Atellica™ IM Fol 10995572Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Folic Acid (Fol) - Atellica IM - RGT - 140T · Class 2

All models of Atellica IM Fol (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00630414598963), received between Sep 30, 2019 and Dec 20, 2022 — classified as 7 malfunctions.

The most frequently reported specific device problem is “High Test Results” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K172201); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 9, 2023Class IIOpen, Classified

    Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.

    Z-1149-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • High Test Results5
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 20, 2022High Test Results; High Test Results

    THE CUSTOMER OBTAINED DISCORDANT ELEVATED ATELLICA IM FOLATE RESULTS FOR TWO PATIENTS. THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN(S), WHO DID NOT QUESTION THE RESULTS.  THE SAME SAMPLES WERE REPEATED USING FOLSERUM ON THE SAME ATELLICA IM ANALYZER, AND THE RESULTS WERE LOW. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FOLATE RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16015775 ↗
  • MalfunctionNov 23, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTS OBSERVATION OF AN ELEVATED PATIENT SAMPLE RESULT WHEN USING ATELLICA IM 1300 FOLATE (FOL) ASSAY. AFTER REPEATING WITH DILUTION, THE RESULTS WERE SIGNIFICANTLY LOWER THAN THE INITIAL RESULT. IT IS UNKNOWN WHICH RESULT IS CORRECT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FOLATE RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15852262 ↗
  • MalfunctionNov 1, 2022High Test Results; High Test Results

    THE CUSTOMER OBTAINED AN ELEVATED ATELLICA IM FOLATE (FOL) RESULT THAT WAS CONSIDERED DISCORDANT WITH THE REPEAT RESULT. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S) AND WAS NOT QUESTIONED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FOLATE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15707418 ↗
  • MalfunctionOct 31, 2022High Test Results; High Test Results

    THE CUSTOMER OBTAINED AN ELEVATED ATELLICA IM FOLATE (FOL) RESULT THAT WAS CONSIDERED DISCORDANT WITH THE REPEAT RESULT. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S) AND WAS NOT QUESTIONED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FOLATE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15699338 ↗
  • MalfunctionOct 28, 2022High Test Results; High Test Results

    THE CUSTOMER OBTAINED AN ELEVATED ATELLICA IM FOLATE (FOL) RESULT THAT WAS CONSIDERED DISCORDANT WITH THE REPEAT RESULT. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S) AND WAS NOT QUESTIONED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FOLATE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15690551 ↗
  • MalfunctionMay 17, 2021High Test Results; High Test Results

    THE CUSTOMER OBSERVED AN ELEVATED ATELLICA IM FOLATE RESULT COMPARED TO REPEAT TESTING. THE ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ELEVATED ATELLICA IM FOLATE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11838661 ↗
  • MalfunctionSep 30, 2019No Apparent Adverse Event; No Apparent Adverse Event

    DISCORDANT ATELLICA IM FOLATE RESULTS WERE OBTAINED ON SAMPLES FROM THREE PATIENTS WHEN COMPARING RESULTS BETWEEN REAGENT LOT 289 AND PREVIOUS LOT 286. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FOLATE RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9134972 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172201. One submission often covers many models, so check the product and lot details in each record.

  • Z-1151-2023 · Jan 9, 2023 · Open, Classified
    Atellica IM Folate 700 test kit (REF)-For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 112
  • Z-1150-2023 · Jan 9, 2023 · Open, Classified
    Atellica IM Folate 700 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 10995573

Other ACID, FOLIC, RADIOIMMUNOASSAY devices

All ACID, FOLIC, RADIOIMMUNOASSAY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page