miniMark Dental Implant System 09270704Recalls & reported problems
Ace Surgical Supply CO., Inc. · miniMARK BALL 1.8/2.3 X 13MM 2MM CUFF RBM IMPLANT W/KEEPER · Class 2 · Implant
All models of miniMark Dental Implant System (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00614950002870), received between Dec 22, 2015 and Oct 23, 2017 — classified as 4 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Osseointegration Problem4
- Failure to Osseointegrate2
Patient problems
- Failure of Implant4
- Pain3
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 23, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
PAIN AND MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 6969287 ↗InjuryOct 23, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
PAIN WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. FRACTURE OF THE IMPLANT WAS ALSO REPORTED.
Patient problems: Failure of Implant; Pain
MAUDE report 6969283 ↗InjuryDec 14, 2016Osseointegration Problem; Osseointegration Problem
PAIN AND MOBILITY WAS REPORTED. NO OSSEOINTEGRATION OF THE IMPLANT 3-6 MONTHS BEFORE DISCOVERY. BONE QUALITY AT THE TIME OF IMPLANT SURGERY TYPE II.
Patient problems: Failure of Implant; Pain
MAUDE report 6173263 ↗InjuryDec 22, 2015Osseointegration Problem
INFECTION AND MOBILITY WAS REPORTED AT THE TIME OF IMPLANT FAILURE. TIME OF LOSS WAS 0-1 MONTHS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT TIME OF FAILURE WAS TYPE II.
Patient problems: Failure of Implant
MAUDE report 5321053 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.