ENDURANT® ENCF2525C49ERecalls & reported problems
Medtronic, Inc. · STENT GRAFT ENCF2525C49E ENDURANT CUFF · Class 3 · Implant
All models of ENDURANT (46)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994762108), received between Aug 11, 2017 and Aug 11, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Leak/Splash” (1 report).
6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Leak/Splash1
Patient problems
- Injury1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2017Leak/Splash; Leak/Splash
AN ENDURANT STENT GRAFT AORTIC CUFF WAS IMPLANTED PROPHYLACTICALLY IN A PATIENT DURING AN ENDOVASCULAR TREATMENT FOR A TYPE II ENDOLEAK. THE PATIENT HAD BEEN LOST TO FOLLOW UP. APPROXIMATELY FOUR YEARS AND TWO MONTHS POST INDEX PROCEDURE, A CT REVEALED THAT THE PATIENT HAD A LIKELY PROXIMAL TYPE I ENDOLEAK WITH A SEAL ZONE REGRESSION OF A COUPLE OF MILLIMETERS. THE PHYSICIAN DETERMINED THE PROXIMAL TYPE I ENDOLEAK TO BE DUE TO PROGRESSIVE PROXIMAL AORTIC NECK DILATATION. THE ANEURYSM MEASURED 13 CM IN DIAMETER. THE PHYSICIAN ELECTED TO IMPLANT AN ENDURANT II AORTIC STENT GRAFT CUFF. IT WAS REPORTED THAT DURING THE SECONDARY PROCEDURE, 20FR SENTRANT SHEATH WAS ADVANCED THROUGH A PREVIOUSLY DEPLOYED BARE METAL STENT HOWEVER UNABLE TO BE ADVANCED TO INTENDED LOCATION. THE PHYSICIAN ELECTED TO REMOVE THE 20FR SENTRANT SHEATH AND MODIFIED THE CATHETER, CUTTING OFF A PORTION OF THE CATHETER. THE 20FR SENTRANT SHEATH WAS REINSERTED IN TO THE PATIENT AND THE ENDURANT AORTIC CUFF DELIVERY SYSTEM AND A FLUSH CATHETER WERE ADVANCED TO THE INTENDED LANDING ZONE. THE PHYSICIAN STATED THAT DURING DELIVERY OF THE ENDURANT AORTIC CUFF THERE WAS EXTREME DRAG, A FEW STENT RINGS WERE EXPOSED HOWEVER
Patient problems: Injury
MAUDE report 6787942 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.
- Z-0291-2022 · Oct 15, 2021 · Open, ClassifiedEndurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
- Z-0292-2022 · Oct 15, 2021 · Open, ClassifiedEndurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
- Z-0290-2022 · Oct 15, 2021 · Open, ClassifiedEndurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
- Z-2135-2021 · Jun 7, 2021 · Open, ClassifiedEndurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat i
- Z-2136-2021 · Jun 7, 2021 · Open, ClassifiedMedtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aort
- Z-1543-2017 · Feb 27, 2017 · TerminatedEndurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page