Skip to content

Valproic Acid 5108438190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630999385), received between Feb 6, 2023 and Jul 15, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K060690); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results2
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2024Low Test Results; Low Test Results

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ONLINE TDM VALPROIC ACID (VALP2) ASSAY ON A COBAS 8000 COBAS C502 MODULE. INITIAL RESULT: <5 MG/L. THE PHYSICIAN QUESTIONED THE RESULT AND REQUESTED A REPEAT OF THE SAMPLE SINCE HE SUSPECTED AN INTENTIONAL DRUG INTOXICATION. REPEAT RESULT: 974 MG/L (AFTER DILUTION 1:10).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19741124 ↗
  • MalfunctionFeb 6, 2023Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results

    THE INITIAL REPORTER QUESTIONED LOW RESULTS FOR 10 PATIENT SAMPLES TESTED FOR ONLINE TDM VALPROIC ACID (VALP2) ON A COBAS 8000 C 502 MODULE. FOR ALL 10 PATIENT SAMPLES, THE INITIAL RESULT WAS 0.1 UG/ML WITH A DATA FLAG. WHEN THE SAMPLES WERE REPEATED, THE RESULTS WERE "MUCH HIGHER." THE CUSTOMER PROVIDED SPECIFIC RESULTS FOR 1 PATIENT SAMPLE. THE INITIAL RESULT WAS 0.1 UG/ML WITH A DATA FLAG. THE REPEAT RESULT WAS 74.9 UG/ML. THE QUESTIONABLE LOW RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE VALP2 REAGENT LOT NUMBER WAS 63608501 WITH AN EXPIRATION DATE OF 30-NOV-2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16306734 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K060690. One submission often covers many models, so check the product and lot details in each record.

  • Z-0920-2007 · Apr 30, 2007 · Terminated
    Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190.

Other ENZYME IMMUNOASSAY, VALPROIC ACID devices