Skip to content

AirTouch F20 63014Recalls & reported problems

Resmed Pty Ltd · AirTouch F20 LGE Sys - MULTI · Class 2

All models of AirTouch F20 (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Nov 20, 2023Open, Classified

    Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact

    Z-0538-2024 · Root cause: Device Design · Recall details · FDA record ↗

Other Ventilator, Non-Continuous (Respirator) devices

All Ventilator, Non-Continuous (Respirator) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page