IMMULITE® 2000 Systems IGFBP-3 10381447Recalls & reported problems

Siemens Healthcare Diagnostics Products Ltd · IGF BP3 - IMMULITE - RGT - 200 Tests · Class 1

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00630414966113), received between Jul 2, 2024 and Aug 14, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2026Low Test Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50252605 ↗
  • MalfunctionJan 7, 2026Low Test Results

    THE CUSTOMER PERFORMED A ROUTINE RESET FOR THE IGFBP-3 (INSULIN-LIKE GROWTH FACTOR BINDING PROTEIN 3) ASSAY, MEASURED ON THE IMMULITE 2000 XPI ANALYZER #O6345, AFTER THE MOVING MEDIAN REGISTERED IN THE CUSTOMER'S SYSTEM WAS REJECTED. THE MOVING MEDIAN IS USED AS A PATIENT-BASED QUALITY CONTROL. FOR EVERY 30 PATIENT RESULTS, THE SYSTEM GENERATES A MEDIAN VALUE, WHICH IS PLOTTED ON A MONITORING GRAPH. FOR THIS ASSAY, THE MEDIAN REGISTERED ON THE GRAPH IS 4.97, WITH A STANDARD DEVIATION (SD) OF 0.27. IF THE MOVING MEDIAN VALUE EXCEEDS THE LIMITS ESTABLISHED BY THE 1X 3SD OR 2X 2SD REJECTION RULES, THE CUSTOMER PERFORMS AN EVALUATION OF THE PREVIOUS ROUTINE. IN THIS CASE, THE PREVIOUS ROUTINE WAS EVALUATED AND DISCREPANCIES IN RESULTS WERE IDENTIFIED IN PATIENT SAMPLES. THE PATIENT SAMPLES WERE REPEATED ON THE SAME ANALYZER THE FOLLOWING DAY. IT IS UNKNOWN WHETHER A DIFFERENT REAGENT WEDGE/BEAD WAS USED FOR THE REPEAT TESTING ON THE SAME ANALYZER. THE INITIAL RESULTS FOR THE DISCORDANT PATIENT SAMPLES WERE REPORTED AND WERE NOT QUESTIONED BY THE PHYSICIAN(S). THE LABORATORY DID NOT SEND CORRECTED RESULTS FOR THESE DISCORDANT PATIENT RESULTS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERV

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23988654 ↗
  • MalfunctionJul 2, 2024Low Test Results; Low Test Results

    THE CUSTOMER REPORTED THAT FOLLOWING THE OBSERVATIONS OF A NEGATIVE BIAS WITH IGFBP-3 QUALITY CONTROL (QC) RESULTS AND FALSELY DEPRESSED IGFBP-3 PATIENT RESULTS INCONSISTENT WITH PREVIOUS RESULTS AND WITH THE PROFILE OF RESULTS OBTAINED IN THE LABORATORY, PREVIOUSLY TESTED PATIENT SAMPLES WERE REPEATED WITH THE IGFBP-3 ASSAY ON AN IMMULITE 2000 XPI INSTRUMENT. THE INITIAL RESULTS HAD BEEN REPORTED TO THE PHYSICIAN(S), WHO DID NOT QUESTION THE RESULTS. THE INITIAL RESULTS WERE FALSELY DEPRESSED IN COMPARISON TO THE REPEAT RESULTS. FOR THREE OF THE PATIENT SAMPLES, THE REPEAT RESULT WAS REPORTED, AS THE CORRECT RESULT, TO THE PHYSICIAN(S). FOR FOURTEEN OF THE PATIENT SAMPLES, THE LABORATORY'S MEDICAL STAFF CHOSE NOT TO RECTIFY THE RESULTS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED IGFBP-3 RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19661350 ↗

Other RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE devices

All RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE devices →