IMMULITE® 2000 Systems BRM BR-MA (CA15-3) 10380983Recalls & reported problems
Siemens Healthcare Diagnostics Products Ltd · CA 15-3 - IMMULITE - RGT - 200 Tests · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00630414961088), received between Nov 17, 2017 and Oct 4, 2023 — classified as 9 malfunctions.
The most frequently reported specific device problem is “High Test Results” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K013984); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- High Test Results4
- Nonstandard Device1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 4, 2023High Test Results; High Test Results
A FALSELY ELEVATED BR-MA (CA15-3) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT. THE SAMPLE WAS REPEATED ON AN ALTERNATE NON-SIEMENS INSTRUMENT. THE REPEAT RESULT WAS LOWER THAN THE ERRONEOUS RESULT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED BR-MA (CA15-3) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17872556 ↗MalfunctionFeb 10, 2022High Test Results; High Test Results
A CUSTOMER OBTAINED AN ELEVATED IMMULITE 2000 XPI BR-MA (CA15-3) RESULT ON A PATIENT SAMPLE (ABOVE REFERENCE RANGE RESULT). THE RESULT WAS REPORTED TO THE PHYSICIAN, WHO QUESTIONED THE RESULT. THE SAMPLE WAS SENT TO ANOTHER SITE AND WAS RETESTED WITH AN ALTERNATE METHOD AND A LOWER RESULT WAS OBTAINED (WITHIN REFERENCE RANGE RESULT). THE INITIAL ELEVATED RESULT WAS CONSIDERED DISCORDANT COMPARED TO THE ALTERNATE TEST METHOD RESULT. THE SAMPLE WAS SENT BACK TO THE CUSTOMER SITE AND WAS RETESTED ON THE SAME IMMULITE 2000 XPI INSTRUMENT. AN ELEVATED RESULT WAS OBTAINED (ABOVE REFERENCE RANGE RESULT). THE RESULT FROM THE ALTERNATE METHOD WAS CONSIDERED TO BE CORRECT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED CA15-3 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13505500 ↗MalfunctionFeb 10, 2022High Test Results; High Test Results
A CUSTOMER OBTAINED AN ELEVATED IMMULITE 2000 XPI BR-MA (CA15-3) RESULT ON A PATIENT SAMPLE (ABOVE REFERENCE RANGE RESULT). THE RESULT WAS REPORTED TO THE PHYSICIAN, WHO QUESTIONED THE RESULT. THE SAMPLE WAS SENT TO ANOTHER SITE AND WAS RETESTED WITH AN ALTERNATE METHOD AND A LOWER RESULT WAS OBTAINED (WITHIN REFERENCE RANGE RESULT). THE INITIAL ELEVATED RESULT WAS CONSIDERED DISCORDANT COMPARED TO THE ALTERNATE TEST METHOD RESULT. THE SAMPLE WAS SENT BACK TO THE CUSTOMER SITE AND WAS RETESTED ON THE SAME IMMULITE 2000 XPI INSTRUMENT. AN ELEVATED RESULT WAS OBTAINED (ABOVE REFERENCE RANGE RESULT). THE RESULT FROM THE ALTERNATE METHOD WAS CONSIDERED TO BE CORRECT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED CA15-3 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13505499 ↗MalfunctionAug 12, 2021High Test Results; High Test Results
THE CUSTOMER OBSERVED A DISCORDANT, ELEVATED RESULT ON A PATIENT SAMPLE USING THE IMMULITE 2000 XPI BR-MA (CA15-3) ASSAY AND QUESTIONED BY THE PHYSICIAN(S). THE SAME SAMPLE WHEN REPEATED WAS ELEVATED. THE SAME SAMPLE WAS REPEATED ON AN ALTERNATE CA15-3 TEST METHOD, RESULTING LOWER. THE LOWER CA15-3 RESULT AGREED WITH THE PATIENT'S CLINICAL PICTURE AND WAS REPORTED TO THE PHYSICIAN(S) AS THE CORRECT RESULT. THERE ARE NO KNOWN REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT BR-MA (CA15-3) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12315688 ↗MalfunctionDec 14, 2017Nonstandard Device; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SIEMENS TECHNICAL OPERATIONS INTERNAL INVESTIGATION HAS IDENTIFIED THAT DISCORDANT FALSELY HIGH RESULTS WERE OBTAINED ON TWO INVESTIGATIONAL SAMPLES ON AN IMMULITE 2000/ IMMULITE 2000 XPI INSTRUMENT WHEN USING THE CARCINOEMBRYONIC (CEA) ASSAY. THE INVESTIGATIONAL SAMPLE WAS SPIKED WITH BIOTIN AT MULTIPLE CONCENTRATIONS, AND RUN FOR INVESTIGATIONAL PURPOSES ON AN IMMULITE 2000/IMMULITE 2000 XPI INSTRUMENT. THE RESULT OF THE SAMPLE SPIKED WITH 1200 NG/ML OF BIOTIN WERE FALSELY HIGH WITH A BIAS OF +31% COMPARED TO THE RESULT OF THE SAMPLE WITHOUT BIOTIN. THERE WERE NO PATIENT SAMPLES IMPACTED AS THIS WAS FOR INVESTIGATIONAL PURPOSES ONLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7118364 ↗MalfunctionDec 14, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SIEMENS TECHNICAL OPERATIONS INTERNAL INVESTIGATION HAS IDENTIFIED THAT DISCORDANT FALSELY LOW RESULTS WERE OBTAINED ON TWO INVESTIGATIONAL SAMPLES ON AN IMMULITE 2000/ IMMULITE 2000 XPI INSTRUMENT WHEN USING THE 3G ALLERGEN SPECIFIC IGE ASSAY. FIVE SPECIFIC ALLERGENS WERE USED FOR THE TESTING - OLIVE SPECIFIC ALLERGEN, CAT DANDER, LATEX ALLERGEN, WHEAT ALLERGEN, AND CHIRONOMUS SPP ALLERGEN. THE TWO INVESTIGATIONAL SAMPLES WERE SPIKED WITH BIOTIN AT MULTIPLE CONCENTRATIONS, FOR EACH SPECIFIC ALLERGEN AND RUN FOR INVESTIGATIONAL PURPOSES ON AN IMMULITE 2000/IMMULITE 2000 XPI INSTRUMENT. A GREATER THAN ONE CLASS CHANGE (FALSELY LOW) IN RESULT WAS OBSERVED AT BIOTIN CONCENTRATIONS OF 250 NG/ML WITH OLIVE ALLERGEN, 100 NG WITH CAT DANDER ALLERGEN, 75 NG/ML WITH LATEX ALLERGEN, 250 NG/ML WITH WHEAT ALLERGEN AND 75 NG/ML OF CHIRONOMUS SPP ALLERGEN RESPECTIVELY COMPARED TO RESULTS OF THE TWO SAMPLES WITHOUT BIOTIN. THERE WERE NO PATIENT SAMPLES IMPACTED AS THIS WAS FOR INVESTIGATIONAL PURPOSES ONLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7118357 ↗MalfunctionDec 14, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SIEMENS TECHNICAL OPERATIONS INTERNAL INVESTIGATION HAS IDENTIFIED THAT DISCORDANT FALSELY HIGH RESULTS WERE OBTAINED ON TWO INVESTIGATIONAL SAMPLES ON AN IMMULITE 2000/ IMMULITE 2000 XPI INSTRUMENT WHEN USING THE OM-MA (CA 125) ASSAY. TWO INVESTIGATIONAL SAMPLES WERE SPIKED WITH BIOTIN AT MULTIPLE CONCENTRATIONS, AND RUN FOR INVESTIGATIONAL PURPOSES ON AN IMMULITE 2000/IMMULITE 2000 XPI INSTRUMENT. THE RESULTS OF THE TWO SAMPLES SPIKED WITH 75 NG/ML AND 250 NG/ML OF BIOTIN WERE FALSELY HIGH WITH A BIAS OF +41% AND +38% RESPECTIVELY COMPARED TO RESULTS OF THE TWO SAMPLES WITHOUT BIOTIN. THERE WERE NO PATIENT SAMPLES IMPACTED AS THIS WAS FOR INVESTIGATIONAL PURPOSES ONLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7118353 ↗MalfunctionDec 14, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SIEMENS TECHNICAL OPERATIONS INTERNAL INVESTIGATION HAS IDENTIFIED THAT DISCORDANT FALSELY HIGH RESULTS WERE OBTAINED ON TWO INVESTIGATIONAL SAMPLES ON AN IMMULITE/ IMMULITE 1000 INSTRUMENT WHEN USING THE OM-MA (CA 125) ASSAY. TWO INVESTIGATIONAL SAMPLES WERE SPIKED WITH BIOTIN AT MULTIPLE CONCENTRATIONS, AND RUN FOR INVESTIGATIONAL PURPOSES ON AN IMMULITE/ IMMULITE 1000 INSTRUMENT. THE RESULTS OF THE TWO SAMPLES SPIKED WITH 75 NG/ML OF BIOTIN WERE FALSELY HIGH WITH A BIAS OF +45% AND +38% RESPECTIVELY COMPARED TO RESULTS OF THE TWO SAMPLES WITHOUT BIOTIN. THERE WERE NO PATIENT SAMPLES IMPACTED AS THIS WAS FOR INVESTIGATIONAL PURPOSES ONLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7118352 ↗MalfunctionNov 17, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SIEMENS TECHNICAL OPERATIONS INTERNAL INVESTIGATION HAS IDENTIFIED THAT DISCORDANT FALSELY LOW RESULTS WERE OBTAINED ON TWO INVESTIGATIONAL SAMPLES ON AN IMMULITE 2000/ IMMULITE 2000 XPI INSTRUMENT WHEN USING THE BR-MA (CA15-3) ASSAY. THE TWO INVESTIGATIONAL SAMPLES WERE SPIKED WITH BIOTIN AT MULTIPLE CONCENTRATIONS AND RUN FOR INVESTIGATIONAL PURPOSES ON AN IMMULITE 2000/IMMULITE 2000 XPI INSTRUMENT. THE RESULTS OF THE TWO SAMPLES SPIKED WITH 500 NG/ML OF BIOTIN WERE FALSELY LOW WITH A BIAS OF -73% AND -53% RESPECTIVELY COMPARED TO RESULTS OF THE TWO SAMPLES WITHOUT BIOTIN. THERE WERE NO PATIENT SAMPLES IMPACTED AS THIS WAS FOR INVESTIGATIONAL PURPOSES ONLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7041463 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K013984. One submission often covers many models, so check the product and lot details in each record.
- Z-1677-2018 · Jan 2, 2018 · TerminatedIMMULITE /IMMULITE 1000 BR-MA (CA15-3)
- Z-1691-2018 · Jan 2, 2018 · TerminatedIMMULITE 2000/IMMULITE 2000 XPi BR-MA (CA15-3)
Other SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR devices
- ADVIA Centaur® CA 15-3® 10327620Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR14 reports
- Atellica IM BR 10995478Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR11 reports
- Atellica IM BR 10995477Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR8 reports
- ADVIA Centaur® BR +G 10333349Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR6 reports
- ADVIA Centaur® BR +G 10334837Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR4 reports
- Dimension Vista® Flex® reagent cartridge CA15-3 10464344Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR4 reports
- Atellica IM CA 15-3 10995485Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR4 reports
- Atellica IM CA 15-3 10995486Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page