Atellica CH TACR 11097542Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Tacrolimus (TACR) Atellica CH - RGT 400T · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00630414238340), received between Dec 14, 2023 and Sep 15, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K060385); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results3
- High Test Results1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Low Test Results
The customer reported that they obtained erroneously depressed Tacrolimus (TACR) results on two patient samples on an Atellica CH 930 Analyzer.The initial results were reported to the physician(s) and were questioned. The samples were repeated on an alternate Atellica CH 930 Analyzer, and the results recovered higher than the initial results and were considered correct. Corrected reports were issued with the repeat results. There are no known reports of patient intervention or adverse health consequences due to the erroneously depressed TACR results.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50771391 ↗MalfunctionFeb 24, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED TO SIEMENS THAT THEY OBTAINED DISCORDANT TACROLIMUS (TACR) RESULTS ON MULTIPLE PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN. THE SAME SAMPLES WERE REPROCESSED ON THE SAME ATELLICA CH 930 ANALYZER. THE REPROCESSED RESULTS OBTAINED WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THE CUSTOMER DECLINED TO PROVIDE THE PATIENT SAMPLE DATA. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT TACROLIMUS (TACR) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21444007 ↗MalfunctionDec 14, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
DISCORDANT ELEVATED TACROLIMUS (TACR) RESULTS WERE OBTAINED ON SAMPLES FROM TWO PATIENTS ON AN ATELLICA CH 930 INSTRUMENT (S/N(B)(6)). THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULTS. THE SAMPLES WERE REPEATED ON THE SAME ATELLICA CH 930 INSTRUMENT. THE REPEAT RESULTS WERE LOWER THAN THE INITIAL RESULTS. CORRECTED REPORT(S) WERE ISSUED. THE PATIENTS WERE BOTH ORGAN TRANSPLANT RECIPIENTS. THERE ARE NO KNOWN REPORTS ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ELEVATED TACROLIMUS (TACR) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18325509 ↗MalfunctionDec 14, 2023Low Test Results; Low Test Results
THE CUSTOMER OBTAINED AN ABOVE ASSAY RANGE RESULT FOR TACROLIMUS (TACR) ON A PATIENT SAMPLE WHEN PROCESSED ON AN ATELLICA CH 930 INSTRUMENT (S/N (B)(6)). THE PATIENT SAMPLE WAS DILUTED AND REPEATED. THE RESULTS WERE LOWER AND WITHIN THE ASSAY RANGE. THE LOWER DILUTED RESULT WAS REPORTED TO THE PHYSICIAN(S) AND WAS QUESTIONED. A NEW SAMPLE FROM THE SAME PATIENT WAS TESTED AND THE RESULT WAS HIGHER AND CONSIDERED TOXIC. THE INITIAL SAMPLE WAS REPEATED AND THE NEAT RESULT WAS SIMILAR TO THE NEW SAMPLE RESULT. THE PATIENT IS AN ORGAN TRANSPLANT RECIPIENT AND WAS GIVEN TACROLIMUS TREATMENT. THE TRANSPLANT TIME FRAME WAS NOT PROVIDED FOR THE PATIENT. THERE ARE NO KNOWN REPORTS ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT TACROLIMUS (TACR) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18325390 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K060385. One submission often covers many models, so check the product and lot details in each record.
- Z-1661-2018 · Jun 29, 2017 · TerminatedSyva Emit 2000 Tacrolimus Assay (Product Number: 8R019UL/10445397) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)
Other ENZYME IMMUNOASSAY, TRACROLIMUS devices
- Dimension® Flex® reagent cartridge TAC 10700795Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, TRACROLIMUS2 recalls121 reports
- Syva® Emit® 2000 Tacrolimus Assay 10445397Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, TRACROLIMUS43 reports
- ARCHITECT 1L77-25Abbott Ireland Diagnostics · ENZYME IMMUNOASSAY, TRACROLIMUS9 reports
- Alinity 09P4220Abbott Ireland Diagnostics · ENZYME IMMUNOASSAY, TRACROLIMUS6 reports
- Elecsys Tacrolimus 07251254190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, TRACROLIMUS6 reports
- Elecsys Tacrolimus 05889057160Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, TRACROLIMUS3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.