Skip to content

ENDURANT® ENLW1613C93ERecalls & reported problems

Medtronic, Inc. · STENT GRAFT ENLW1613C93E ENDURANT LIMB · Class 3 · Implant

All models of ENDURANT (46)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994761989), received between Aug 16, 2017 and Aug 30, 2017 — classified as 1 death, 1 injury.

The most frequently reported specific device problem is “Failure To Adhere Or Bond” (1 report).

6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Failure To Adhere Or Bond1

Patient problems

  • Death1
  • Injury1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 30, 2017Failure To Adhere Or Bond; Failure To Adhere Or Bond

    A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE PATIENT PRESENTED WITH A DISTAL TYPE I ENDOLEAK IN THE ILIAC LIMB. THE PHYSICIAN IMPLANTED AN ENDURANT STENT GRAFT. THE ENDOLEAK WAS RESOLVED. APPROXIMATELY 4 YEARS LATER, THERE WAS LOSS OF SEAL OF THE ENDURANT STENT GRAFT. THE PHYSICIAN IMPLANTED ANOTHER ENDURANT ILIAC LIMB TO RESOLVE THE LOSS OF SEAL. PER THE PHYSICIAN, THE CAUSE OF THE EVENTS WERE RELATED TO PATIENT ANATOMY, ILIAC LIMB DILATION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

    Patient problems: Injury

    MAUDE report 6829543 ↗
  • DeathAug 16, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A TALENT/ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE PATIENT EXPIRED. THE CAUSE OF DEATH HAS NOT BEEN OBTAINED.

    Patient problems: Death

    MAUDE report 6797373 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0292-2022 · Oct 15, 2021 · Open, Classified
    Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
  • Z-0291-2022 · Oct 15, 2021 · Open, Classified
    Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
  • Z-0290-2022 · Oct 15, 2021 · Open, Classified
    Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
  • Z-2136-2021 · Jun 7, 2021 · Open, Classified
    Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aort
  • Z-2135-2021 · Jun 7, 2021 · Open, Classified
    Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat i
  • Z-1543-2017 · Feb 27, 2017 · Terminated
    Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page