Atellica Gamma-Glutamyl Transferase (GGT) Reagent - Product… recallZ-0722-2021 · Siemens Healthcare Diagnostics, Inc. · initiated Nov 23, 2020
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability (within-run) listed in the Instructions For Use (IFU) on some analyzers.
FDA root cause category: Device Design
What the company is doing
Siemens Healthcare Diagnostics Urgent Field Safety Notice letter beginning November 23, 2020 via Fed Ex and outside the US the distribution is determined at the country level. Letter states reason for recall , health risk and action to take: For the product listed above, please perform the following steps: 1. Review and apply the preliminary precision performance characteristics of d8% CV at a GGT concentration of approximately 27 U/L 42 U/L provided. 2. Please review this letter with your Medical Director. 3. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. 4. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Customer Care Center or your local Siemens technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Customer Care Center or your local Siemens technical support representative. The precision section of the IFUs will be updated when the investigation is complete. The updated IFUs will be uploaded into Document Library where all registered users who opt in to receive alerts will be notified of the updated IFU.
Recall details
- Recall number
- Z-0722-2021
- Status
- Open, Classified
- Initiated by the company
- Nov 23, 2020
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 731 units US; 16824 units OUS
- Product code
- JQA
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Australia, Austria, Bahamas, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam. ¿¿¿¿ ¿¿¿¿¿
Product description
Atellica Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma. Siemens Material Number (SMN): 11097597
Code information (models, lots, serial numbers)
All lot codes. UDI: 00630414596440
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH GGT 11097597Siemens Healthcare Diagnostics Inc. · ELECTROPHORETIC, GAMMA-GLUTAMYL TRANSPEPTIDASE ISOENZYMES2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 86900).
- Z-0721-2021 · Open, ClassifiedADVIA Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasm…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page