ENDURANT® ENLW1624C93ERecalls & reported problems
Medtronic, Inc. · STENT GRAFT ENLW1624C93E ENDURANT LIMB · Class 3 · Implant
All models of ENDURANT (46)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00613994761897), received between May 19, 2017 and Feb 19, 2019 — classified as 5 injuries.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Leak/Splash2
- Migration or Expulsion of Device1
Patient problems
- Injury2
- Aneurysm1
- Hypoxia1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryFeb 19, 2019Leak/Splash; Leak/Splash
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED ON AN UNKNOWN DATE, CT A TYPE IB ENDOLEAK WAS OBSERVED FROM THE RIGHT CIA WHICH HAD LOST SEAL AND HAS NOW INCREASED IN SIZE. APPROXIMATELY SEVEN YEARS POST THE INDEX PROCEDURE, THE PHYSICIAN COILED THE RIGHT HYPOGASTRIC AND PLACED AN EXTENSION ENDURANT LIMB. ADDITIONALLY, THE PHYSICIAN ELECTED TO IMPLANT AN AORTIC CUFF DUE TO CONCERN FOR A TYPE IA ENDOELAK. PER THE PHYSICIAN THE CAUSE OF THE EVENT WAS ANATOMY RELATED DUE TO CONTINUOUS EXPANSION OF THE RIGHT CIA. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: Injury
MAUDE report 8349218 ↗InjuryMay 17, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN UNKNOWN ABDOMINAL AORTIC DISEASE. IT WAS REPORTED APPROXIMATELY 7 YEARS POST THE INDEX PROCEDURE A CT WAS PERFORMED WHICH SHOWED THAT THE RIGHT ILIAC LIMB HAD PULLED OUT OF THE ILIAC AND WAS NOW IN THE AORTA. INTERVENTION WAS PERFORMED. DURING THE INTERVENTION, THE PHYSICIAN SUCCESSFULLY GOT A WIRE THROUGH THE ENDURANT LIMB BUT IT WAS DECIDED THAT THE LIMB WAS AT TOO BAD OF AN ANGLE AND THE WIRE COULD NOT STRAIGHTEN. IT WAS DECIDED TO PLACE AN AUI DEVICE ON THE LEFT SIDE AND OCCLUDE THE RIGHT COMMON ILIAC ARTERY. A FINAL ANGIO WAS PERFORMED AND THERE WAS NO ENDOLEAK. AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT WAS ANATOMY RELATED WITH THE RIGHT COMMON ILIAC MEASURING 38MM AND THAT AT THE TIME OF IMPLANT, THEY MAY NOT HAVE IMPLANTED FAR ENOUGH INTO THE RIGHT COMMON ILIAC. DUE TO EXPANSION OF THE RIGHT COMMON ILIAC, THE ENDURANT LIMB NOW HAD NOTHING TO ATTACH TO. NO ADDITIONAL CLINICAL SEQUELA WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: Injury
MAUDE report 7521454 ↗InjuryMay 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
HELI-FX ENDOANCHORS WERE IMPLANTED IN AN ENDURANT STENT GRAFT DURING THE ENDOVASCULAR TREATMENT OF AN 61MM ABDOMINAL AORTIC ANEURYSM. THE AORTIC NECK LENGTH WAS OBSERVED TO BE 20MM IN LENGTH, WITH AN AORTIC DIAMETER OF 17.67MM AT 1MM, 16.8MM AT 10MM, 17.67 AT 15MM AND 18.39 AT 20MM BELOW THE LOWEST RENAL ARTERY. THE PROXIMAL AORTIC NECK ANGLE WAS 80 DEGREES. THREE ENDOANCHORS WERE IMPLANTED DUE TO A CONCERN FOR LATE FAILURE. IT WAS REPORTED THAT APPROXIMATELY 6 YEARS POST INDEX PROCEDURE, THE PATIENT EXPERIENCED PULMONARY ISSUES, SPECIFICALLY HYPOXEMIA. THE EVENT RELATIONSHIP WAS DETERMINED TO BE UNCERTAIN. THE PATIENT WAS GIVEN OXYGEN AT HOME AS TREATMENT. THE EVENT WAS RESOLVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED.
Patient problems: Hypoxia
MAUDE report 7498087 ↗InjuryNov 14, 2017Leak/Splash; Leak/Splash
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 55MM DIAMETER ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE PATIENT WAS ADMITTED TO THE HOSPITAL APPROXIMATELY 6 YEARS AND 2 MONTHS POST INDEX PROCEDURE WITH ABDOMINAL PAIN. A CT SCAN SHOWED THAT THE ANEURYSM HAS GROWN FROM 5.5 CM TO 8.6CM A PROXIMAL TYPE I ENDOLEAK AND A DISTAL TYPE I ENDOLEAK WAS ALSO NOTED ON THE RIGHT. THE LEFT APPEARED TO HAVE A SEAL. THE PHYSICIAN IMPLANTED A 36X36X49 ENDURANT CUFF PROXIMALLY AND A 16X24X93 ENDURANT RIGHT SIDE EXTENSION DISTALLY. THE PROXIMAL ENDOLEAK HAS RESOLVED WITH THE CUFF PLACEMENT. THE RIGHT SIDE EXTENSION WAS PLACED WITH A SMALL RESIDUAL ENDOLEAK THAT WAS BALLOONED MULTIPLE TIMES. THE ENDOLEAK IMPROVED BUT THE DECISION WAS MADE TO HAVE A FOLLOW UP CT SCAN AFTER DISCHARGE. A RETROGRADE ANGIOGRAM WAS DONE ON THE LEFT SIDE AND NO ENDOLEAK WAS VISUALIZED. IT WAS REPORTED THAT THE PHYSICIAN CONFIRMED THAT, POST RE-INTERVENTION, THE PATIENT WAS NO LONGER HAVING ABDOMINAL PAIN OR TENDERNESS. PER THE PHYSICIAN THE EVENT WAS ANATOMY RELATED (DISEASE PROGRESSION). NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT FINE.
Patient problems: Pain
MAUDE report 7028187 ↗InjuryMay 19, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE PATIENT RETURNED FOR FOLLOW-UP AND THE CT SHOWED THERE WAS DISEASE PROGRESSION OF THE ILIAC ARTERY ON THE RIGHT SIDE. PER THE PHYSICIAN THE CAUSE OF THE EVENT IS ANATOMY RELATED, DUE TO A 4CM ILIAC ANEURYSM BELOW THE EXISTING ENDURANT LIMB ON RIGHT SIDE OF PATIENT. THE DECISION WAS MADE TO COIL THE HYPOGASTRIC ARTERY ON THE RIGHT SIDE AND EXTEND WITH AN ENDURANT STENT GRAFT AND THIS SUCCESSFULLY RESOLVED THE EVENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: Aneurysm
MAUDE report 6579118 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.
- Z-0292-2022 · Oct 15, 2021 · Open, ClassifiedEndurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
- Z-0290-2022 · Oct 15, 2021 · Open, ClassifiedEndurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
- Z-0291-2022 · Oct 15, 2021 · Open, ClassifiedEndurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
- Z-2135-2021 · Jun 7, 2021 · Open, ClassifiedEndurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat i
- Z-2136-2021 · Jun 7, 2021 · Open, ClassifiedMedtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aort
- Z-1543-2017 · Feb 27, 2017 · TerminatedEndurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page