ADVIA Centaur® SYPH 10492493Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Syphilis - CENTAUR - RGT - 200 tests · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00630414580173), received between Jun 9, 2017 and Nov 27, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K112343); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- False Negative Result2
- Adverse Event Without Identified Device or Use Problem1
- Off-Label Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 27, 2025False Negative Result; False Negative Result
THE CUSTOMER REPORTS AN OBSERVATION OF A FALSE NEGATIVE (NON-REACTIVE) ADVIA CENTAUR XP SYPHILIS (SYPH) RESULT FOR A PATIENT SAMPLE, WHICH WAS DISCORDANT RELATIVE TO REPEAT TESTING. AN INITIAL NON-REACTIVE RESULT WAS OBTAINED FOR A 49-YEAR-OLD FEMALE PATIENT. THIS INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S) AND QUESTIONED. THE SAMPLE WAS REPEATED ON THE ORIGINAL ADVIA CENTAUR ANALYZER AND OBTAINED A SIMILAR NON-REACTIVE RESULT. THE SAMPLE WAS THEN TESTED ON AN ALTERNATE METHOD AND OBTAINED A REACTIVE (POSITIVE) RESULT, WHICH WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). FOR TROUBLESHOOTING PURPOSE, THE SAME SAMPLE WAS REPEATED ON AN ALTERNATE ATELLICA IM ANALYZER AND A REACTIVE RESULT WAS OBTAINED. T. PALLIDUM PARTICLE AGGLUTINATION [TPPA] AND RAPID PLASMA REAGIN (RPR) RESULTS ON THIS SAMPLE WERE ALSO REACTIVE. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS IN DIAGNOSIS IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23665784 ↗MalfunctionJun 3, 2021False Negative Result; Off-Label Use; False Negative Result; Off-Label Use
DISCORDANT, FALSE NONREACTIVE ADVIA CENTAUR XPT SYPHILIS (SYPH) REPEAT RESULTS WERE OBTAINED FROM A REHYDRATED BLOOD SPOT SAMPLE AFTER AN INITIAL SYPH REACTIVE RESULT. THE NONREACTIVE REPEAT RESULTS WERE TESTED WITH A DIFFERENT REHYDRATED BLOOD SPOT SAMPLE. THE CUSTOMER PERFORMED REPEAT TESTING WITH AN ALTERNATE RPR (NON-TREPONEMAL) TEST METHOD ON ANOTHER REHYDRATED BLOOD SPOT SAMPLE, RESULTING REACTIVE. THE SYPH RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, NONREACTIVE ADVIA CENTAUR XPT SYPHILIS (SYPH) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11933825 ↗MalfunctionJun 9, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CUSTOMER OBSERVED A (B)(6) ADVIA CENTAUR XP (B)(6) RESULT FOR A SAMPLE THAT WAS (B)(6) BY OTHER METHODS. THERE ARE NOT REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE (B)(6) ADVIA CENTAUR XP (B)(6) RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6631081 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112343. One submission often covers many models, so check the product and lot details in each record.
- Z-1674-2021 · Apr 21, 2021 · Open, ClassifiedCentaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative determination of antibodies to Treponema pallidum in human serum or plasma
- Z-2228-2012 · Jul 17, 2012 · TerminatedSiemens ADVIA Centaur Syphilis (SYPH) Assay , 200 ReadyPack Reference Number 10492493 Product Usage: The ADVIA Centaur Syphilis (SYPH) assay is an in vitro diagnostic immunoass
Other ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM devices
- IMMULITE® 2000 Systems SYP Syphilis Screen 10471981Siemens Healthcare Diagnostics Products Ltd · ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM1 recall0 reports
- Elecsys Syphilis 09015051190Roche Diagnostics GmbH · ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM28 reports
- VITROS 6177333Ortho Clinical Diagnostics · ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM13 reports
- Elecsys Syphilis 09014977190Roche Diagnostics GmbH · ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM13 reports
- Elecsys Syphilis 06923348190Roche Diagnostics GmbH · ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM7 reports
- Elecsys Syphilis 07251378190Roche Diagnostics GmbH · ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM7 reports
- Atellica IM Syph 10995675Siemens Healthcare Diagnostics Inc. · ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM3 reports
- Trinity Biotech 800-970Clark Laboratories, Inc. · Enzyme Linked Immunoabsorption Assay, Treponema Pallidum1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page