CD HORIZON® Spinal System 1606000500Recalls & reported problems
Medtronic Sofamor Danek, Inc. · ROD 1606000500 6.0 CCM PLS NS STRT 500MM · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613994974297), received between Jul 14, 2017 and Jun 22, 2021 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K113174); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Device Damaged by Another Device1
Patient problems
- Bone Fracture(s)1
- No Code Available1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 22, 2021Device Damaged by Another Device; Device Damaged by Another Device
THE INFORMATION WAS RECEIVED FROM HEALTH CARE PROFESSIONAL VIA A MANUFACTURING REPRESENTATIVE REGARDING A PATIENT WITH SPINE STABILIZATION/FUSION FOR THORACOLUMBAR FUSION SPINAL THERAPY. IT WAS REPORTED THAT FROM RIGHT SIDE IMPLANTS THE SET SCREW WAS LOOSE. SADDLE OF THE BALLAST SCREW SEEMED DEFECTIVE WHEN HE USED THE TULIP ALIGNER FOR SCREW REMOVAL. FROM LEFT SIDE IMPLANTS SET SCREW SEEMED TIGHT AND NORMAL. TULIP ANGLE OF BALLAST SCREW SEEMED CHANGED SINCE POST OP STANDING FILMS. THE PATIENT HAD PELVIC FRACTURE SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT. REMOVAL OF BALLAST SCREWS, SET SCREWS AND RODS. REPLACED WITH LARGER DIAMETER BALLAST SCREWS, NEW SET SCREWS AND NEW RODS. IT WAS UNSURE IF THE RODS HAD ANY PART OF THE PRODUCT FAILURE. SENT THEM WITH THE SCREWS AND THE SET SCREWS SO ALL AVAILABLE IMPLANTS WERE TOGETHER TO GET TESTED. WE CAN NOT SAY WITH CERTAINTY THE FRACTURE WAS A RESULT OF THE IMPLANTS.
Patient problems: Bone Fracture(s)
MAUDE report 12038725 ↗MalfunctionJul 20, 2017Break; Break
PROCEDURE: SCOLIOSIS ADULT LEVELS: S1-ILIAC-T3 IT WAS REPORTED THAT ON (B)(6) 2017, POST-OP, ROD WAS FOUND TO BE BROKEN IN X-RAY IMAGE. PATIENT STATUS WAS REPORTED UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6727371 ↗InjuryJul 14, 2017Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
PATIENT CLINICAL ID: LUCUS 1-2-3-4 PRE-OP DIAGNOSIS: DISBALANCE SAGITAL LUMBAR PROCEDURE: SCOLIOSIS DISBALANCE SAGITAL LEVEL: T3-ILLIAC IT WAS REPORTED THAT PATIENT UNDERWENT A SURGERY. POST-OP, PAIN AND INSTABILITY WAS OBSERVED IN THE PATIENT.
Patient problems: No Code Available; Pain
MAUDE report 6713265 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113174. One submission often covers many models, so check the product and lot details in each record.
- Z-2603-2018 · Jul 9, 2018 · TerminatedMedtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page