Atellica IM CA 15-3 10995485Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · CA 15-3 Atellica IM - RGT 100T · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00630414598123), received between May 14, 2021 and Oct 25, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results3
- High Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 25, 2024Low Test Results; Low Test Results
THE CUSTOMER IDENTIFIED DISCORDANT (DEPRESSED) ATELLICA IM CA15-3, LOT 218, PATIENT SAMPLE RESULTS. THE DISCORDANCE WAS FOUND AFTER AN ISSUE WITH QUALITY CONTROL (QC) RESULTS WAS RESOLVED AND PATIENT SAMPLES WERE RETESTED WERE SUBSEQUENTLY RETESTED ON A DIFFERENT ATELLICA IM ANALYZER. THE INITIAL DEPRESSED PATIENT RESULTS WERE REPORTED, BUT NOT QUESTIONED BY PHYSICIAN(S). CORRECTED REPORTS WERE ISSUED FOR THE SAMPLES USING THE HIGHER RETEST RESULTS. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS IN ASSOCIATION WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20539857 ↗MalfunctionMay 14, 2021Low Test Results; Low Test Results
FALSELY LOW, ATELLICA IM CA 15-3 (CA 15-3) RESULTS WERE OBTAINED ON FOUR PATIENT SAMPLES AND CONSIDERED DISCORDANT COMPARED TO HIGHER RESULTS OBSERVED WHEN THE SAME SAMPLE WERE RE-TESTED ON ALTERNATE ATELLICA SYSTEMS. THE CUSTOMER'S QUALITY CONTROL (QC) RESULTS WERE OBSERVED TO BE OUT OF RANGE AT THE TIME OF THE INCIDENT. THE HIGHER REPEAT RESULTS FROM THE ALTERNATE ATELLICA SYSTEMS WERE REPORTED AS THE CORRECTED RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, ATELLICA IM CA 15-3 (CA 15-3) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11826397 ↗MalfunctionMay 14, 2021High Test Results; High Test Results
A FALSELY ELEVATED, ATELLICA IM CA 15-3 (CA 15-3) RESULT WAS OBTAINED ON A PATIENT SAMPLE AND CONSIDERED DISCORDANT COMPARED TO A LOWER RESULT OBSERVED WHEN THE SAME SAMPLE WAS RE-TESTED AFTER ASSAY RECALIBRATION USING A NEW REAGENT READYPACK. THE CUSTOMER'S QUALITY CONTROL (QC) RESULTS WERE OBSERVED TO BE OUT OF RANGE AT THE TIME OF THE INCIDENT AND REPEAT TESTING WAS PERFORMED AFTER ACCEPTABLE QC RESULTS WERE OBTAINED. THE LOWER REPEAT RESULT WAS REPORTED AS THE CORRECTED RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, ATELLICA IM CA 15-3 (CA 15-3) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11826393 ↗MalfunctionMay 14, 2021Low Test Results; Low Test Results
A FALSELY LOW, ATELLICA IM CA 15-3 (CA 15-3) RESULT WAS OBTAINED ON A PATIENT SAMPLE AND CONSIDERED DISCORDANT COMPARED TO A HIGHER RESULT OBSERVED WHEN THE SAME SAMPLE WAS RE-TESTED AFTER ASSAY RECALIBRATION USING A NEW REAGENT READYPACK. THE CUSTOMER'S QUALITY CONTROL (QC) RESULTS WERE OBSERVED TO BE OUT OF RANGE AT THE TIME OF THE INCIDENT AND REPEAT TESTING WAS PERFORMED AFTER ACCEPTABLE QC RESULTS WERE OBTAINED. THE HIGHER REPEAT RESULT WAS REPORTED AS THE CORRECTED RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, ATELLICA IM CA 15-3 (CA 15-3) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11826379 ↗
Other SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR devices
- ADVIA Centaur® CA 15-3® 10327620Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR14 reports
- Atellica IM BR 10995478Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR11 reports
- IMMULITE® 2000 Systems BRM BR-MA (CA15-3) 10380983Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR9 reports
- Atellica IM BR 10995477Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR8 reports
- ADVIA Centaur® BR +G 10333349Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR6 reports
- ADVIA Centaur® BR +G 10334837Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR4 reports
- Dimension Vista® Flex® reagent cartridge CA15-3 10464344Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR4 reports
- Atellica IM CA 15-3 10995486Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page