ADVIA Centaur® FOL 10325366Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Folic Acid - CENTAUR - RGT - 500 tests · Class 2
All models of ADVIA Centaur FOL (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00630414450940), received between Aug 11, 2017 and Aug 11, 2017 — classified as 1 malfunction.
1 FDA recall record name this device's identifier, 1 still open.
13 other recalls cover devices cleared under the same FDA submission (K010050); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jan 9, 2023Class IIOpen, Classified
Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
Z-1153-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DISCORDANT ADVIA CENTAUR XP FOLATE RESULTS WERE OBTAINED ON A PATIENT SAMPLE BETWEEN NEAT AND DILUTION, THE INITIAL RESULT WAS GREATER THAN 24 NG/ML. THE SAMPLE WAS DILUTED AND TESTED MULTIPLE TIMES. THE RESULTS WERE LOWER WITH DILUTION. THE SAMPLE WAS TESTED AGAIN DILUTED ON A SECOND ADVIA CENTAUR XP AND THE RESULT WAS LOWER THAN THE NEAT RESULT. THE DILUTED RESULTS WERE CONSIDERED CORRECT. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FOLATE RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6789785 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K010050. One submission often covers many models, so check the product and lot details in each record.
- Z-1152-2023 · Jan 9, 2023 · Open, ClassifiedADVIA Centaur Folate 100 test kit-For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 1031030
- Z-1154-2023 · Jan 9, 2023 · Open, ClassifiedADVIA Centaur Folate 500 test kit (REF)-For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 1
- Z-0829-2020 · Jul 22, 2019 · TerminatedSiemens ADVIA Centaur Folate 100 test kit-for IVD of folate in serum or red blood cells SMN:10310308
- Z-0830-2020 · Jul 22, 2019 · TerminatedSiemens ADVIA Centaur Folate (500 Test Kit)-for IVD of folate in serum or red blood cells SMN: 10325366
- Z-0831-2020 · Jul 22, 2019 · TerminatedSiemens ADVIA Centaur Folate (500 Test Kit Reference)- for IVD of folate in serum or red blood cell SMN: 10331250
- Z-1689-2018 · Jan 2, 2018 · TerminatedADVIA Centaur Folate (FOL)
- Z-2739-2015 · Jul 27, 2015 · TerminatedADVIA Centaur Folate Ascorbic Acid/Ascorbic Acid Diluent; SMN 10316877. Used to prepare the Red Blood Cell hemolysate for use with the ADVIA Centaur systems folate assay.
- Z-1564-2014 · Mar 25, 2014 · TerminatedSiemens ADVIA Centaur Folate Cat No. 06891541 SMN 10325366 (500 tests) Cat No. 06367974 SMN 10310308 (100 tests) Cat No. 00203473 SMN 10331250 (Ref) (500 tests) Cat No. 09132781 S
- Z-2170-2013 · Jul 25, 2013 · TerminatedADVIA Centaur Folate (500 Tests) SMN 10325366, Cat No. 06891541 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
- Z-2171-2013 · Jul 25, 2013 · TerminatedADVIA Centaur Folate (500 tests), REF SMN 10331250, Cat No. 00203473 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
- Z-2172-2013 · Jul 25, 2013 · TerminatedADVIA Centaur Folate (2500 tests), REF SMN 10340209, Cat No. 09132781 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
- Z-2169-2013 · Jul 25, 2013 · TerminatedADVIA Centaur Folate (100 Tests), SMN 10310308, Cat No. 06367974 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
- Z-0511-2013 · Nov 14, 2012 · TerminatedSiemens Healthcare ADVIA Centaur Folate (FOL) 500 Test 06891541 100 Test 06367974 Ref Kit (US only) 09132781 Ref Kit (US only) 00203473 Intended Use For
Other ACID, FOLIC, RADIOIMMUNOASSAY devices
- Atellica™ IM Fol 10995572Siemens Healthcare Diagnostics Inc. · ACID, FOLIC, RADIOIMMUNOASSAY1 recall7 reports
- ADVIA Centaur® FOL 10310308Siemens Healthcare Diagnostics Inc. · ACID, FOLIC, RADIOIMMUNOASSAY1 recall3 reports
- Atellica™ IM Fol 10995573Siemens Healthcare Diagnostics Inc. · ACID, FOLIC, RADIOIMMUNOASSAY1 recall3 reports
- ADVIA Centaur® FOL 10331250Siemens Healthcare Diagnostics Inc. · ACID, FOLIC, RADIOIMMUNOASSAY1 recall0 reports
- Atellica IM Fol 11206252Siemens Healthcare Diagnostics Inc. · ACID, FOLIC, RADIOIMMUNOASSAY1 recall0 reports
- Elecsys Folate 07027290190Roche Diagnostics GmbH · ACID, FOLIC, RADIOIMMUNOASSAY10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page