Astral 150Recalls & reported problems

Resmed Pty Ltd · Portable ventilator · Class 2

All models of Astral (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This ResMed Astral 150 is a portable ventilator for home use that helps people breathe. Its 2,363 reports often describe battery and charging problems, calibration failures and error messages, and mention 24 injuries and 17 deaths. One recall, started in June 2026 and open, concerns a leaking supercapacitor on the circuit board that may damage the charger chip and interrupt ventilation therapy.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,340
Deaths reported
17
Injuries reported
24
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2,340 adverse event reports that list this exact device (DI 00619498270033), received between Jun 30, 2016 and Jun 5, 2026 — classified as 17 deaths, 24 injuries, 2,299 malfunctions.

The most frequently reported specific device problem is “Protective Measures Problem” (760 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 22 reportsJan 2024: 12 reportsFeb 2024: 6 reportsMar 2024: 6 reportsApr 2024: 8 reportsMay 2024: 12 reportsJun 2024: 20 reportsJul 2024: 8 reportsAug 2024: 6 reportsSep 2024: 2 reportsOct 2024: 12 reportsNov 2024: 6 reportsDec 2024: 0 reportsJan 2025: 12 reportsFeb 2025: 0 reportsMar 2025: 8 reportsApr 2025: 4 reportsMay 2025: 6 reportsJun 2025: 10 reportsJul 2025: 20 reportsAug 2025: 12 reportsSep 2025: 12 reportsOct 2025: 30 reportsNov 2025: 22 reportsDec 2025: 8 reportsJan 2026: 28 reportsFeb 2026: 20 reportsMar 2026: 10 reportsApr 2026: 20 reportsMay 2026: 12 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202630
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2026Class IOpen, Classified

    Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.

    Z-2735-2026 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Protective Measures Problem760
  • Battery Problem434
  • Failure to Calibrate328
  • Device Displays Incorrect Message263
  • Inadequate User Interface132
  • Failure to Charge129
  • Failure of Device to Self-Test127
  • Display or Visual Feedback Problem113

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,189
  • No Consequences Or Impact To Patient860
  • Insufficient Information137
  • No Patient Involvement114
  • Low Oxygen Saturation10
  • Death5
  • Loss of consciousness4
  • Respiratory Failure4

Recent adverse event reports

The 25 most recent of 2,340 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2026Protective Measures Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF 179) RELATED TO A SUPER CAPACITOR ISSUE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403488 ↗
  • MalfunctionJun 5, 2026Protective Measures Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF 179) RELATED TO A SUPER CAPACITOR ISSUE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403477 ↗
  • MalfunctionMay 29, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF140) RELATED TO ALARM PRIORITY. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326571 ↗
  • MalfunctionMay 29, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF140) RELATED TO ALARM PRIORITY. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326564 ↗
  • MalfunctionMay 19, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25221359 ↗
  • MalfunctionMay 19, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25221353 ↗
  • MalfunctionMay 19, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO REPORT OF HARM OR INJURY AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219726 ↗
  • MalfunctionMay 19, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO REPORT OF HARM OR INJURY AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219725 ↗
  • MalfunctionMay 14, 2026Inadequate User Interface

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25178380 ↗
  • MalfunctionMay 14, 2026Inadequate User Interface

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25178379 ↗
  • MalfunctionMay 14, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL 150 DEVICE DISPLAYED AN ERROR MESSAGE (SF 82) RELATED TO THE HARDWARE ALARM CONTROLLER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25167110 ↗
  • MalfunctionMay 14, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL 150 DEVICE DISPLAYED AN ERROR MESSAGE (SF 82) RELATED TO THE HARDWARE ALARM CONTROLLER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25167063 ↗
  • MalfunctionMay 11, 2026Protective Measures Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF140) RELATED TO ALARM PRIORITY. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25144893 ↗
  • MalfunctionMay 11, 2026Protective Measures Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF140) RELATED TO ALARM PRIORITY. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25144892 ↗
  • MalfunctionApr 30, 2026Battery Problem; Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25050045 ↗
  • MalfunctionApr 30, 2026Battery Problem; Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25050024 ↗
  • MalfunctionApr 22, 2026Battery Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN INTERNAL BATTERY WITH A REDUCED LEVEL OF CAPACITY. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24961439 ↗
  • MalfunctionApr 22, 2026Battery Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN INTERNAL BATTERY WITH A REDUCED LEVEL OF CAPACITY. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24961437 ↗
  • MalfunctionApr 22, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24961359 ↗
  • MalfunctionApr 22, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24961358 ↗
  • MalfunctionApr 22, 2026Battery Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD A DEFECTIVE INTERNAL BATTERY. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24958049 ↗
  • MalfunctionApr 22, 2026Battery Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD A DEFECTIVE INTERNAL BATTERY. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24958028 ↗
  • MalfunctionApr 21, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF82) RELATED TO THE ALARM SYSTEM. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24947589 ↗
  • MalfunctionApr 21, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF82) RELATED TO THE ALARM SYSTEM. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24947587 ↗
  • MalfunctionApr 17, 2026Battery Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24913395 ↗

FDA clearance and classification

The FDA regulates product code NOU (continuous, ventilator, home use) under 21 CFR 868.5895 as a Class II device. 21 CFR 868.5895 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

FDA definition of this device type: This product code was needed for the home use indication. CBK is not a tracked device and NOU is a tracked device.

Common questions about the Astral 150

Has the Astral 150 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2735-2026 (Class I), began on Jun 25, 2026: Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Astral 150?

The FDA's MAUDE database lists 2,340 adverse event reports for this device, received between Jun 30, 2016 and Jun 5, 2026. In those reports, the most common specific device problems are “Protective Measures Problem” (760), “Battery Problem” (434) and “Failure to Calibrate” (328); the most common patient problems are “Low Oxygen Saturation” (10), “Death” (5) and “Loss of consciousness” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Astral 150?

17 reports are classified as deaths, 24 as injuries and 2,299 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Astral 150 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Astral 150, and is it still sold?

The FDA's device database lists Resmed Pty Ltd as the company responsible for it (brand Astral, model 150). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.