Elecsys Tacrolimus 05889057160Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630947102), received between Feb 22, 2023 and Oct 2, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Incorrect Measurement1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 2, 2023High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S WHOLE BLOOD SAMPLES TESTED WITH ELECSYS TACROLIMUS ASSAY ON A COBAS E411 IMMUNOASSAY ANALYZER. ON (B)(6) 2023: INITIAL RESULT: >30 NG/ML. 1ST REPEAT RESULT: 9.1 NG/ML. 2ND REPEAT RESULT: 10.2 NG/ML (TESTED ON A DIFFERENT E411). 3RD REPEAT RESULT: 9.7 NG/ML (TESTED ON A DIFFERENT E411). ON (B)(6) 2023: INITIAL RESULT: 9.4 NG/ML. 1ST REPEAT RESULT: 5.6 NG/ML (NEW TREATED SAMPLE). 2ND REPEAT RESULT: 6.1 NG/ML (TESTED ON A DIFFERENT E411). THE CUSTOMER NOTICED THAT THE RESULTS DID NOT MATCH THE PATIENT'S PREVIOUS RESULTS. NO ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY AND THE SAMPLES WERE REPEATED. ALL THE REPEAT RESULTS WERE DEEMED TO BE CORRECT. THE REPEAT RESULTS OF 9.1 NG/ML AND 5.6 NG/ML WERE REPORTED OUTSIDE THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17850332 ↗MalfunctionJul 20, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS TACROLIMUS ASSAY RESULTS FROM MULTIPLE PATIENT SAMPLES TESTED ON THE COBAS E411 RACK. THE INITIAL RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE REPORTER STATED THAT THEY HAVE AN ONGOING ISSUE WITH THE TACROLIMUS ASSAY. THEY ARE RUNNING THE PATIENT SAMPLES ON BOTH OF THEIR E 411 ANALYZERS. PLEASE REFER TO THE ATTACHMENT PT-83173 IN THE MEDWATCH FOR THE TABLE CONTAINING EXAMPLES OF QUESTIONABLE RESULTS AND THE RESULTS REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17363151 ↗MalfunctionFeb 22, 2023Incorrect Measurement; High Test Results; Incorrect Measurement; High Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ELECSYS TACROLIMUS ASSAY ON COBAS E 411 ANALYZERS (RACK SYSTEM). THE SAMPLE WAS INITIALLY TESTED ON COBAS E 411, SERIAL NUMBER: (B)(4), AND RESULTED IN A TACROLIMUS VALUE OF 30 NG/ML, ACCOMPANIED BY A DATA FLAG. THE SAMPLE WAS REPEATED ON COBAS E 411, SERIAL NUMBER: (B)(4), AND RESULTED IN A TACROLIMUS VALUE OF 10.89NG/ML. IT IS UNKNOWN WHICH RESULT WAS DEEMED CORRECT. NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16420595 ↗
Other ENZYME IMMUNOASSAY, TRACROLIMUS devices
- Dimension® Flex® reagent cartridge TAC 10700795Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, TRACROLIMUS2 recalls121 reports
- Syva® Emit® 2000 Tacrolimus Assay 10445397Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, TRACROLIMUS43 reports
- ARCHITECT 1L77-25Abbott Ireland Diagnostics · ENZYME IMMUNOASSAY, TRACROLIMUS9 reports
- Alinity 09P4220Abbott Ireland Diagnostics · ENZYME IMMUNOASSAY, TRACROLIMUS6 reports
- Elecsys Tacrolimus 07251254190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, TRACROLIMUS6 reports
- Atellica CH TACR 11097542Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, TRACROLIMUS4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page