Atellica CH ECRE3 11537216Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Enzymatic Creatinine_3 (ECre3) Atellica CH Assay 2000 Tests · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00630414610948), received between Nov 21, 2023 and Apr 25, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results3
  • High Test Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2024High Test Results; High Test Results

    THE CUSTOMER REPORTED TO SIEMENS THAT THEY OBTAINED AN ERRONEOUSLY ELEVATED ENZYMATIC CREATININE_3 (ECRE3) PATIENT SAMPLE RESULT ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUSLY ELEVATED RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME ATELLICA CH 930 ANALYZER. THE LOWER REPROCESSED RESULT OBTAINED WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). CUSTOMER STATED THAT THE PATIENT DISCHARGE WAS DELAYED DUE TO THE EVENT AND CONFIRMED THAT THERE WAS NO ADVERSE EFFECT OR IMPACT ON PATIENT CARE DUE TO THE EVENT. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED ENZYMATIC CREATININE_3 (ECRE3) RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19180890 ↗
  • MalfunctionApr 22, 2024High Test Results; High Test Results

    THE CUSTOMER REPORTED TO SIEMENS THAT THEY OBTAINED TWO (2) ERRONEOUSLY ELEVATED ENZYMATIC CREATININE_3 (ECRE3) RESULTS ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUSLY ELEVATED RESULTS WERE REPORTED TO THE PHYSICIAN(S), AND THE RESULTS WERE QUESTIONED. A NEW SAMPLE WAS DRAWN FROM THE SAME PATIENT, PROCESSED, AND THEN REPROCESSED ON AN ALTERNATE SYSTEM ON SUBSEQUENT DAYS IN ANOTHER LABORATORY. LOWER RESULTS WERE OBTAINED AND CONSIDERED CORRECT. ONE OF THE LOWER RESULTS FROM THE ALTERNATE SYSTEM WAS REPORTED AS THE CORRECT RESULT TO THE PHYSICIAN(S). THE CUSTOMER STATED THAT THE PATIENT WAS HOSPITALIZED FOR 2 DAYS BASED ON THE ERRONEOUSLY ELEVATED ECRE3 RESULTS. NO UNNECESSARY MEDICAL PROCEDURE OR TREATMENT WAS PERFORMED ON THE PATIENT. THE PATIENT WAS UNDER OBSERVATION IN THE HOSPITAL AND WAS RELEASED. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED ENZYMATIC CREATININE_3 (ECRE3) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19153305 ↗
  • MalfunctionDec 26, 2023Low Test Results; Low Test Results

    THE CUSTOMER REPORTED THAT DUE TO OUT-OF-RANGE QUALITY CONTROL IN THE EVENING (CLOSING), FOUR HUNDRED PATIENT SAMPLES HAD TO BE REPEATED FOR THE ATELLICA CH ENZYMATIC CREATININE_3 (ECRE3) ASSAY. THE INITIAL TESTING WAS PROCESSED ON THE ATELLICA CH 930 ANALYZER S/N (B)(6) AND THE REPEAT TESTING WAS PROCESSED ON THE SAME ATELLICA CH 930 ANALYZER AFTER A NEW CALIBRATION AND QUALITY CONTROL IN RANGE. THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN(S), WHO DID NOT QUESTION THE RESULTS. A CORRECTED REPORT WAS ISSUED. PATIENT DATA WAS NOT PROVIDED FOR EVALUATION, AND IT IS UNKNOWN IF RESULTS WERE DISCORDANT. ONE EXAMPLE WAS PROVIDED SHOWING AN INITIAL RESULT WAS LOWER THAN THE REPEAT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ENZYMATIC CREATININE_3 (ECRE3) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18393574 ↗
  • MalfunctionNov 21, 2023Low Test Results; Low Test Results

    THE CUSTOMER OBTAINED DISCORDANT ENZYMATIC CREATININE_3 (ECRE3) FOR A PATIENT SERUM SAMPLE PROCESSED ON THE ATELLICA CH 930 ANALYZER, SERIAL NUMBER (S/N) (B)(6) AND REAGENT LOT 123531. THE DISCORDANT RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE DISCORDANT RESULT WAS DISCOVERED WHEN THE CUSTOMER PERFORMED POST-ANALYSIS/REVIEW OF RESULTS FROM PATIENT SAMPLES WITH INITIAL LOW ECRE3 RESULTS. THE CUSTOMER RETESTED THE SAMPLES ON A DIFFERENT ANALYZER AND SAMPLE ID 1739795101 WAS FOUND TO BE DISCORDANT. THERE ARE NO ALLEGATIONS OF PATIENT OR USER INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ECRE3 RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18181522 ↗
  • MalfunctionNov 21, 2023Low Test Results; Low Test Results

    THE CUSTOMER OBTAINED DISCORDANT ENZYMATIC CREATININE_3 (ECRE3) RESULTS FOR PATIENT SERUM SAMPLES PROCESSED ON THE ATELLICA CH 930 ANALYZER, SERIAL NUMBER (S/N) (B)(6) REAGENT LOT 123531. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE DISCORDANT RESULTS WERE DISCOVERED WHEN THE CUSTOMER PERFORMED POST-ANALYSIS/REVIEW OF RESULTS FROM PATIENT SAMPLES WITH INITIAL LOW ECRE3 RESULTS. THE CUSTOMER RETESTED THE SAMPLES ON A DIFFERENT ANALYZER AND INITIAL RESULTS FOR SAMPLE ID (B)(4) AND SAMPLE ID (B)(4) WERE FOUND TO BE DISCORDANT. THERE ARE NO ALLEGATIONS OF PATIENT OR USER INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ECRE3 RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18181156 ↗

Other ENZYMATIC METHOD, CREATININE devices

All ENZYMATIC METHOD, CREATININE devices →