Atellica IM CA 15-3 10995485Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · CA 15-3 Atellica IM - RGT 100T · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00630414598123), received between May 14, 2021 and Oct 25, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results3
  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 25, 2024Low Test Results; Low Test Results

    THE CUSTOMER IDENTIFIED DISCORDANT (DEPRESSED) ATELLICA IM CA15-3, LOT 218, PATIENT SAMPLE RESULTS. THE DISCORDANCE WAS FOUND AFTER AN ISSUE WITH QUALITY CONTROL (QC) RESULTS WAS RESOLVED AND PATIENT SAMPLES WERE RETESTED WERE SUBSEQUENTLY RETESTED ON A DIFFERENT ATELLICA IM ANALYZER. THE INITIAL DEPRESSED PATIENT RESULTS WERE REPORTED, BUT NOT QUESTIONED BY PHYSICIAN(S). CORRECTED REPORTS WERE ISSUED FOR THE SAMPLES USING THE HIGHER RETEST RESULTS. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS IN ASSOCIATION WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20539857 ↗
  • MalfunctionMay 14, 2021Low Test Results; Low Test Results

    FALSELY LOW, ATELLICA IM CA 15-3 (CA 15-3) RESULTS WERE OBTAINED ON FOUR PATIENT SAMPLES AND CONSIDERED DISCORDANT COMPARED TO HIGHER RESULTS OBSERVED WHEN THE SAME SAMPLE WERE RE-TESTED ON ALTERNATE ATELLICA SYSTEMS. THE CUSTOMER'S QUALITY CONTROL (QC) RESULTS WERE OBSERVED TO BE OUT OF RANGE AT THE TIME OF THE INCIDENT. THE HIGHER REPEAT RESULTS FROM THE ALTERNATE ATELLICA SYSTEMS WERE REPORTED AS THE CORRECTED RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, ATELLICA IM CA 15-3 (CA 15-3) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11826397 ↗
  • MalfunctionMay 14, 2021High Test Results; High Test Results

    A FALSELY ELEVATED, ATELLICA IM CA 15-3 (CA 15-3) RESULT WAS OBTAINED ON A PATIENT SAMPLE AND CONSIDERED DISCORDANT COMPARED TO A LOWER RESULT OBSERVED WHEN THE SAME SAMPLE WAS RE-TESTED AFTER ASSAY RECALIBRATION USING A NEW REAGENT READYPACK. THE CUSTOMER'S QUALITY CONTROL (QC) RESULTS WERE OBSERVED TO BE OUT OF RANGE AT THE TIME OF THE INCIDENT AND REPEAT TESTING WAS PERFORMED AFTER ACCEPTABLE QC RESULTS WERE OBTAINED. THE LOWER REPEAT RESULT WAS REPORTED AS THE CORRECTED RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, ATELLICA IM CA 15-3 (CA 15-3) RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11826393 ↗
  • MalfunctionMay 14, 2021Low Test Results; Low Test Results

    A FALSELY LOW, ATELLICA IM CA 15-3 (CA 15-3) RESULT WAS OBTAINED ON A PATIENT SAMPLE AND CONSIDERED DISCORDANT COMPARED TO A HIGHER RESULT OBSERVED WHEN THE SAME SAMPLE WAS RE-TESTED AFTER ASSAY RECALIBRATION USING A NEW REAGENT READYPACK. THE CUSTOMER'S QUALITY CONTROL (QC) RESULTS WERE OBSERVED TO BE OUT OF RANGE AT THE TIME OF THE INCIDENT AND REPEAT TESTING WAS PERFORMED AFTER ACCEPTABLE QC RESULTS WERE OBTAINED. THE HIGHER REPEAT RESULT WAS REPORTED AS THE CORRECTED RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, ATELLICA IM CA 15-3 (CA 15-3) RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11826379 ↗

Other SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR devices

All SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR devices →