ALLEN 108001Recalls & reported problems
Hill-Rom, Inc. · Boot Assembly, Traction Device,OT 1000 · Class 1
All models of ALLEN (20)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00615521000851), received between Feb 6, 2024 and Feb 7, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Feb 14, 2024Class IIOpen, Classified
Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted device is used during surgery, the boot may detach from the traction system, potentially leading to patient harm and a delay in surgery.
Z-1289-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Manufacturing, Packaging or Shipping Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
BAXTER RECEIVED A REPORT FROM A DISTRIBUTOR NOTING IT WAS FOUND 2 NON-SERIALIZED TRACTION BOOTS FROM SAME LOT (23-268), WHERE THE BOOT BASE HAS BEEN INSTALLED IS FACING THE WRONG DIRECTION. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18660886 ↗MalfunctionFeb 7, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
BAXTER RECEIVED A REPORT FROM A CUSTOMER NOTING TRACTION BOOT WHERE THE BOOT BASE HAS BEEN INSTALLED IS FACING THE WRONG DIRECTION. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18660707 ↗MalfunctionFeb 6, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
BAXTER RECEIVED A REPORT FROM A CUSTOMER NOTING TRACTION BOOT WHERE THE BOOT BASE HAS BEEN INSTALLED IS FACING THE WRONG DIRECTION. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18652130 ↗
FDA clearance and classification
The FDA regulates product code FWZ (operating room accessories table tray) under 21 CFR 878.4950 as a Class I device. 21 CFR 878.4950 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ALLEN 108001
Has the ALLEN 108001 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1289-2024 (Class II), began on Feb 14, 2024: Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ALLEN 108001?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Feb 6, 2024 and Feb 7, 2024. In those reports, the most common specific device problem is “Manufacturing, Packaging or Shipping Problem” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ALLEN 108001?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ALLEN 108001, and is it still sold?
The FDA's device database lists Hill-Rom, Inc. as the company responsible for it (brand ALLEN, model 108001). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Operating Room Accessories Table Tray devices
- MAYFIELD® A2101Integra Lifesciences Corporation · Operating Room Accessories Table Tray100 reports
- NA 210080Mako Surgical Corp. · Operating Room Accessories Table Tray74 reports
- ALLEN 104002Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall8 reports
- N/A 7210569Smith & Nephew, Inc. · Accessories, operating-room, table (kit)57 reports
- ALLEN 107004Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall5 reports
- NA 7210551Smith & Nephew, Inc. · Accessories, operating-room, table (kit)41 reports
- MAYFIELD® A1018Integra Lifesciences Corporation · Operating Room Accessories Table Tray36 reports
- MAYFIELD® A3101Integra Lifesciences Corporation · Operating Room Accessories Table Tray33 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page