Alinity 08P4320Recalls & reported problems
Abbott GmbH · Alinity c Hemoglobin A1c Reagent Kit 1300 Tests · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00380740135591), received between Jun 28, 2022 and Aug 2, 2022 — classified as 7 malfunctions.
The most frequently reported specific device problem is “High Test Results” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jun 9, 2022Open, Classified
There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
Z-1455-2022 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- High Test Results7
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 2, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C HEMOGLOBIN A1C RESULT FOR ONE PATIENT SAMPLE THAT WAS QUESTIONED BY THE PHYSICIAN. THE SAMPLE WAS REPEATED, AND THE RESULT WAS NORMAL. THE FOLLOWING DATA WAS PROVIDED (NORMAL RANGE: < 6.5%): SID (B)(6) INITIAL RESULT = > 14%, REPEAT RESULT = 6.3%. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15145045 ↗MalfunctionJul 8, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY C HEMOGLOBIN A1C RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: SAMPLE ID (B)(6) RESULT WAS 129.5 MMOL/MOL, WHICH WAS REPORTED TO THE PHYSICIAN. THE PATIENT¿S PREVIOUS RESULT, IN JANUARY, WAS 42 MMOL/MOL. BASED ON THE ELEVATED HEMOGLOBIN A1C RESULT, THE PATIENT HAS BEEN TREATED; HOWEVER, NO OTHER INFORMATION WAS PROVIDED. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14955385 ↗MalfunctionJul 8, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY C HEMOGLOBIN A1C RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT WAS ¿UNDETECTABLE¿, REPEATED, IN THE MAIN LAB, WAS 13.8%. THE SAMPLE WAS REPEATED, ON ANOTHER PLATFORM, AND THE RESULT WAS 4.6%, WHICH AGREED WITH THE GLUCOSE RESULT OF 5.2 MMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14955343 ↗MalfunctionJun 28, 2022High Test Results; High Test Results
THE CUSTOMER STATED THAT A FALSELY ELEVATED ALINITY C HEMOGLOBIN A1C RESULT WAS GENERATED FOR A PATIENT ON (B)(6) 2022. THE FOLLOWING DATA WAS PROVIDED. SAMPLE NUMBER: (B)(4). HBA1CWB: 63.7512 UMOL/L. THBWB: 83.3212 UMOL/L. %A1CWB: >14.00%. THE PATIENT HAS A KNOWN HISTORY OF DIABETES AND HAS AN INSULIN PUMP. THE PHYSICIAN QUESTIONED THE RESULT BECAUSE THE PATIENT HAS A HISTORY OF STABLE %A1CWB VALUES. IT WAS SUSPECTED THAT THE PUMP WAS POSSIBLY NOT WORKING CORRECTLY SINCE THE ELEVATED RESULT WAS NOT EXPECTED. THE SAME SAMPLE WAS RETESTED ON (B)(6) 2022 WITH THE FOLLOWING RESULTS. HBA1CWB: 236.1069 UMOL/L. THBWB: 4179.2451 UMOL/L. %A1CWB: 7.32%. REVIEW OF INSTRUMENT LOGS SHOWED A DIFFERENT ASPIRATION CURVE FOR THE FALSELY ELEVATED RESULT INDICATING A POTENTIAL SAMPLE ISSUE SUCH AS A CLOT. THE CUSTOMER WILL BE IMPLEMENTING A NEW RETEST RULE FOR EVERY A1CWB ABOVE 11.3%. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14844031 ↗MalfunctionJun 28, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED DISCREPANT ALINITY C HEMOGLOBIN A1C RESULTS ON MULTIPLE PATIENT SAMPLES THAT WERE NORMAL UPON REPEAT ON THE ALINITY ANALYZER AND VARIANT (HPLC). THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2022. SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 5.70%, VARIANT (HPLC) = 6.20%, SID (B)(6) INITIAL %HBA1C RESULT = > 14.0 %, REPEAT = 7.90%, VARIANT (HPLC) = 9.30%, SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 7.40%, VARIANT (HPLC) = 4.50%, SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 6.90%, VARIANT (HPLC) = 7.10%. (B)(6) 2022. SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = > 14%, VARIANT (HPLC) = 4.20%, SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 5.60%, VARIANT (HPLC) = 6.00%, SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 6.20%, VARIANT (HPLC) = 6.20%. (B)(6) 2022 . SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 6.00%, VARIANT (HPLC) = 6.40%, SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 8.00%, VARIANT (HPLC) = 7.80%, SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 7.30%, VARIANT (HPLC) = 7.40%, SID (B)(6) INITIAL %HBA1C RESULT = > 14.0%, REPEAT = 6.60%, VARIANT (HPLC) = 6.60%, SID (B)(6) INITIAL %HBA1C RESULT =
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14843108 ↗MalfunctionJun 28, 2022High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED HEMOGLOBIN A1C RESULTS GENERATED ON THE ALINITY C ANALYZER ON MULTIPLE PATIENTS. EXAMPLE SIDS PROVIDED: (B)(6). THE CUSTOMER OBSERVES > 129 MMOL/ML AND UPON REPEAT THE RESULTS ARE NORMAL. RESULTS PROVIDED: SID (B)(6). FIRST RUN >129.51 MMOL/MOL / >129.51 MMOL/MOL. 1RST RERUN 38.89 MMOL/MOL / 56.33 MMOL/MOL. 2ND RERUN 38.44 MMOL/MOL / 56.23 MMOL/MOL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14842667 ↗MalfunctionJun 28, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY C HEMOGLOBIN A1C RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S NORMAL REFERENCE RANGE IS <5.7%): SAMPLE ID (B)(6) INITIAL HBA1CWB RESULT WAS 53.5221, THBWB RESULT WAS 461.0438, RESULT WAS %A1CWB 12.8%, REPEAT HBA1CWB WAS 94.5087, THBWB WAS 3,240.5947, %A1CWB WAS 4.8%. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14842618 ↗
Other Hemoglobin A1c test system devices
- VITROS 6842905Ortho-Clinical Diagnostics, Inc. · Hemoglobin A1c test system1 recall21 reports
- ARCHITECT 4P52-20Abbott GmbH · Hemoglobin A1c test system1 recall2 reports
- Tina-quant Hemoglobin A1cDx Gen.3 08056668190Roche Diagnostics GmbH · Hemoglobin A1c test system28 reports
- Tina-quant Hemoglobin A1cDx Gen.3 07559674190Roche Diagnostics GmbH · Hemoglobin A1c test system22 reports
- Dimension Clinical Chemistry System A1C Kit 10720883Siemens Healthcare Diagnostics Inc. · Hemoglobin A1c test system6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.