Skip to content

Atellica CH VPA 11097512Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Valproic Acid (VPA) Atellica CH - RGT 400T · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00630414596013), received between Mar 4, 2024 and Mar 4, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2024High Test Results; High Test Results

    A DISCORDANT FALSELY ELEVATED VALPROIC ACID (VPA) RESULT WAS OBTAINED ON A PATIENT SAMPLE WHEN THE SAMPLE WAS REPROCESSED ON AN ATELLICA CH 930 ANALYZER. THE FALSELY ELEVATED RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE INITIAL LOW RESULT OBTAINED FOR THE SAME SAMPLE ON THE PREVIOUS DAY WAS REPORTED AS THE CORRECT RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED VALPROIC ACID (VPA) RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18824925 ↗
  • MalfunctionMar 4, 2024Low Test Results; Low Test Results

    DISCORDANT FALSELY DEPRESSED VALPROIC ACID (VPA) RESULTS WERE OBTAINED ON SIX (6) PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE FALSELY DEPRESSED RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). TWO (2) SAMPLES WERE REPROCESSED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE OTHER FOUR (4) SAMPLES WERE REPROCESSED ON THE ORIGINAL ATELLICA CH 930 ANALYZER. HIGHER RESULTS, CONSIDERED CORRECT, WERE OBTAINED AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED VALPROIC ACID (VPA) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18824920 ↗

Other ENZYME IMMUNOASSAY, VALPROIC ACID devices