AIRFIT F30I 63332Recalls & reported problems
Resmed Pty Ltd · AIRFIT F30I SYS - M/STD - AMER · Class 2
All models of AIRFIT F30I (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00619498633326), received between Nov 8, 2024 and Nov 8, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Electromagnetic Interference” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Nov 20, 2023Class IOpen, Classified
Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact
Z-0542-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Electromagnetic Interference2
Patient problems
- Confusion/ Disorientation2
- Dizziness2
- Fall2
- Irritability2
- Loss of consciousness2
- Memory Loss/Impairment2
- Psychosis2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 8, 2024Electromagnetic Interference; Electromagnetic Interference
IT WAS REPORTED TO RESMED THAT THE MAGNETS OF AN AIRFIT F30I MASK ALLEGEDLY CAUSED THE SETTINGS OF A PATIENT'S ADJUSTABLE MAGNETIC VENTRICULOPERITONEAL SHUNT TO SHIFT RESULTING IN TOO MUCH FLOW OF CEREBROSPINAL FLUID. PATIENT REQUIRED TRANSPORT TO THE HOSPITAL. PATIENT HAS SWITCHED TO A NON-MAGNETIC MASK. PATIENT DID NOT HAVE SYMPTOMS PRIOR TO USE OF THE MASK.
Patient problems: Confusion/ Disorientation; Dizziness; Fall; Irritability; Loss of consciousness; Memory Loss/Impairment; Psychosis
MAUDE report 20645928 ↗InjuryNov 8, 2024Electromagnetic Interference; Electromagnetic Interference
IT WAS REPORTED TO RESMED THAT THE MAGNETS OF AN AIRFIT F30I MASK ALLEGEDLY CAUSED THE SETTINGS OF A PATIENT'S ADJUSTABLE MAGNETIC VENTRICULOPERITONEAL SHUNT TO SHIFT RESULTING IN TOO MUCH FLOW OF CEREBROSPINAL FLUID. PATIENT REQUIRED TRANSPORT TO THE HOSPITAL. PATIENT HAS SWITCHED TO A NON-MAGNETIC MASK. PATIENT DID NOT HAVE SYMPTOMS PRIOR TO USE OF THE MASK.
Patient problems: Confusion/ Disorientation; Dizziness; Fall; Irritability; Loss of consciousness; Memory Loss/Impairment; Psychosis
MAUDE report 20644726 ↗
Other Ventilator, Non-Continuous (Respirator) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- DreamStation DreamStation Auto CPAP with Humidifier, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)44 deaths5,793 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page