OF BROKEN PEDICLE SCREW STILL REMAINS IN PATIENT AT S1. NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6269215 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K091974. One submission often covers many models, so check the product and lot details in each record.
- Z-2876-2011 · Jun 28, 2011 · TerminatedCD HORIZON (R) SOLERA (TM) FAS, REF 54410006035, pedicle screw, Rx only, Medtronic Sofamor Danek USA, 1800 Pyramid Place, Memphis, TN 38132 Intended to help provide immobilizatio
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page