Astral 150Recalls & reported problems

Resmed Pty Ltd · Portable ventilator · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This ResMed Astral 150 is a portable ventilator for home use that helps people breathe. Its 2,363 reports often describe battery and charging problems, calibration failures and error messages, and mention 24 injuries and 17 deaths. One recall, started in June 2026 and open, concerns a leaking supercapacitor on the circuit board that may damage the charger chip and interrupt ventilation therapy.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,361
Deaths reported
17
Injuries reported
24
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2,361 adverse event reports that list this exact device (DI 00619498270033), received between Jun 30, 2016 and Sep 29, 2026 — classified as 17 deaths, 24 injuries, 2,320 malfunctions.

The most frequently reported specific device problem is “Protective Measures Problem” (771 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 32 reportsDec 2023: 22 reportsJan 2024: 12 reportsFeb 2024: 6 reportsMar 2024: 6 reportsApr 2024: 8 reportsMay 2024: 12 reportsJun 2024: 20 reportsJul 2024: 8 reportsAug 2024: 6 reportsSep 2024: 2 reportsOct 2024: 12 reportsNov 2024: 6 reportsDec 2024: 0 reportsJan 2025: 12 reportsFeb 2025: 0 reportsMar 2025: 8 reportsApr 2025: 4 reportsMay 2025: 6 reportsJun 2025: 10 reportsJul 2025: 20 reportsAug 2025: 12 reportsSep 2025: 12 reportsOct 2025: 30 reportsNov 2025: 22 reportsDec 2025: 8 reportsJan 2026: 28 reportsFeb 2026: 20 reportsMar 2026: 10 reportsApr 2026: 24 reportsMay 2026: 14 reportsJun 2026: 13 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports20242025202632
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2026Open, Classified

    Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.

    Z-2735-2026 · Root cause: Component design/selection · FDA record ↗

Most reported problems

Device problems

  • Protective Measures Problem771
  • Battery Problem436
  • Failure to Calibrate328
  • Device Displays Incorrect Message263
  • Inadequate User Interface136
  • Failure to Charge129
  • Failure of Device to Self-Test127
  • Display or Visual Feedback Problem113

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,210
  • No Consequences Or Impact To Patient860
  • Insufficient Information137
  • No Patient Involvement114
  • Low Oxygen Saturation10
  • Death5
  • Loss of consciousness4
  • Respiratory Failure4

Recent adverse event reports

The 25 most recent of 2,361 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Protective Measures Problem

    During Resmed evaluation, an Astral device displayed an error message (SF140) related to alarm priority. There was no harm or serious injury reported as a result of the incident.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927463 ↗
  • MalfunctionSep 29, 2026Protective Measures Problem

    During Resmed evaluation, an Astral device displayed an error message (SF140) related to alarm priority. There was no harm or serious injury reported as a result of the incident.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927462 ↗
  • MalfunctionSep 29, 2026Protective Measures Problem

    It was reported to Resmed that an Astral device displayed an error message (SF140) related to alarm priority. There was no harm or serious injury reported as a result of the incident.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927431 ↗
  • MalfunctionSep 29, 2026Protective Measures Problem

    It was reported to Resmed that an Astral device displayed an error message (SF140) related to alarm priority. There was no harm or serious injury reported as a result of the incident.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927430 ↗
  • MalfunctionJun 26, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO A SUPER CAPACITOR ISSUE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25641921 ↗
  • MalfunctionJun 26, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO A SUPER CAPACITOR ISSUE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25641777 ↗
  • MalfunctionJun 23, 2026Battery Problem; Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25603117 ↗
  • MalfunctionJun 23, 2026Battery Problem; Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED A BATTERY ERROR MESSAGE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25603104 ↗
  • MalfunctionJun 22, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. DURING EVALUATION, THE DEVICE DISPLAYED ERROR MESSAGE (SF140) RELATED TO ALARM PRIORITY AND ERROR MESSAGE (SF 82) RELATED TO THE HARDWARE ALARM CONTROLLER. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25588369 ↗
  • MalfunctionJun 5, 2026Protective Measures Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF 179) RELATED TO A SUPER CAPACITOR ISSUE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403488 ↗
  • MalfunctionJun 5, 2026Protective Measures Problem

    DURING RESMED EVALUATION, AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF 179) RELATED TO A SUPER CAPACITOR ISSUE. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403477 ↗
  • MalfunctionJun 5, 2026Complete Loss of Power

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD A POWER SHUT DOWN DURING FLOW CALIBRATION. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403442 ↗
  • MalfunctionJun 5, 2026Complete Loss of Power

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD A POWER SHUT DOWN DURING FLOW CALIBRATION. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403415 ↗
  • MalfunctionJun 4, 2026Complete Loss of Power

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE SHUT DOWN DURING FLOW CALIBRATION. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403074 ↗
  • MalfunctionJun 4, 2026Complete Loss of Power

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE SHUT DOWN DURING FLOW CALIBRATION. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403073 ↗
  • MalfunctionJun 4, 2026Power Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE WAS UNABLE TO POWER OFF. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25401479 ↗
  • MalfunctionJun 4, 2026Power Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE WAS UNABLE TO POWER OFF. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25401202 ↗
  • MalfunctionMay 29, 2026Inadequate User Interface

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326667 ↗
  • MalfunctionMay 29, 2026Inadequate User Interface

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE HAD AN UNRESPONSIVE TOUCHSCREEN. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326651 ↗
  • MalfunctionMay 29, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF140) RELATED TO ALARM PRIORITY. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326571 ↗
  • MalfunctionMay 29, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF140) RELATED TO ALARM PRIORITY. THERE WAS NO HARM OR INJURY REPORTED AS A RESULT OF THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326564 ↗
  • MalfunctionMay 19, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25221359 ↗
  • MalfunctionMay 19, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25221353 ↗
  • MalfunctionMay 19, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO REPORT OF HARM OR INJURY AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219726 ↗
  • MalfunctionMay 19, 2026Protective Measures Problem

    IT WAS REPORTED TO RESMED THAT AN ASTRAL DEVICE DISPLAYED AN ERROR MESSAGE (SF179) RELATED TO THE SUPER CAPACITOR. THERE WAS NO REPORT OF HARM OR INJURY AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219725 ↗

Other Continuous, Ventilator, Home Use devices

All Continuous, Ventilator, Home Use devices →