Skip to content

MAYFIELD® A2101Recalls & reported problems

Integra Lifesciences Corporation · MAYFIELD® ULTRA BASE UNIT · Class 1

All models of MAYFIELD (32)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Mayfield Ultra Base Unit is an operating-room accessory table tray. Reports mostly describe mechanical loosening or slipping during surgery, sometimes causing lacerations, abrasions, or bleeding. There are 98 adverse event reports, no deaths, 14 injuries, and 84 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
100
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 100 adverse event reports that list this exact device (DI 10381780267997), received between Sep 18, 2020 and Sep 3, 2026 — classified as 14 injuries, 85 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (33 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 3 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem33
  • Device Slipped22
  • Mechanics Altered16
  • Unstable13
  • Unintended Movement5
  • Physical Resistance/Sticking4
  • Mechanical Jam3
  • Naturally Worn3

Patient problems

  • No Clinical Signs, Symptoms or Conditions81
  • Laceration(s)7
  • Insufficient Information5
  • Abrasion3
  • Hemorrhage/Blood Loss/Bleeding1
  • Neck Stiffness1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 25 most recent of 100 reports, in the reporter's own words as published by the FDA.

  • OtherSep 3, 2026Device Slipped

    Patient problems: Laceration(s)

    MAUDE report 50599183 ↗
  • MalfunctionAug 28, 2026Unstable

    Patient problems: Insufficient Information

    MAUDE report 50501972 ↗
  • MalfunctionAug 11, 2026Physical Resistance/Sticking

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50216691 ↗
  • MalfunctionMay 8, 2026Unstable

    THIS IS 1 OF 2 REPORTS LINKED TO MFG REPORT NUMBER 3004608878-2026-00055: A FACILITY REPORTED THAT DURING SURGERY, THE MAYFIELD ULTRA BASE UNIT (A2101) LOOSENED TWICE WHILE DRILLING INTO THE PATIENT¿S SKULL. THE ISSUE OCCURRED INTRAOPERATIVELY, WHILE THE DEVICE WAS IN USE TO SECURE THE PATIENT¿S HEAD. THE MAYFIELD CLAMP WAS RETAINED AS THE SURGERY WAS ALREADY UNDERWAY, AND THE SURGICAL PROCEDURE WAS COMPLETED WITHOUT CHANGING THE DEVICE. THERE WAS NO INJURY OR INCREASED SURGERY TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25125801 ↗
  • MalfunctionMay 1, 2026Device Slipped

    A FACILITY REPORTED THAT AFTER PATIENT WAS POSITIONED PRONE USING THE MAYFIELD HEAD HOLDER, A SLIGHT DOWNWARD SLIP OCCURRED. THE SURGICAL TEAM CONFIRMED THAT THE SLIP DID NOT OCCUR WITHIN THE MAYFIELD CLAMP, AND THE MAYFIELD ULTRA BASE (A2101), IS SUSPECTED TO HAVE CAUSED THE SLIP. A SIMILAR SLIP OCCURRED 2 WEEKS PRIOR, AND ALL OF THE HEAD HOLDERS WERE INSPECTED, BUT NO DEFECTS WERE FOUND AT THE TIME OF INSPECTION. AFTER THE SECOND OCCURRENCE, IT WAS FOUND THAT COUPLE OF BOLTS ON THE BASE MIGHT NOT BE TIGHTENED TO SPECIFICATIONS. NO INJURY OCCURRED AND NO INFORMATION ON SURGICAL DELAY HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25060277 ↗
  • InjuryMar 14, 2026Unstable

    THIS IS 2 OF 3 REPORTS LINKED TO MFG REPORT NUMBERS: 3004608878-2026-00007 AND 3004608878-2026-00031: A FACILITY REPORTED THAT THE MAYFIELD ULTRA BASE UNIT (A2101) CAME LOOSE DURING SURGERY, RESULTING IN A SINGLE SCRATCH TO THE PATIENT'S HAIRLINE/FOREHEAD, AND A BANDAGE WAS APPLIED TO THE FOREHEAD. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THE SURGERY TIME WAS NOT SIGNIFICANTLY INCREASED.

    Patient problems: Abrasion

    MAUDE report 24599763 ↗
  • MalfunctionMar 9, 2026Device Slipped

    THIS IS 1 OF 3 REPORTS LINKED TO MFG REPORT NUMBERS 3004608878-2026-00026 AND 3004608878-2026-00027: A FACILITY REPORTED THAT THE MAYFIELD ULTRA BASE UNIT (A2101) WAS PART OF A CERVICAL STABILIZATION CASE WHERE THE PHYSICIAN STATED THAT THE ENTIRE SYSTEM FELT LIKE IT SLIPPED. THE PROCEDURE WAS COMPLETED PER USUAL WITH NO INJURY TO THE PATIENT, WHO IS RECOVERING AS EXPECTED. ADDITIONALLY, THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24549058 ↗
  • MalfunctionFeb 27, 2026Device Slipped

    A FACILITY REPORTED THAT AFTER THE PATIENT WAS POSITIONED PRONE USING THE MAYFIELD HEAD HOLDER, AND A SLIGHT DOWNWARD SLIP OCCURRED. THE SURGICAL TEAM CONFIRMED THAT THE SLIP DID NOT OCCUR WITHIN THE MAYFIELD CLAMP, AND THE MAYFIELD ULTRA BASE UNIT (A2101) IS SUSPECTED TO HAVE CAUSED THE SLIP. IT IS SUSPECTED THAT COUPLE OF BOLTS ON THE BASE MIGHT NOT BE TIGHTENED TO SPECIFICATIONS. THERE WAS NO PATIENT INJURY AND SURGICAL DELAY IS UNKNOWN. ADDITIONAL INFORMATION IS BEING REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24476952 ↗
  • InjuryFeb 20, 2026Device Slipped

    THIS IS 1 OF 3 REPORTS LINKED TO MFG REPORT NUMBERS 3004608878-2026-00017 AND 3004608878-2026-00019: A FACILITY REPORTED THAT THE PATIENT WAS IN THE PRONE POSITION FOR A CRANIOTOMY PROCEDURE AND SLIPPED FROM THE MAYFIELD ULTRA BASE UNIT (A2101). THIS RESULTED IN AN INCH AND A HALF DEEP LACERATION TO THE PATIENT'S LEFT SCALP WHICH REQUIRED SUTURING. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Laceration(s)

    MAUDE report 24405450 ↗
  • InjuryFeb 12, 2026Unstable

    N/A.

    Patient problems: Laceration(s)

    MAUDE report 24329068 ↗
  • MalfunctionJan 29, 2026Unstable

    A FACILITY REPORTED THAT AT THE BEGINNING OF SURGERY IN A PATIENT WITH SPHENOIDAL MENINGIOMA, THE INTERMEDIARY PART (TRANSITIONAL MEMBER) OF THE MAYFIELD ULTRA BASE UNIT (A2101) BECAME LOOSE. THE PATIENT'S HEAD WAS NO LONGER PROPERLY FIXED/ATTACHED, AND THE OPERATING POSITION HAD BEEN CHANGED. THERE WERE NO OTHER CONSEQUENCE FOR THE PATIENT. HOWEVER, THERE COULD HAVE BEEN A RISK/A WRONG OPERATING "MOVE" IF IT WOULD HAVE OCCURRED DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24209620 ↗
  • MalfunctionDec 12, 2025Mechanics Altered; Mechanics Altered

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23796812 ↗
  • MalfunctionDec 12, 2025Mechanics Altered; Mechanics Altered

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23796702 ↗
  • MalfunctionNov 26, 2025Device Slipped; Device Slipped

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23662847 ↗
  • MalfunctionNov 20, 2025Material Twisted/Bent; Material Twisted/Bent

    A FACILITY REPORTED THAT THE MAYFIELD ULTRA BASE UNIT (A2101) WAS BENT AND DOES NOT FIT IN THE OPERATING TABLE FIXTURE. THE PATIENT WAS ASLEEP UNDER ANESTHESIA; HOWEVER, THERE WAS NO PATIENT INJURY. ANOTHER MAYFIELD WAS USED AND THERE WAS INCREASED SURGERY TIME OF 30 MINUTES DUE TO THE REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23605664 ↗
  • InjuryJul 21, 2025Unstable; Unstable

    THIS IS 4 OF 4 REPORTS FOR THE MAYFIELD SYSTEM LINKED TO MFG REPORT NUMBERS: 3004608878-2025-00148, 3004608878-2025-00149, 3004608878-2025-00151. A FACILITY REPORTED THAT THE MAYFIELD ULTRA BASE UNIT (A2101) NEEDS TO BE CHECKED AS IT GOT LOOSE/BECAME UNLOCKED DURING SURGERY, RESULTING IN WHIPLASH FOR THE PATIENT. IT IS UNCLEAR IF THE EVENT LED TO INCREASED SURGERY TIME. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Neck Stiffness

    MAUDE report 22555449 ↗
  • MalfunctionJun 20, 2025Mechanics Altered; Mechanics Altered

    AN AUTHORIZED DISTRIBUTOR REPORTED THAT DURING THE OPERATION, THE MAYFIELD ULTRA BASE UNIT (A2101) DID NOT MAINTAIN CORRECT TIGHTENING. SURGERY TIME WAS INCREASED FOR MORE THAN 30 MINUTES. THERE WAS NO PATIENT INJURY. END USER HOSPITAL: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22296947 ↗
  • MalfunctionJun 20, 2025Device Slipped; Device Slipped

    A FACILITY REPORTED THAT A PATIENT SLIPPED WHILE THE MAYFIELD ULTRA BASE UNIT (A2101) WAS BEING SET UP, AND THE GOLD HANDLE WAS FOUND TO BE LOOSE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 22296945 ↗
  • MalfunctionJun 10, 2025Unstable; Unstable

    THIS IS 1 OF 2 REPORTS LINKED TO MFG REPORT NUMBER 3004608878-2025-00121: A FACILITY REPORTED THE MAYFIELD ULTRA BASE UNIT (A2101) DOES NOT CLOSE WELL AND MOVES EVEN WHEN LOCKED. THE PATIENT WAS ASLEEP, AND THERE WAS INCREASED SURGERY TIME IN BEGINNING THE SURGERY. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22180013 ↗
  • MalfunctionMay 21, 2025Mechanics Altered; Mechanics Altered

    THIS IS 1 OF 2 REPORTS LINKED TO MFG REPORT NUMBER 3004608878-2025-00108: A FACILITY REPORTED THAT THE MAYFIELD ULTRA BASE UNIT (A2101) WAS USED IN A CASE, AND THE JOINT COULD NOT HOLD. AS A RESULT, THE DEVICE WAS CHANGED. THERE WAS NO PATIENT INJURY; HOWEVER, THERE WAS SIGNIFICANT INCREASED SURGICAL TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22060296 ↗
  • MalfunctionApr 24, 2025Unstable; Unstable

    THIS IS 1 OF 3 REPORTS LINKED TO MFG REPORT NUMBERS 3004608878-2025-00080 AND 3004608878-2025-00081: A FACILITY REPORTED THAT ONCE FIXED, THE MAYFIELD ULTRA BASE UNIT (A2101) WAS NOT STABLE. THE PROBLEM WAS FOUND ON THE PATIENT UNDER ANESTHESIA, BEFORE THE SURGERY. SINCE THE PATIENT COULD NOT GIVE UP THIS SURGERY, WE MINIMIZED THE MOVEMENT OF THE HEADBOARD AS MUCH AS POSSIBLE, TAKING INTO ACCOUNT ITS IMPERFECT STABILITY FOR THE ENTIRE DURATION OF THE SURGERY. THERE WAS NO INCREASED SURGERY TIME OR PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21901683 ↗
  • MalfunctionApr 3, 2025Unstable; Unstable

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21764806 ↗
  • MalfunctionMar 20, 2025Degraded; Degraded

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21657302 ↗
  • MalfunctionMar 13, 2025Unstable; Unstable

    THIS REPORT IS 3 OF 3 AND LINKED TO MFG REPORT NUMBERS 3004608878-2025-00044 AND 3004608878-2025-00049: A FACILITY REPORTED THAT THE MAYFIELD ULTRA BASE UNIT (A2101) IS MOBILE WHEN IT SHOULD NOT BE - TIGHTENING PROBLEM. THE ISSUE WAS NOTED BEFORE SURGERY WHILE THE PATIENT WAS UNDER ANESTHESIA. THERE WAS NO PATIENT INJURY AND NO INCREASED SURGERY TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21602011 ↗
  • MalfunctionFeb 13, 2025Device Slipped; Device Slipped

    THIS REPORT IS 3 OF 3 REPORTS LINKED TO MFG REPORT NUMBERS: 3004608878-2025-00022, 3004608878-2025-00023. A FACILITY REPORTED THAT THE MAYFIELD ULTRA BASE UNIT (A21019) WAS SLIPPING AND NOT MAINTAINING THE FULL POSITION DURING SURGERY. ANOTHER HEADREST WAS USED TO COMPLETE THE SURGERY, AND THERE WAS NO DELAY OR PATIENT INJURY REPORTED. FACILITY HAS INDICATED THAT THE COMPLETE MAYFIELD SYSTEM IS BEING RETURNED TO KEEP ALL PARTS TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21381343 ↗

FDA clearance and classification

The FDA regulates product code FWZ (operating room accessories table tray) under 21 CFR 878.4950 as a Class I device. 21 CFR 878.4950 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the MAYFIELD® A2101

Has the MAYFIELD® A2101 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10381780267997), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the MAYFIELD® A2101?

The FDA's MAUDE database lists 100 adverse event reports for this device, received between Sep 18, 2020 and Sep 3, 2026. In those reports, the most common specific device problems are “Mechanical Problem” (33), “Device Slipped” (22) and “Mechanics Altered” (16); the most common patient problems are “Laceration(s)” (7), “Abrasion” (3) and “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MAYFIELD® A2101?

None of the reports are classified as deaths, 14 as injuries and 85 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the MAYFIELD® A2101 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the MAYFIELD® A2101, and is it still sold?

The FDA's device database lists Integra Lifesciences Corporation as the company responsible for it (brand MAYFIELD®, model A2101). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Operating Room Accessories Table Tray devices

All Operating Room Accessories Table Tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page