7D Surgical Pedicle Probe (Lumbar) 11-0013Recalls & reported problems
7D Surgical ULC · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00628341520027), received between Mar 10, 2025 and Aug 20, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- Foreign Body In Patient1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 20, 2026Fracture
On 22 JUL 2026 Orthofix was made aware of an event involving the 7D Surgical Pedicle Probe (Lumbar) fracturing during a lumbar fusion revision procedure. The reporter stated that while advancing the probe through highly sclerotic sacral bone, the shaft of the probe would not rotate with the handle. Upon removal, it was observed that the distal end had twisted, and the tip had fractured off. The fragment remained within the bone.No patient injury was reported. The surgeon did not attempt to retrieve the fragment.
Patient problems: Foreign Body In Patient
MAUDE report 50358999 ↗MalfunctionAug 28, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ON 30 JUL 2025, ORTHOFIX WAS MADE AWARE OF THE FOLLOWING EVENT DURING NAVIGATION USING 7D SURGICAL FLASH IMAGING. PER THE REPORTER, THE SURGEON NAVIGATED THE RIGHT T10 AND RIGHT T9 SCREWS USING THE 7D SURGICAL PEDICLE PROBE (LUMBAR). THE SURGEON COMPLAINED OF THE PROBES INACCURACY AND SWITCHED TO ANOTHER PROBE TO COMPLETE THE REMAINING SCREW NAVIGATION. AFTER ALL SCREWS WERE IMPLANTED, AN X-RAY WAS TAKEN, REVEALING THAT THE T10 AND T9 SCREWS WERE INCORRECTLY POSITIONED. THE TWO SCREWS WERE REVISED, RESULTING IN A 45 MIN SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22913121 ↗MalfunctionMar 10, 2025Fracture; Fracture
THE CUSTOMER REPORTED THAT WHILE CANNULATING THE PATIENT'S PELVIS, THE SURGEON FELT A SNAP. WHEN HE PULLED THE INSTRUMENT OUT, HE FOUND THAT THE TIP OF THE PROBE HAD SNAPPED OFF. THE SURGEON PROCEEDED WITH NAVIGATION. APPROXIMATELY 10 MM OF THE INSTRUMENT TIP WAS LEFT 110MM DEEP IN THE PATIENT'S PELVIS. NO PATIENT INJURY WAS REPORTED. NO REVISION SURGERY IS NECESSARY.
Patient problems: Insufficient Information
MAUDE report 21564523 ↗
Other Orthopedic Stereotaxic Instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page