IMMULITE® 2000 Systems IGFBP-3 10381447Recalls & reported problems
Siemens Healthcare Diagnostics Products Ltd · IGF BP3 - IMMULITE - RGT - 200 Tests · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00630414966113), received between Jul 2, 2024 and Jan 7, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 7, 2026Low Test Results
THE CUSTOMER PERFORMED A ROUTINE RESET FOR THE IGFBP-3 (INSULIN-LIKE GROWTH FACTOR BINDING PROTEIN 3) ASSAY, MEASURED ON THE IMMULITE 2000 XPI ANALYZER #O6345, AFTER THE MOVING MEDIAN REGISTERED IN THE CUSTOMER'S SYSTEM WAS REJECTED. THE MOVING MEDIAN IS USED AS A PATIENT-BASED QUALITY CONTROL. FOR EVERY 30 PATIENT RESULTS, THE SYSTEM GENERATES A MEDIAN VALUE, WHICH IS PLOTTED ON A MONITORING GRAPH. FOR THIS ASSAY, THE MEDIAN REGISTERED ON THE GRAPH IS 4.97, WITH A STANDARD DEVIATION (SD) OF 0.27. IF THE MOVING MEDIAN VALUE EXCEEDS THE LIMITS ESTABLISHED BY THE 1X 3SD OR 2X 2SD REJECTION RULES, THE CUSTOMER PERFORMS AN EVALUATION OF THE PREVIOUS ROUTINE. IN THIS CASE, THE PREVIOUS ROUTINE WAS EVALUATED AND DISCREPANCIES IN RESULTS WERE IDENTIFIED IN PATIENT SAMPLES. THE PATIENT SAMPLES WERE REPEATED ON THE SAME ANALYZER THE FOLLOWING DAY. IT IS UNKNOWN WHETHER A DIFFERENT REAGENT WEDGE/BEAD WAS USED FOR THE REPEAT TESTING ON THE SAME ANALYZER. THE INITIAL RESULTS FOR THE DISCORDANT PATIENT SAMPLES WERE REPORTED AND WERE NOT QUESTIONED BY THE PHYSICIAN(S). THE LABORATORY DID NOT SEND CORRECTED RESULTS FOR THESE DISCORDANT PATIENT RESULTS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERV
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23988654 ↗MalfunctionJul 2, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED THAT FOLLOWING THE OBSERVATIONS OF A NEGATIVE BIAS WITH IGFBP-3 QUALITY CONTROL (QC) RESULTS AND FALSELY DEPRESSED IGFBP-3 PATIENT RESULTS INCONSISTENT WITH PREVIOUS RESULTS AND WITH THE PROFILE OF RESULTS OBTAINED IN THE LABORATORY, PREVIOUSLY TESTED PATIENT SAMPLES WERE REPEATED WITH THE IGFBP-3 ASSAY ON AN IMMULITE 2000 XPI INSTRUMENT. THE INITIAL RESULTS HAD BEEN REPORTED TO THE PHYSICIAN(S), WHO DID NOT QUESTION THE RESULTS. THE INITIAL RESULTS WERE FALSELY DEPRESSED IN COMPARISON TO THE REPEAT RESULTS. FOR THREE OF THE PATIENT SAMPLES, THE REPEAT RESULT WAS REPORTED, AS THE CORRECT RESULT, TO THE PHYSICIAN(S). FOR FOURTEEN OF THE PATIENT SAMPLES, THE LABORATORY'S MEDICAL STAFF CHOSE NOT TO RECTIFY THE RESULTS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED IGFBP-3 RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19661350 ↗
FDA clearance and classification
The FDA regulates product code CFL (radioimmunoassay, human growth hormone) under 21 CFR 862.1370 as a Class I device. 21 CFR 862.1370 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the IMMULITE® 2000 Systems IGFBP-3 10381447
Has the IMMULITE® 2000 Systems IGFBP-3 10381447 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00630414966113), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the IMMULITE® 2000 Systems IGFBP-3 10381447?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jul 2, 2024 and Jan 7, 2026. In those reports, the most common specific device problem is “Low Test Results” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the IMMULITE® 2000 Systems IGFBP-3 10381447?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the IMMULITE® 2000 Systems IGFBP-3 10381447 FDA approved?
Devices of this type (product code CFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Siemens Healthcare Diagnostics Products Ltd.
Who makes the IMMULITE® 2000 Systems IGFBP-3 10381447, and is it still sold?
The FDA's device database lists Siemens Healthcare Diagnostics Products Ltd as the company responsible for it (brand IMMULITE® 2000 Systems IGFBP-3, model 10381447). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE devices
- IMMULITE® 2000 Systems IGF-I 11128543Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE21 reports
- Elecsys IGF-1 07475896190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE7 reports
- Elecsys IGF-1 07475918190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE5 reports
- Elecsys IGFBP-3 07574690190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE2 reports
- IMMULITE 2000 Systems Growth Hormone 10381451Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE2 reports
- IMMULITE® 2000 Systems IGF-1 10381448Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE1 report
- Elecsys hGH 5390125190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page