Skip to content

7D Surgical Pedicle Probe (Lumbar) 11-0013Recalls & reported problems

7D Surgical ULC · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00628341520027), received between Mar 10, 2025 and Aug 20, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 20, 2026Fracture

    On 22 JUL 2026 Orthofix was made aware of an event involving the 7D Surgical Pedicle Probe (Lumbar) fracturing during a lumbar fusion revision procedure. The reporter stated that while advancing the probe through highly sclerotic sacral bone, the shaft of the probe would not rotate with the handle. Upon removal, it was observed that the distal end had twisted, and the tip had fractured off. The fragment remained within the bone.No patient injury was reported. The surgeon did not attempt to retrieve the fragment.

    Patient problems: Foreign Body In Patient

    MAUDE report 50358999 ↗
  • MalfunctionAug 28, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ON 30 JUL 2025, ORTHOFIX WAS MADE AWARE OF THE FOLLOWING EVENT DURING NAVIGATION USING 7D SURGICAL FLASH IMAGING. PER THE REPORTER, THE SURGEON NAVIGATED THE RIGHT T10 AND RIGHT T9 SCREWS USING THE 7D SURGICAL PEDICLE PROBE (LUMBAR). THE SURGEON COMPLAINED OF THE PROBES INACCURACY AND SWITCHED TO ANOTHER PROBE TO COMPLETE THE REMAINING SCREW NAVIGATION. AFTER ALL SCREWS WERE IMPLANTED, AN X-RAY WAS TAKEN, REVEALING THAT THE T10 AND T9 SCREWS WERE INCORRECTLY POSITIONED. THE TWO SCREWS WERE REVISED, RESULTING IN A 45 MIN SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22913121 ↗
  • MalfunctionMar 10, 2025Fracture; Fracture

    THE CUSTOMER REPORTED THAT WHILE CANNULATING THE PATIENT'S PELVIS, THE SURGEON FELT A SNAP. WHEN HE PULLED THE INSTRUMENT OUT, HE FOUND THAT THE TIP OF THE PROBE HAD SNAPPED OFF. THE SURGEON PROCEEDED WITH NAVIGATION. APPROXIMATELY 10 MM OF THE INSTRUMENT TIP WAS LEFT 110MM DEEP IN THE PATIENT'S PELVIS. NO PATIENT INJURY WAS REPORTED. NO REVISION SURGERY IS NECESSARY.

    Patient problems: Insufficient Information

    MAUDE report 21564523 ↗

Other Orthopedic Stereotaxic Instrument devices

All Orthopedic Stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page