Midas Rex® AS075Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT AS075 LEGEND 7.5CM STRAIGHT 2.4MM · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169087583), received between Aug 21, 2024 and May 29, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
9 other recalls cover devices cleared under the same FDA submission (K020069); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- No Apparent Adverse Event1
- Noise, Audible1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 29, 2026Detachment of Device or Device Component; No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION
Patient problems: Insufficient Information
MAUDE report 25327616 ↗MalfunctionMay 25, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE ATTACHMENT BEARING WAS DETACHED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25269732 ↗MalfunctionAug 21, 2024Detachment of Device or Device Component; Noise, Audible; Detachment of Device or Device Component; Noise, Audible
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF NOISE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. REPAIR ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO DETACHED BEARING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20019686 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K020069. One submission often covers many models, so check the product and lot details in each record.
- Z-1602-2016 · Apr 8, 2016 · TerminatedMedtronic Midas Rex Dissecting Tool Legend 10 cm 4mm BA DIAM. Intended for use in neurosurgery.
- Z-0321-2016 · Oct 21, 2015 · TerminatedTool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
- Z-0546-2010 · Nov 30, 2009 · TerminatedMedtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotom
- Z-2156-2008 · May 30, 2008 · TerminatedMedtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General S
- Z-1574-05 · Sep 1, 2005 · TerminatedMedtronic Midas Rex Classic, GS and Legend dissecting tools manufactured by Medtronic Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, tX 76137 USA, Rx only, S
- Z-1067-05 · Jun 29, 2005 · TerminatedMedtronic MIDAS REX Legend High Speed Pneumatic System, Gold Touch Motor. Manufactured by Medtronic, Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, TX 76137
- Z-0278-05 · Nov 11, 2004 · TerminatedMedtronic Midas Rex Legend High Speed Pneumatic System
- Z-0213-05 · Nov 3, 2004 · TerminatedMedtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
- Z-0419-03 · Nov 11, 2002 · TerminatedLegend Lubricant System (Legend Gold, Gold Touch and Platinum motors and Legend Lubricant/Diffuser cartridge)
Other MOTOR, DRILL, PNEUMATIC devices
- NA 05.001.080Synthes GmbH · MOTOR, DRILL, PNEUMATIC432 reports
- NA CSR60Synthes GmbH · MOTOR, DRILL, PNEUMATIC221 reports
- Midas Rex® AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC163 reports
- Maestro Air 5407-200-000Stryker Corporation · MOTOR, DRILL, PNEUMATIC112 reports
- Maestro 5400210060Stryker Corporation · MOTOR, DRILL, PNEUMATIC95 reports
- MIDAS REX™ AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC91 reports
- CORE, Maestro 5400200000Stryker Corporation · MOTOR, DRILL, PNEUMATIC75 reports
- Maestro Air 5407-200-00HStryker Corporation · MOTOR, DRILL, PNEUMATIC56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page