Skip to content

Atellica CH GGT 11097597Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · G-Glutamyltransferase (GGT) Atellica CH - RGT 1792T · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 10, 2023Class IIOpen, Classified

    Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from AtellicaCH Total Bilirubin_2 (TBil_2), Atellica CH LDL Cholesterol (LDLC), Atellica CH Gamma-Glutamyl Transferase (GGT) and Atellica CH HDL Cholesterol (HDLC)

    Z-2473-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 23, 2020Class IIOpen, Classified

    Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability (within-run) listed in the Instructions For Use (IFU) on some analyzers.

    Z-0722-2021 · Root cause: Device Design · Recall details · FDA record ↗

Other ELECTROPHORETIC, GAMMA-GLUTAMYL TRANSPEPTIDASE ISOENZYMES devices

All ELECTROPHORETIC, GAMMA-GLUTAMYL TRANSPEPTIDASE ISOENZYMES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page