Astral 100Recalls & reported problems
Resmed Pty Ltd · Portable ventilator · Class 2
All models of Astral (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This ResMed Astral 100 portable home ventilator provides breathing support. Most of the 864 reports describe battery and charging problems, calibration problems and screens that show wrong messages or stop responding. The reports include 9 deaths and 6 injuries; some describe a patient who became unresponsive during ventilator support, with caregivers reportedly hearing no alarm. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 859
- Deaths reported
- 9
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 859 adverse event reports that list this exact device (DI 00619498270019), received between Mar 3, 2016 and Aug 28, 2026 — classified as 9 deaths, 8 injuries, 842 malfunctions.
The most frequently reported specific device problem is “Protective Measures Problem” (315 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Protective Measures Problem315
- Battery Problem168
- Failure to Calibrate99
- Device Displays Incorrect Message91
- Failure to Charge79
- Display or Visual Feedback Problem58
- Inadequate User Interface42
- Failure of Device to Self-Test39
Patient problems
- No Clinical Signs, Symptoms or Conditions384
- No Consequences Or Impact To Patient366
- Insufficient Information50
- No Patient Involvement49
- Death5
- Loss of consciousness5
- Low Oxygen Saturation4
- Choking2
Recent adverse event reports
The 25 most recent of 859 reports, in the reporter's own words as published by the FDA.
DeathAug 28, 2026No Audible Alarm
It was reported to Resmed that a patient expired while using an Astral 100 device. It was reported that the patient became unresponsive while receiving ventilatory support from the device and no alarms were reportedly heard by the caregivers at the time of the event. The mother called for assistance, and family members initiated CPR prior to the arrival of emergency responders. Upon arrival of emergency responders, they assumed resuscitation efforts, and transported the patient by ambulance to a hospital, after which the patient was transferred to another hospital for a higher level of care. The patient was pronounced deceased at the hospital.
Patient problems: Loss of consciousness
MAUDE report 50505593 ↗DeathAug 28, 2026No Audible Alarm
Patient problems: Loss of consciousness
MAUDE report 50505590 ↗InjuryAug 19, 2026Use of Device Problem
It was reported to Resmed that an Astral device displayed a 'low pressure' alarm while in use. Learn Circuit was performed and the device displayed 'without SpO2 monitoring'. The patient was placed on a different device. The patient's SpO2 was 88% and manual ventilation was provided.
Patient problems: Low Oxygen Saturation
MAUDE report 50332128 ↗InjuryAug 19, 2026Use of Device Problem
It was reported to Resmed that an Astral device displayed a 'low pressure' alarm while in use. Learn Circuit was performed and the device displayed 'without SpO2 monitoring'. The patient was placed on a different device. The patient's SpO2 was 88% and manual ventilation was provided.
Patient problems: Low Oxygen Saturation
MAUDE report 50331994 ↗MalfunctionJun 10, 2026Failure to Power Up
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25462695 ↗MalfunctionJun 5, 2026Battery Problem
DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED A BATTERY INOPERABLE ALARM AND FAILED TO CHARGE ITS INTERNAL BATTERY. THERE WAS NO PATIENT HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25417974 ↗MalfunctionJun 1, 2026Power Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD A TOTAL POWER FAILURE DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25351668 ↗MalfunctionMay 14, 2026Power Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE ALARMED WHEN PLUGGED IN AND DID NOT POWER ON. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25178451 ↗MalfunctionMay 14, 2026Power Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE ALARMED WHEN PLUGGED IN AND DID NOT POWER ON. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25178450 ↗MalfunctionMay 7, 2026Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25115795 ↗MalfunctionMay 7, 2026Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25115794 ↗MalfunctionApr 22, 2026Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF82) RELATED TO THE ALARM SYSTEM. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24957898 ↗MalfunctionApr 22, 2026Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF82) RELATED TO THE ALARM SYSTEM. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24957849 ↗MalfunctionApr 14, 2026Battery Problem; Protective Measures Problem
DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24877352 ↗MalfunctionApr 14, 2026Battery Problem; Protective Measures Problem
DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24877350 ↗MalfunctionApr 9, 2026Battery Problem; Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED BATTERY ERROR MESSAGES. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24836181 ↗MalfunctionApr 9, 2026Battery Problem; Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED BATTERY ERROR MESSAGES. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24836177 ↗MalfunctionApr 9, 2026Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF147) RELATED TO THE MAIN BLOWER. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24835209 ↗MalfunctionApr 9, 2026Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF147) RELATED TO THE MAIN BLOWER. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24835208 ↗MalfunctionApr 9, 2026Inadequate User Interface
DURING RESMED EVALUATION, AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24835104 ↗MalfunctionApr 9, 2026Inadequate User Interface
DURING RESMED EVALUATION, AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24835103 ↗MalfunctionApr 2, 2026Protective Measures Problem; Complete Loss of Power
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A TOTAL POWER FAILURE ALARM AND POWER FAULT CODES. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THE INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24770133 ↗MalfunctionApr 2, 2026Protective Measures Problem; Complete Loss of Power
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A TOTAL POWER FAILURE ALARM AND POWER FAULT CODES. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THE INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24770131 ↗MalfunctionApr 1, 2026Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A POWER FAULT/NO CHARGING ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760417 ↗MalfunctionApr 1, 2026Protective Measures Problem
IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A POWER FAULT/NO CHARGING ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760402 ↗
FDA clearance and classification
The FDA regulates product code NOU (continuous, ventilator, home use) under 21 CFR 868.5895 as a Class II device. 21 CFR 868.5895 ↗
Devices of this type normally need 510(k) clearance before sale.
The FDA lists this type of device as life-sustaining or life-supporting.
FDA definition of this device type: This product code was needed for the home use indication. CBK is not a tracked device and NOU is a tracked device.
Common questions about the Astral 100
Has the Astral 100 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00619498270019), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Astral 100?
The FDA's MAUDE database lists 859 adverse event reports for this device, received between Mar 3, 2016 and Aug 28, 2026. In those reports, the most common specific device problems are “Protective Measures Problem” (315), “Battery Problem” (168) and “Failure to Calibrate” (99); the most common patient problems are “Death” (5), “Loss of consciousness” (5) and “Low Oxygen Saturation” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Astral 100?
9 reports are classified as deaths, 8 as injuries and 842 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Astral 100 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Astral 100, and is it still sold?
The FDA's device database lists Resmed Pty Ltd as the company responsible for it (brand Astral, model 100). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Continuous, Ventilator, Home Use devices
- Astral 150Resmed Pty Ltd · Continuous, Ventilator, Home Use1 recall17 deaths2,340 reports
- Trilogy EV300, USA DS2200X11BRespironics, Inc. · Continuous, ventilator, home use5 recalls2,069 reports
- Trilogy 00Respironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE15 deaths2,059 reports
- Trilogy Evo, Japan JP2110X16BRespironics, Inc. · Continuous, ventilator, home use5 recalls26 deaths1,095 reports
- Trilogy Trilogy Evo Ventilator, IntlRespironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE4 recalls3 deaths645 reports
- Life2000 MS-01-0118Welch Allyn, Inc. · Continuous, ventilator, home use10 recalls2 reports
- Trilogy Evo GB2110X15BRespironics, Inc. · Continuous, ventilator, home use5 recalls1 death140 reports
- Trilogy Evo FX2100X15BRespironics, Inc. · Continuous, ventilator, home use5 recalls1 death125 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page