Skip to content

ENDURANT® ENBF3220C145ERecalls & reported problems

Medtronic, Inc. · STENT GRAFT ENBF3220C145E ENDURANT BIF · Class 3 · Implant

All models of ENDURANT (46)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994762405), received between Nov 1, 2021 and Nov 1, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 1, 2021Leak/Splash; Leak/Splash

    AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 50MM  ABDOMINAL AORTIC ANEURYSM.  IT WAS REPORTED THAT DURING A FOLLOW UP CT SCAN, A SMALL TYPE 1A ENDOLEAK WAS NOTED IN THE BIFURCATE GRAFT . NO COMMUNICATION INTO THE AAA IS EVIDENT . A SLIGHT TYPE 1B ENDOLEAK WAS OBSERVED IN THE RIGHT COMMON ILIAC ARTERY. IT WAS STATED THAT THE RIGHT ILIAC HAS ONLY 6 MM OF SEAL APPROXIMATELY.  AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.  NO ADDITIONAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12727090 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0292-2022 · Oct 15, 2021 · Open, Classified
    Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
  • Z-0291-2022 · Oct 15, 2021 · Open, Classified
    Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
  • Z-0290-2022 · Oct 15, 2021 · Open, Classified
    Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
  • Z-2135-2021 · Jun 7, 2021 · Open, Classified
    Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat i
  • Z-2136-2021 · Jun 7, 2021 · Open, Classified
    Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aort
  • Z-1543-2017 · Feb 27, 2017 · Terminated
    Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page