ENDURANT® ENLW1616C82ERecalls & reported problems
Medtronic, Inc. · STENT GRAFT ENLW1616C82E ENDURANT LIMB · Class 3 · Implant
All models of ENDURANT (46)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994761965), received between Sep 11, 2017 and Dec 6, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).
6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Deformation Due to Compressive Stress1
- Leak/Splash1
Patient problems
- Injury1
- Obstruction/Occlusion1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 6, 2021Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED DURING AN UNKNOWN ENDOVASCULAR PROCEDURE. IT WAS REPORTED BY THE PATIENT VIA TECHNICAL SERVICES THAT DURING IMPLANTATION OF ONE OF THE GRAFTS THERE WAS A KINK AND THE PATIENT LATER DEVELOPED CLAUDICATION. THE CLAUDICATION RESOLVED AFTER 1-2 YEARS WITH MEDICATION AND EXERCISE. IT WAS DETERMINED THAT COLLATERAL CIRCULATION DEVELOPED AND CLAUDICATION DIMINISHED. AFTER 4-5 YEARS DEVELOPED CLOGGING IN RIGHT FEMORAL ARTERY, AND DISTAL CREATING CLAUDICATION. NO CAUSE OF THE EVENT HAS BEEN PROVIDED. NO ADDITIONAL CLINICAL SEQUELAE HAS BEEN PROVIDED AND THE PATIENT IS FINE.
Patient problems: Obstruction/Occlusion; Pain
MAUDE report 12932610 ↗InjurySep 11, 2017Leak/Splash; Leak/Splash
AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN 8.0 CM ABDOMINAL AORTIC ANEURYSM. IT WAS NOTED THAT THERE WAS 30 DEGREES OF ANGULATION. IT WAS REPORTED THAT THE PATIENT PREVIOUSLY HAD ANOTHER MANUFACTURER¿S CUFF AND LIMB IMPLANTED FOR A TYPE IA AND IB, RESPECTIVELY. THE TYPE IB REPORTEDLY RESOLVED BUT THE TYPE IA REMAINED. THE PATIENT THEN PRESENTED EMERGENTLY WITH A MIGRATION AND TYPE IA ENDOLEAK. THE 18 HELI-FX ENDOANCHORS WERE IMPLANTED ALONG WITH A 32 ENDURANT AORTIC CUFF AND ANOTHER MANUFACTURER¿S STENT. A LEAK STILL REMAINED BUT IT COULD NOT BE DETERMINED IF IT WAS A TYPE IA OR A TYPE II. THE PHYSICIAN STATED THAT THE CAUSE OF THE MIGRATION AND TYPE IA WAS DUE TO DISEASE PROGRESSION; AND THE CAUSE OF THE TYPE IB WAS DUE TO UNDER-SIZING. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS BEING MONITORED BY THEIR PHYSICIAN.
Patient problems: Injury
MAUDE report 6855441 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.
- Z-0292-2022 · Oct 15, 2021 · Open, ClassifiedEndurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
- Z-0291-2022 · Oct 15, 2021 · Open, ClassifiedEndurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
- Z-0290-2022 · Oct 15, 2021 · Open, ClassifiedEndurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
- Z-2136-2021 · Jun 7, 2021 · Open, ClassifiedMedtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aort
- Z-2135-2021 · Jun 7, 2021 · Open, ClassifiedEndurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat i
- Z-1543-2017 · Feb 27, 2017 · TerminatedEndurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page