Dimension Clinical Chemistry System A1C Kit 10720883Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · A1C Reagent Kit · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00630414006659), received between Mar 2, 2020 and May 15, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “High Test Results” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results6
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2026High Test Results
THE CUSTOMER REPORTED OBSERVATION OF AN ELEVATED DIMENSION HEMOGLOBIN A1C (HBA1C) RESULT WHICH WAS DISCORDANT RELATIVE TO REPEAT TESTING WHEN USING LOT GA6344. THE CUSTOMER USED AN INVALID CALIBRATION WHICH RESULTED IN ABOVE ASSAY RANGE RESULT ACCOMPANIED BY AN INSTRUMENT ERROR. THE OPERATOR¿S GUIDE/INSTRUCTIONS FOR USE (IFU) INSTRUCTIONS WAS NOT FOLLOWED, AS THE CUSTOMER DISREGARDED THE ERROR MESSAGE AND DISCORDANT HBA1C PATIENT SAMPLE RESULT WAS REPORTED FOR ONE (1) PATIENT SAMPLE. THE RESULT WAS NOT QUESTIONED BY THE PHYSICIAN. THE PATIENT SAMPLE WAS SUBSEQUENTLY RETESTED ON THE ORIGINAL ANALYZER WHICH YIELDED A LOWER RESULT. THE LOWER RESULT WAS CONSIDERED CORRECT AND WAS ISSUED ON CORRECTED REPORT TO THE PHYSICIAN. THERE WERE NO ALLEGATIONS OF PATIENT HARM DUE TO THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25191966 ↗MalfunctionMay 15, 2026High Test Results
THE CUSTOMER REPORTED OBSERVATION OF ELEVATED DIMENSION HEMOGLOBIN A1C (HBA1C) RESULTS WHICH WERE DISCORDANT RELATED TO REPEAT TESTING WHEN USING LOT: GA6344. THE CUSTOMER USED AN INVALID CALIBRATION WHICH RESULTED IN ABOVE ASSAY RANGE RESULTS ACCOMPANIED BY AN INSTRUMENT ERROR. THE OPERATOR¿S GUIDE/INSTRUCTIONS FOR USE (IFU) INSTRUCTIONS WERE NOT FOLLOWED, AS THE CUSTOMER DISREGARDED THE ERROR MESSAGES AND DID NOT PERFORM THE REQUIRED MANUAL DILUTION PROCEDURE PER THE IFU. DISCORDANT HBA1C PATIENT SAMPLE RESULTS WERE REPORTED FOR TWO (2) PATIENT SAMPLES. FOR ONE (1) PATIENT SAMPLE WITH SID: (B)(6), THE DISCORDANT RESULT WAS QUESTIONED BY THE PHYSICIAN WHICH CONTRIBUTED TO THE CANCELLATION OF A SCHEDULED SURGERY. THE PATIENT WAS RETESTED AT AN ALTERNATE LABORATORY WHICH YIELDED A LOWER (NORMAL) RESULT. THE INITIAL PATIENT SAMPLE WAS SUBSEQUENTLY RETESTED ON THE ORIGINAL ANALYZER USING THE INVALID CALIBRATION AND CONTINUED TO YIELD A HIGHER RESULT. FOR THE OTHER PATIENT SAMPLE, THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN BUT WAS NOT QUESTIONED. THE PATIENT SAMPLE WAS SUBSEQUENTLY RETESTED ON THE ORIGINAL ANALYZER WHICH YIELDED A LOWER RESULT. THE LOWER RESULT WAS CONSIDERED COR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25187212 ↗MalfunctionApr 26, 2026High Test Results
THE CUSTOMER REPORTED THAT AN ERRONEOUSLY ELEVATED HEMOGLOBIN A1C ASSAY (A1C) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION EXL WITH LM INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPROCESSED ON THE ORIGINAL SYSTEM. THE REPROCESSED RESULT WAS LOWER THAN THE ERRONEOUS RESULT. THE REPROCESSED RESULT WAS REPORTED AS THE CORRECT RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED HEMOGLOBIN A1C ASSAY (A1C) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24993495 ↗MalfunctionApr 26, 2026High Test Results
THE CUSTOMER REPORTED THAT AN ERRONEOUSLY ELEVATED HEMOGLOBIN A1C ASSAY (A1C) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION EXL WITH LM INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS SENT TO AN ALTERNATE LAB AND REPROCESSED ON A NON-SIEMENS SYSTEM. THE REPROCESSED RESULT WAS LOWER THAN THE ERRONEOUS RESULT. THE REPROCESSED RESULT WAS REPORTED AS THE CORRECT RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED HEMOGLOBIN A1C ASSAY (A1C) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24993489 ↗MalfunctionMar 2, 2020High Test Results; High Test Results
A FALSELY ELEVATED HEMOGLOBIN A1C (A1C) RESULT WAS OBTAINED ON A PATIENT WHOLE BLOOD SAMPLE ON A DIMENSION EXL WITH LM INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPEATED ON THE SAME INSTRUMENT AFTER RECALIBRATION OF THE ASSAY AND DILUTION OF THE SAMPLE. A LOWER RESULT WAS OBTAINED AND A CORRECTED RESULT WAS ISSUED. THE ACCOUNT STATED THAT A CARDIAC CATHETERIZATION WAS DELAYED. THERE ARE NO KNOWN REPORTS ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED AIC RESULT OR THE DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9774734 ↗MalfunctionMar 2, 2020High Test Results; High Test Results
A FALSELY ELEVATED HEMOGLOBIN A1C (A1C) RESULT WAS OBTAINED ON A PATIENT WHOLE BLOOD SAMPLE ON A DIMENSION EXL WITH LM INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPEATED ON THE SAME INSTRUMENT AFTER RECALIBRATION OF THE ASSAY AND DILUTION OF THE SAMPLE. A LOWER RESULT WAS OBTAINED AND A CORRECTED RESULT WAS ISSUED. THE ACCOUNT STATED THAT A CARDIAC CATHETERIZATION WAS DELAYED. THERE ARE NO KNOWN REPORTS ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED AIC RESULT OR THE DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9774733 ↗
Other Hemoglobin A1c test system devices
- VITROS 6842905Ortho-Clinical Diagnostics, Inc. · Hemoglobin A1c test system1 recall21 reports
- Alinity 08P4320Abbott GmbH · Hemoglobin A1c test system1 recall7 reports
- ARCHITECT 4P52-20Abbott GmbH · Hemoglobin A1c test system1 recall2 reports
- Tina-quant Hemoglobin A1cDx Gen.3 08056668190Roche Diagnostics GmbH · Hemoglobin A1c test system28 reports
- Tina-quant Hemoglobin A1cDx Gen.3 07559674190Roche Diagnostics GmbH · Hemoglobin A1c test system22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page