Elecsys Tacrolimus 05889057160Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630947102), received between Feb 22, 2023 and Oct 2, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Incorrect Measurement1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2023High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S WHOLE BLOOD SAMPLES TESTED WITH ELECSYS TACROLIMUS ASSAY ON A COBAS E411 IMMUNOASSAY ANALYZER. ON (B)(6) 2023: INITIAL RESULT: >30 NG/ML. 1ST REPEAT RESULT: 9.1 NG/ML. 2ND REPEAT RESULT: 10.2 NG/ML (TESTED ON A DIFFERENT E411). 3RD REPEAT RESULT: 9.7 NG/ML (TESTED ON A DIFFERENT E411). ON (B)(6) 2023: INITIAL RESULT: 9.4 NG/ML. 1ST REPEAT RESULT: 5.6 NG/ML (NEW TREATED SAMPLE). 2ND REPEAT RESULT: 6.1 NG/ML (TESTED ON A DIFFERENT E411). THE CUSTOMER NOTICED THAT THE RESULTS DID NOT MATCH THE PATIENT'S PREVIOUS RESULTS. NO ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY AND THE SAMPLES WERE REPEATED. ALL THE REPEAT RESULTS WERE DEEMED TO BE CORRECT. THE REPEAT RESULTS OF 9.1 NG/ML AND 5.6 NG/ML WERE REPORTED OUTSIDE THE LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17850332 ↗
  • MalfunctionJul 20, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS TACROLIMUS ASSAY RESULTS FROM MULTIPLE PATIENT SAMPLES TESTED ON THE COBAS E411 RACK. THE INITIAL RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE REPORTER STATED THAT THEY HAVE AN ONGOING ISSUE WITH THE TACROLIMUS ASSAY. THEY ARE RUNNING THE PATIENT SAMPLES ON BOTH OF THEIR E 411 ANALYZERS. PLEASE REFER TO THE ATTACHMENT PT-83173 IN THE MEDWATCH FOR THE TABLE CONTAINING EXAMPLES OF QUESTIONABLE RESULTS AND THE RESULTS REPORTED OUTSIDE OF THE LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17363151 ↗
  • MalfunctionFeb 22, 2023Incorrect Measurement; High Test Results; Incorrect Measurement; High Test Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ELECSYS TACROLIMUS ASSAY ON COBAS E 411 ANALYZERS (RACK SYSTEM). THE SAMPLE WAS INITIALLY TESTED ON COBAS E 411, SERIAL NUMBER: (B)(4), AND RESULTED IN A TACROLIMUS VALUE OF 30 NG/ML, ACCOMPANIED BY A DATA FLAG. THE SAMPLE WAS REPEATED ON COBAS E 411, SERIAL NUMBER: (B)(4), AND RESULTED IN A TACROLIMUS VALUE OF 10.89NG/ML. IT IS UNKNOWN WHICH RESULT WAS DEEMED CORRECT. NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE OF THE LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16420595 ↗