Skip to content

ENDURANT® ENBF3620C166ERecalls & reported problems

Medtronic, Inc. · STENT GRAFT ENBF3620C166E ENDURANT BIF · Class 3 · Implant

All models of ENDURANT (46)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00613994762351), received between May 18, 2017 and Aug 12, 2021 — classified as 5 injuries.

The most frequently reported specific device problem is “Leak/Splash” (4 reports).

6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Leak/Splash4
  • Migration or Expulsion of Device2
  • Adverse Event Without Identified Device or Use Problem1
  • Material Deformation1

Patient problems

  • Insufficient Information2
  • Rupture2
  • Abdominal Pain1
  • Hemorrhage/Blood Loss/Bleeding1
  • Injury1
  • Low Blood Pressure/ Hypotension1
  • Myocardial Infarction1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 12, 2021Leak/Splash; Migration or Expulsion of Device; Leak/Splash; Migration or Expulsion of Device

    AN ENDURANT BIFURCATE STENT GRAFT SYSTEM WAS IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A 80MM ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT ON AN UNKNOWN DATE OF A CT, A TYPE IA ENDOLEAK AND STENT GRAFT MIGRATED WERE IDENTIFIED. IT WAS REPORTED IT WAS THE MAIN BODY BIFURCATE THAT MIGRATED. INTERVENTION WAS PERFORMED, WHERE A FENESTRATED NON MDT STENT GRAFT WAS IMPLANTED. THIS RESOLVED THE EVENT. AS PER THE PHYSICIAN THE CAUSE OF THE EVENT WAS ANATOMY AND MORPHOLOGY CHANGES LEADING TO A JUXTA RENAL ANEURYSM. NO ADDITIONAL CLINICAL SEQUELAE WERE PROVIDED AND THE PATIENT IS FINE.

    Patient problems: Insufficient Information

    MAUDE report 12311599 ↗
  • InjuryApr 16, 2021Leak/Splash; Material Deformation; Leak/Splash; Material Deformation

    AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN AN UNKNOWN ENDOVASCULAR PROCEDURE. IT WAS REPORTED BY THE PATIENT THROUGH TECHNICAL SERVICES, THAT APPROXIMATELY 7 YEAR POST INDEX PROCEDURE, A REPAIR WAS PERFORMED TO THE PREVIOUS IMPLANTED STENTS TO STRAIGHTEN THEM BACK OUT. AS PER PATIENT, IT WAS FOUND OUT THAT INSTEAD OF AN UPSIDE-DOWN "Y" SHAPE, THE STENTS GOTTEN INTO A "C" SHAPE. THE PATIENT DOES NOT THINK THE STENTS GRAFTS WERE EXPLANTED AND DOES NOT KNOW IF NEW STENT GRAFTS WERE IMPLANTED, BUT NOTES IT IS THOUGHT THEY WERE "REWRAPPED". NO CAUSE OF THE EVENT WAS REPORTED. NO ADDITIONAL CLINICAL SEQUALAE WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 11676763 ↗
  • InjurySep 6, 2018Leak/Splash; Migration or Expulsion of Device; Leak/Splash; Migration or Expulsion of Device

    AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN UNKNOWN DIAMETER ABDOMINAL AORTIC ANEURYSM. APPROX. 7 YEARS, 4 MONTHS POST INDEX PROCEDURE, IT WAS REPORTED THAT THE PATIENT PRESENTED EMERGENTLY WITH ABDOMINAL PAIN. A CT REVEALED A LARGE AAA WITH AN ENDOLEAK AND INTRA-ABDOMINAL BLOOD SUSPICIOUS FOR AN AORTIC RUPTURE. THE CT SHOWED THAT BOTH IMPLANTED LIMBS WERE ¿FLOATING¿ IN THE AAA SAC DUE TO TYPE IB ENDOLEAKS FROM DISEASE PROGRESSION. THE RIGHT LIMB ALONG WITH THE ENTIRE ILIAC ARTERY WAS OCCLUDED. IT WAS NOTED THAT THE NECK AREA APPEARED TO BE QUITE DILATED WITH A POSSIBLE TYPE IA ENDOLEAK AND IT LOOKED LIKE THE STENT GRAFT HAD SLIPPED DOWN 6MM BELOW THE RENALS. THE AORTIC NECK DIAMETER WAS 34MM AND THE AAA DIAMETER WAS 82MM. NECK LENGTH WAS 6MM WITH A PROXIMAL NECK ANGULATION OF 45 DEGREES. INTERVENTION WAS CARRIED OUT WITH THE IMPLANTATION OF AN ENDURANT AUI AND EXTENSION TO THE EXTERNAL ILIAC ARTERY WITH TWO FURTHER LIMBS. TOUCH UP WAS THEN PERFORMED USING A RELIANT BALLOON. 4 ENDOANCHORS WERE IMPLANTED NEAR THE TOP OF THE AORTIC AUI IMPLANT TO ENHANCE SEAL AND FIXATION. ANGIOGRAM SHOWED GOOD PATENCY AND NO SIGN OF AN ENDOLEAK. A

    Patient problems: Abdominal Pain; Myocardial Infarction; Rupture

    MAUDE report 7849699 ↗
  • InjuryAug 7, 2018Leak/Splash

    AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 50MM DIAMETER ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT APPROXIMATELY 7 YEARS POST INDEX PROCEDURE, THE PATIENT PRESENTED EMERGENTLY WITH A TYPE IA ENDOLEAK. THE PHYSICIAN CARRIED OUT TREATMENT WITH THE PLACEMENT OF ANOTHER MANUFACTURER STENT GRAFT PROXIMALLY AND ANOTHER MANUFACTURER CUFF TO SUCCESSFULLY TREAT THE TYPE IA ENDOLEAK. ALTHOUGH BOTH EXISTING LIMBS WERE PATENT, THE PHYSICIAN CHOOSE TO RE-LINE THEM AS BOTH HAD CIRCUMFERENTIAL THROMBUS BILATERALLY. PER THE PHYSICIAN, THE CAUSE OF THE EVENT CANNOT BE DETERMINED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Injury; Thrombus

    MAUDE report 7755894 ↗
  • InjuryMay 18, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AN ENDURANT STENT GRAFT SYSTEM WERE IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT FOR AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT A PATIENT PRESENTED WITH BACK PAIN AND LOW BLOOD PRESSURE, BUT WAS IN STABLE CONDITION. THE PHYSICIAN STATED THAT IT WAS NOT CAUSED BY THE PREVIOUS ENDURANT STENT GRAFT BUT THE AORTA ABOVE THE RENAL ARTERIES HAD EXPANDED AND POSSIBLY RUPTURED, BUT WAS CONTAINED. THE FILMS WERE REVIEWED AND THE PHYSICIAN DECIDED TO IMPLANT AN ENDURANT AORTIC CUFF ABOVE THE RENAL ARTERY BUT BELOW THE SUPERIOR MESENTERIC ARTERY. THIS WAS DONE SUCCESSFULLY WITH AN ENDURANT 36X36X70 AORTIC CUFF. AN ANGIOGRAM WAS THEN PERFORMED AND THERE WAS CONTRAST GETTING AROUND THE STENT GRAFT PROXIMALLY BECAUSE THE GRAFT WASN'T LONG ENOUGH TO REACH THE SEAL ZONE. THE PHYSICIAN DECIDED TO CANNULATE THE SMA COMING FROM THE BRACHIAL ARTERY. ANOTHER MANUFACTURER¿S COVERED STENT WAS ADVANCED INTO THE SMA UP TO THE LEVEL OF THE CELIAC ARTERY. ANOTHER ENDURANT 36X36X49 CUFF WAS THEN IMPLANTED TO THE LEVEL OF THE OTHER MANUFACTURER¿S COVERED STENT AND THE CELIAC ARTERY. BOTH WERE DEPLOYED SIMULTANEOUSLY AND A SNORKEL WAS CREATED. THE ENDURANT CUFF WAS THEN BALLOONED USING A REL

    Patient problems: Low Blood Pressure/ Hypotension; Pain; Rupture

    MAUDE report 6575826 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0291-2022 · Oct 15, 2021 · Open, Classified
    Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
  • Z-0290-2022 · Oct 15, 2021 · Open, Classified
    Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
  • Z-0292-2022 · Oct 15, 2021 · Open, Classified
    Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
  • Z-2135-2021 · Jun 7, 2021 · Open, Classified
    Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infra
  • Z-2136-2021 · Jun 7, 2021 · Open, Classified
    Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoili
  • Z-1543-2017 · Feb 27, 2017 · Terminated
    Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page