Atellica IM PRL 10995655Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Prolactin (PRL) Atellica IM - RGT 250 · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00630414599731), received between Dec 4, 2020 and Nov 4, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Low Test Results3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 4, 2021Low Test Results; Low Test Results
A CUSTOMER OBSERVED A LOW ATELLICA IM PROLACTIN (PRL) RESULT ON 2 DIFFERENT SAMPLES TAKEN FROM THE SAME PATIENT. THE DEPRESSED RESULTS ARE CONSIDERED DISCORDANT COMPARED TO ALTERNATE METHOD TESTING RESULTS AND PATIENT CLINICAL HISTORY. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. A CORRECTED REPORT HAS NOT BEEN ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT PROLACTIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12753598 ↗MalfunctionNov 4, 2021Low Test Results; Low Test Results
A CUSTOMER OBSERVED A LOW ATELLICA IM PROLACTIN (PRL) RESULT ON 2 DIFFERENT SAMPLES TAKEN FROM THE SAME PATIENT. THE DEPRESSED RESULTS ARE CONSIDERED DISCORDANT COMPARED TO ALTERNATE METHOD TESTING RESULTS AND PATIENT CLINICAL HISTORY. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. A CORRECTED REPORT HAS NOT BEEN ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT PROLACTIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12753597 ↗MalfunctionDec 4, 2020Low Test Results; Low Test Results
A FALSE LOW ATELLICA IM PROLACTIN (PRL) RESULT WAS OBTAINED ON A PATIENT AND THE REPORTED RESULT WAS QUESTIONED BY THE PHYSICIAN(S). THE CUSTOMER PERFORMED REPEAT NEAT (UNDILUTED) TESTING OF THE SAMPLE, RESULTING LOW. FURTHER SAMPLE DILUTIONS WERE PERFORMED AND THE PRL RESULTS INCREASED AS THE DILUTION FACTOR INCREASED. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, LOW ATELLICA IM PROLACTIN (PRL) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10946377 ↗
Other RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN) devices
- VITROS 1849793Ortho Clinical Diagnostics · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)1 recall0 reports
- ARCHITECT 7K76-25Abbott Ireland · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)37 reports
- Alinity 07P6620Abbott Ireland · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)31 reports
- Elecsys Prolactin II 07027737190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)18 reports
- Elecsys Prolactin II 03203093190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)13 reports
- ARCHITECT 7K76-35Abbott Ireland · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)12 reports
- Alinity 07P6630Abbott Ireland · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)10 reports
- Atellica IM PRL 10995656Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page