ENDURANT® ENBF2313C166ERecalls & reported problems

Medtronic, Inc. · STENT GRAFT ENBF2313C166E ENDURANT BIF · Class 3 · Implant

All models of ENDURANT (46)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00613994762313), received between Jun 5, 2017 and Feb 22, 2023 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (4 reports).

6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Leak/Splash4
  • Adverse Event Without Identified Device or Use Problem1
  • Occlusion Within Device1

Patient problems

  • Injury2
  • Blood Loss1
  • Hematoma1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Occlusion1
  • Pain1
  • Rupture1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 22, 2023Leak/Splash; Leak/Splash

    AN ENDURANT BIFURCATE STENT GRAFT SYSTEM WAS IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A AAA.  A CTA WAS PERFORMED , WHICH IDENTIFIED A PROXIMAL TYPE IA ENDOLEAK AND BILATERAL TYPE IB ENDOLEAKS.  PER THE PHYSICIAN THE CAUSE OF THE ENDOLEAKS  IS DISEASE PROGRESSION.  NO ADDITIONAL CLINICAL SEQUALAE WERE PROVIDED AND THE PATIENT WILL BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16416408 ↗
  • InjuryJun 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AN ENDURANT BIFURCATE STENT GRAFT SYSTEM WAS IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A >50MM ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE PATIENT PRESENTED EMERGENTLY ON THE 8 YEARS AND 3 MONTHS LATER WITH COMPLAINTS OF ABDOMINAL PAIN. A CT PERFORMED REVEALED A HEMATOMA AND LABORATORY DIAGNOSTIC TESTS CONFIRMED THE PATIENT WAS EXPERIENCING BLOOD LOSS CONSISTENT WITH A RUPTURED ANEURYSM. THERE WAS NO ENDOLEAK VISIBLE. INTERVENTION WAS PERFORMED ON THE SAME DATE AND A NON-MDT STENT GRAFT WAS IMPLANTED AS TREATMENT. AS PER THE PHYSICIAN THE CAUSE OF THE EVENT CAN NOT BE DETERMINED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT WILL BE MONITORED.

    Patient problems: Blood Loss; Hematoma; Pain; Rupture

    MAUDE report 10162272 ↗
  • InjuryFeb 10, 2020Leak/Splash; Leak/Splash

    AN ENDURANT BIFURCATE AND AN ENDURANT LIMB WERE IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A 80MM ABDOMINAL AORTIC ANEURYSM. IT WAS NOTED THAT WHEN THE PATIENT RETURNED FOR FOLLOW-UP, A CT PERFORMED SEVEN YEARS AND SEVEN MONTHS LATER, SHOWED A TYPE IA ENDOLEAK PRESENT IN THE BIFURCATE. A ENDOLEAK TYPE IB WAS PRESENT IN THE RIGHT COMMON ILIAC ARTERY IN THE ENDURANT LIMB ALSO. NO CAUSE OF THE EVENT HAS BEEN REPORTED. NO ADDITIONAL CLINICAL SEQUALAE WERE REPORTED AND THE PATIENT WILL BE MONITORED.

    Patient problems: Injury; No Consequences Or Impact To Patient

    MAUDE report 9687700 ↗
  • InjuryApr 4, 2018Leak/Splash; Leak/Splash

    AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 50MM DIAMETER ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT, APPROXIMATELY 4 YEARS POST INDEX PROCEDURE THE PATIENT PRESENTED WITH ABDOMINAL TENDERNESS. ON CT EXAM THERE APPEARED TO BE A TYPE IA ENDOLEAK. THE NEXT DAY, AFTER ENDOANCHORS DEPLOYMENT FOR THE TYPE IA ENDOLEAK, THE ANGIOGRAM REVEALED A JETTING ENDOLEAK FROM THE WALL OF THE MAIN BODY OF THE BIFURCATED DEVICE. AN AORTIC CUFF WAS DEPLOYED AND SEALED THE HOLE BUT ON THE NEXT ANGIOGRAM ANOTHER HOLE WAS OBSERVED IN THE CONTRALATERAL LIMB. THE PHYSICIAN RELINED THE LIMB WITH A 16X16X82 ENDURANT LIMB AND THE FINAL ANGIOGRAM SHOWED THE ENDOLEAK HAD RESOLVED. AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT CANNOT BE DETERMINED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

    Patient problems: Injury

    MAUDE report 7395049 ↗
  • InjuryJun 5, 2017Leak/Splash; Occlusion Within Device; Leak/Splash; Occlusion Within Device

    AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 5 CM DIAMETER ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE CT REVEALED A DISTAL TYPE I ENDOLEAK FROM THE ENDURANT IPSILATERAL LIMB OF THE BIFURCATED STENT GRAFT. THE ANEURYSM IS CURRENTLY 4.8 CM IN DIAMETER. THE PATIENT¿S ILIAC ARTERY HAS DILATED TO 24MM-18.5 MM IN DIAMETER. THE PHYSICIAN ELECTED TO IMPLANT AN ENDURANT ILIAC LIMB EXTENSION TO THE ORIGIN OF THE INTERNAL ARTERY. THE DISTAL TYPE I ENDOLEAK WAS RESOLVED. THERE IS A TYPE IV ENDOLEAK PRESENT IN THE IPSILATERAL EXTENSION HOWEVER THE PHYSICIAN STATED THAT THIS WAS DUE TO A LARGE DOSE OF ANTI-COAGULATE MEDICATION GIVEN AND SHORT PROCEDURE TIME. IT WAS ALSO NOTED THAT THE ENDURANT CONTRALATERAL LIMB WAS OCCLUDED. IT IS UNKNOWN WHEN THE OCCLUSION OCCURRED, SINCE THERE IS BLOOD FLOW THERE HAS NOT BEEN AND WILL NOT BE TREATMENT FOR THE OCCLUSION. PER THE PHYSICIAN, THE CAUSE OF THE DISTAL TYPE I ENDOLEAK WAS DUE TO USER ERROR. THE IPSILATERAL LIMB SHOULD HAVE BEEN EXTENDED DISTALLY AT THE INDEX PROCEDURE. THE CAUSE OF THE OCCLUSION IS RELATED TO THE PATIENT¿S ANATOMY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATI

    Patient problems: Occlusion

    MAUDE report 6612900 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0290-2022 · Oct 15, 2021 · Open, Classified
    Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
  • Z-0292-2022 · Oct 15, 2021 · Open, Classified
    Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
  • Z-0291-2022 · Oct 15, 2021 · Open, Classified
    Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
  • Z-2135-2021 · Jun 7, 2021 · Open, Classified
    Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infra
  • Z-2136-2021 · Jun 7, 2021 · Open, Classified
    Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoili
  • Z-1543-2017 · Feb 27, 2017 · Terminated
    Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates