Elecsys IGF-1 07475918190Recalls & reported problems
Roche Diagnostics GmbH · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613336140864), received between Oct 5, 2022 and Aug 7, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Low Test Results1
- Non Reproducible Results1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50175662 ↗MalfunctionNov 3, 2025Output Problem; Output Problem
THE INITIAL REPORTER COMPLAINED OF HIGH RESULTS FOR MULTIPLE PATIENT SAMPLES TESTED FOR ELECSYS IGF-1 (IGF-1) STARTING AT THE BEGINNING OF (B)(6) 2025 ON A COBAS 8000 E 801 MODULE. THE CUSTOMER HAD PREVIOUSLY USED THE IMMULITE INSTRUMENT, AND THE IGF-1 RESULTS WERE LOWER. AN EXAMPLE OF QUESTIONABLE RESULTS WAS PROVIDED FOR 1 PATIENT SAMPLE. ON (B)(6) 2025, THE RESULT FROM THE SIEMENS METHOD WAS 706.0 NG/ML. ON (B)(6) 2025, THE RESULT FROM THE E801 MODULE WAS 1107 NG/ML. THE REPEAT RESULT WAS 1153 NG/ML. THE PATIENT HAD A PREVIOUS RESULT (B)(6) 2025) FROM THE IMMULITE INSTRUMENT OF 673 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23450187 ↗MalfunctionAug 29, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS IGF-1 ASSAY ON A COBAS 8000 CORE UNIT. INITIAL RESULT: 1232 NG/ML. THE CUSTOMER QUESTIONED THE INITIAL RESULT AS IT DID NOT MATCH THE PATIENT'S CLINICAL DIAGNOSIS AND REPEATED THE SAMPLE. NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE THE LABORATORY. REPEAT RESULT: 940 NG/ML. THE REPEAT RESULT WAS DEEMED TO BE CORRECT AND REPORTED TO THE CLINIC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22920613 ↗MalfunctionNov 4, 2022High Test Results; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS IGF-1 ASSAY RESULTS FOR 1 PATIENT SAMPLE ON A COBAS E 801 MODULE COMPARED TO A SIEMENS IMMULITE. ON (B)(6) 2022, THE IGF-1 RESULT WAS 89 NMOL/L. ON (B)(6) 2022, THE SIEMENS IGF-1 RESULT WAS 60 NMOL/L. THE EXACT DATE OF TESTING WAS REQUESTED BUT NOT PROVIDED. ON (B)(6) 2022, THE SAMPLE WAS REPEATED ON THE E801 MODULE AND THE RESULT WAS 94 NMOL/L. THE ROCHE REFERENCE RANGE IS 15 - 64 NMOL/L. THE SIEMENS REFERENCE RANGE IS 23 - 66 NMOL/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE E801 MODULE SERIAL NUMBER IS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15729514 ↗MalfunctionOct 5, 2022Low Test Results; Low Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE LOW IGF-1 ELECSYS RESULTS FROM COBAS E 801 ANALYTICAL UNIT SERIAL NUMBER (B)(4) THAT DID NOT MATCH THE CLINICAL PICTURE OF THE PATIENTS. ONE EXAMPLE WAS PROVIDED OF AN INITIAL RESULT OF 80.2 NG/ML AND A REPEAT RESULT OF 149 NG/ML. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY AND WERE QUESTIONED BY THE PHYSICIAN. THE REPEAT RESULTS FIT THE PATIENTS' HISTORIES AND WERE CONSIDERED VALID.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15544191 ↗
Other RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE devices
- IMMULITE® 2000 Systems IGF-I 11128543Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE21 reports
- Elecsys IGF-1 07475896190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE7 reports
- IMMULITE® 2000 Systems IGFBP-3 10381447Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE3 reports
- Elecsys IGFBP-3 07574690190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE2 reports
- IMMULITE 2000 Systems Growth Hormone 10381451Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE2 reports
- Elecsys hGH 5390125190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE1 report
- IMMULITE® 2000 Systems IGF-1 10381448Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, HUMAN GROWTH HORMONE1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page