Skip to content

Astral 100Recalls & reported problems

Resmed Pty Ltd · Portable ventilator · Class 2

Quick summary

This ResMed Astral 100 portable home ventilator provides breathing support. Most of the 864 reports describe battery and charging problems, calibration problems and screens that show wrong messages or stop responding. The reports include 9 deaths and 6 injuries; some describe a patient who became unresponsive during ventilator support, with caregivers reportedly hearing no alarm. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
865
Deaths reported
9
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 865 adverse event reports that list this exact device (DI 00619498270019), received between Mar 3, 2016 and Aug 28, 2026 — classified as 9 deaths, 7 injuries, 849 malfunctions.

The most frequently reported specific device problem is “Protective Measures Problem” (317 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 4 reportsSep 2024: 2 reportsOct 2024: 2 reportsNov 2024: 8 reportsDec 2024: 1 reportsJan 2025: 6 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 0 reportsMay 2025: 6 reportsJun 2025: 0 reportsJul 2025: 8 reportsAug 2025: 8 reportsSep 2025: 8 reportsOct 2025: 6 reportsNov 2025: 6 reportsDec 2025: 3 reportsJan 2026: 8 reportsFeb 2026: 6 reportsMar 2026: 4 reportsApr 2026: 16 reportsMay 2026: 4 reportsJun 2026: 8 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Protective Measures Problem317
  • Battery Problem169
  • Failure to Calibrate99
  • Device Displays Incorrect Message91
  • Failure to Charge79
  • Display or Visual Feedback Problem58
  • Inadequate User Interface44
  • Failure of Device to Self-Test39

Patient problems

  • No Clinical Signs, Symptoms or Conditions391
  • No Consequences Or Impact To Patient366
  • Insufficient Information50
  • No Patient Involvement49
  • Death5
  • Loss of consciousness5
  • Low Oxygen Saturation3
  • Choking2

Recent adverse event reports

The 25 most recent of 865 reports, in the reporter's own words as published by the FDA.

  • DeathAug 28, 2026No Audible Alarm

    It was reported to Resmed that a patient expired while using an Astral 100 device. It was reported that the patient became unresponsive while receiving ventilatory support from the device and no alarms were reportedly heard by the caregivers at the time of the event. The mother called for assistance, and family members initiated CPR prior to the arrival of emergency responders. Upon arrival of emergency responders, they assumed resuscitation efforts, and transported the patient by ambulance to a hospital, after which the patient was transferred to another hospital for a higher level of care. The patient was pronounced deceased at the hospital.

    Patient problems: Loss of consciousness

    MAUDE report 50505593 ↗
  • DeathAug 28, 2026No Audible Alarm

    Patient problems: Loss of consciousness

    MAUDE report 50505590 ↗
  • InjuryAug 19, 2026Use of Device Problem

    It was reported to Resmed that an Astral device displayed a 'low pressure' alarm while in use. Learn Circuit was performed and the device displayed 'without SpO2 monitoring'. The patient was placed on a different device. The patient's SpO2 was 88% and manual ventilation was provided.

    Patient problems: Low Oxygen Saturation

    MAUDE report 50332128 ↗
  • MalfunctionJun 10, 2026Failure to Power Up

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25462712 ↗
  • MalfunctionJun 10, 2026Failure to Power Up

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25462695 ↗
  • MalfunctionJun 8, 2026Battery Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED A BATTERY INOPERABLE ALARM AND FAILED TO CHARGE ITS INTERNAL BATTERY. THERE WAS NO PATIENT HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25423216 ↗
  • MalfunctionJun 5, 2026Battery Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED A BATTERY INOPERABLE ALARM AND FAILED TO CHARGE ITS INTERNAL BATTERY. THERE WAS NO PATIENT HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25417974 ↗
  • MalfunctionJun 5, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF 140) RELATED TO ALARM PRIORITY. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403504 ↗
  • MalfunctionJun 5, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF 140) RELATED TO ALARM PRIORITY. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403503 ↗
  • MalfunctionJun 1, 2026Power Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD A TOTAL POWER FAILURE DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25351668 ↗
  • MalfunctionJun 1, 2026Power Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD A TOTAL POWER FAILURE DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25351640 ↗
  • MalfunctionMay 14, 2026Power Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE ALARMED WHEN PLUGGED IN AND DID NOT POWER ON. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25178451 ↗
  • MalfunctionMay 14, 2026Power Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE ALARMED WHEN PLUGGED IN AND DID NOT POWER ON. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25178450 ↗
  • MalfunctionMay 7, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25115795 ↗
  • MalfunctionMay 7, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25115794 ↗
  • MalfunctionApr 30, 2026Inadequate User Interface

    DURING RESMED EVALUATION, AN ASTRAL DEVICE HAD AN INOPERABLE TOUCHSCREEN. THERE WAS NO REPORTED HARM OR INJURY AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25050759 ↗
  • MalfunctionApr 30, 2026Inadequate User Interface

    DURING RESMED EVALUATION, AN ASTRAL DEVICE HAD AN INOPERABLE TOUCHSCREEN. THERE WAS NO REPORTED HARM OR INJURY AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25050754 ↗
  • MalfunctionApr 22, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF82) RELATED TO THE ALARM SYSTEM. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24957898 ↗
  • MalfunctionApr 22, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF82) RELATED TO THE ALARM SYSTEM. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24957849 ↗
  • MalfunctionApr 14, 2026Battery Problem; Protective Measures Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24877352 ↗
  • MalfunctionApr 14, 2026Battery Problem; Protective Measures Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24877350 ↗
  • MalfunctionApr 9, 2026Battery Problem; Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED BATTERY ERROR MESSAGES. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24836181 ↗
  • MalfunctionApr 9, 2026Battery Problem; Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED BATTERY ERROR MESSAGES. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24836177 ↗
  • MalfunctionApr 9, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF147) RELATED TO THE MAIN BLOWER. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24835209 ↗
  • MalfunctionApr 9, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF147) RELATED TO THE MAIN BLOWER. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24835208 ↗

Other Continuous, Ventilator, Home Use devices

All Continuous, Ventilator, Home Use devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.