Skip to content

Atellica IM PRGE 10995660Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Progesterone (PRGE) Atellica IM - RGT 90T · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00630414599786), received between Aug 20, 2020 and Feb 9, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “High Test Results” (3 reports).

11 other recalls cover devices cleared under the same FDA submission (K932955); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2024High Test Results; High Test Results

    THE CUSTOMER REPORTS OBSERVATION OF REPRODUCIBLY ELEVATED ATELLICA IM PROGESTERONE (PRGE) RESULTS FOR ONE PATIENT WHICH WERE DISCORDANT RELATIVE TO CLINICAL EXPECTATION AND ALTERNATE-METHOD TESTING. DISCORDANT ELEVATED PROGESTERONE RESULTS WERE OBSERVED FOR A 42-YEAR-OLD FEMALE PATIENT. AN INITIAL ELEVATED PRGE RESULT WAS OBTAINED AND QUESTIONED BY THE PHYSICIAN. A NEW SAMPLE WAS DRAWN AND TESTED AT ANOTHER LABORATORY (USING AN ALTERNATE TEST METHOD), AND A MUCH LOWER RESULT WAS RETURNED. THIS LOWER RESULT WAS CONSIDERED TO BE IN AGREEMENT WITH OTHER CLINICAL INDICATIONS. LATER REPEAT TESTING REPRODUCED THE INITIAL OBSERVATIONS (REPORTED SEPARATELY). THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS OR TREATMENT IN ASSOCIATION WITH THE OBSERVED DISCORDANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18679534 ↗
  • MalfunctionFeb 9, 2024High Test Results; High Test Results

    THE CUSTOMER REPORTS OBSERVATION OF REPRODUCIBLY ELEVATED ATELLICA IM PROGESTERONE (PRGE) RESULTS FOR ONE PATIENT WHICH WERE DISCORDANT RELATIVE TO CLINICAL EXPECTATION AND ALTERNATE-METHOD TESTING. DISCORDANT ELEVATED PROGESTERONE RESULTS WERE OBSERVED FOR A 42-YEAR-OLD FEMALE PATIENT. AN INITIAL DISCORDANT, ELEVATED PRGE RESULT WAS OBSERVED WHICH CONFLICTED WITH ALTERNATE-METHOD TESTING (REPORTED SEPARATELY). LATER REPEAT TESTING REPRODUCED THE INITIAL OBSERVATIONS (DISCORDANT PRGE RESULT ADDRESSED IN THIS REPORT). A NEW SAMPLE WAS DRAWN AND TESTED AT ANOTHER LABORATORY (USING AN ALTERNATE TEST METHOD), AND A MUCH LOWER RESULT WAS RETURNED. THIS LOWER RESULT WAS CONSIDERED TO BE IN AGREEMENT WITH OTHER CLINICAL INDICATIONS. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS OR TREATMENT IN ASSOCIATION WITH THE OBSERVED DISCORDANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18679522 ↗
  • MalfunctionAug 20, 2020High Test Results; High Test Results

    THE CUSTOMER OBSERVED ELEVATED ATELLICA IM 1300 PROGESTERONE (PRGE) RESULTS WHEN COMPARED TO THE IMMULITE 2000 PROGESTERONE ASSAY AND AN ALTERNATE METHOD. BOTH THE ATELLICA IM 1300 PRGE AND IMMULITE 2000 PROGESTERONE RESULTS WERE REPORTED TO THE PHYSICIAN. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ELEVATED ATELLICA IM 1300 PRGE RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10432294 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K932955. One submission often covers many models, so check the product and lot details in each record.

  • Z-1421-2017 · Jan 4, 2017 · Terminated
    IMMULITE 2000 Progesterone; Catalog Number: L2KPW2 L2KPW6; Siemens Material Number: 10381181, 10381170 Product Usage: For in vitro diagnostic use with the IMMULITE and IMMULITE 1
  • Z-1418-2017 · Jan 4, 2017 · Terminated
    ADVIA Centaur Progesterone; Catalog Number: 10310305 10315522 10333111 Product Usage: The PROG method is an in vitro diagnostic test for the quantitative measurement of progestero
  • Z-1420-2017 · Jan 4, 2017 · Terminated
    IMMULITE/ IMMULITE 1000 Progesterone; Catalog Number: LKPW1; Siemens Material Number: 10381128 Product Usage: For in vitro diagnostic use with the IMMULITE2000 Systems Analyzers-
  • Z-1419-2017 · Jan 4, 2017 · Terminated
    Dimension Vista LOCI Progesterone; Catalog Number: K6464; Siemens Material Number: 10461743 Product Usage: For in vitro diagnostic use in the quantitative determination of progest
  • Z-1078-2016 · Feb 4, 2016 · Terminated
    Siemens Healthcare Diagnostics ADVIA Centaur Calibrator E- 6- Pack Catalog Number: 04634762 SMN: 10321075 ADVIA Centaur Calibrator E For in vitro diagnostic use in calibration the
  • Z-1079-2016 · Feb 4, 2016 · Terminated
    Siemens Healthcare Diagnostic ADVIA Centaur Calibrator E- 6- Pack (Reference) Catalog Number: 04636889 SMN: 10335532 ADVIA Centaur Calibrator E For in vitro diagnostic use in cal
  • Z-1077-2016 · Feb 4, 2016 · Terminated
    Siemens Healthcare Diagnostics ADVIA Centaur Calibrator E - 2-Pack Catalog Number: 04634452 SMN: 10309079 ADVIA Centaur Calibrator E For in vitro diagnostic use in calibration t
  • Z-1230-2015 · Jan 27, 2015 · Terminated
    ADVIA Centaur Systems Progesterone Kit, (5 -pack); In Vitro Diagnostic; Catalog number: 01586287; SMN:10315522.
  • Z-1229-2015 · Jan 27, 2015 · Terminated
    ADVIA Centaur Systems Progesterone Kit, (1 -pack) In Vitro Diagnostic; Catalog number: 02382928; SMN: 10310305.
  • Z-1231-2015 · Jan 27, 2015 · Terminated
    ADVIA Centaur Systems Progesterone Kit, (5-pack Ref); In Vitro Diagnostic Catalog number: 02177364; SMN: 10333111.
  • Z-0164-2015 · Sep 26, 2014 · Terminated
    Siemens Calibrator E; SMN 10309079 (2 pack), SMN 10321075 (6 pack), SMN 10335532 (6 pack Ref)

Other RADIOIMMUNOASSAY, PROGESTERONE devices

All RADIOIMMUNOASSAY, PROGESTERONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page